DOI QR코드

DOI QR Code

A Study on Quality Improvement and Advancement of Negative Pressure Isolation Stretcher: Using FOCUS-PDCA

음압격리들것의 품질개선과 고도화 연구: FOCUS-PDCA 모형을 활용하여

  • Choi, Hyunchul (Department of Health Services Management, Graduate Colleage, Kyung Hee University) ;
  • Seo, Seul-Ki (Department of Health Services Management, Graduate Colleage, Kyung Hee University) ;
  • Byun, Sungkwan (WooSung Engineering Co., Ltd., Research Institute) ;
  • Chang, Hyejung (Department of Management, School of Management, Kyung Hee University) ;
  • Park, Sangchan (Department of Technology and Society, The State University of New York Korea)
  • 최현철 (경희대학교 일반대학원 의료경영학과) ;
  • 서슬기 (경희대학교 일반대학원 의료경영학과) ;
  • 변성관 ((주)우성이엔지 기업부설연구소) ;
  • 장혜정 (경희대학교 경영대학 경영학부) ;
  • 박상찬 (한국뉴욕주립대학교 기술경영학과)
  • Received : 2023.11.29
  • Accepted : 2023.12.04
  • Published : 2023.12.31

Abstract

Purpose: Authors suggest a new alternative of preparedness capabilities for health care institutes to react to the new types of communicable diseases and the pandemic situations through the quality improvement and advancement of Negative Pressure Isolation Stretcher (NPIS). Methods: We employed FOCUS-PDCA, a representative quality improvement model. First, we identified and materialized necessary components of existing NPIS to be improved. Second, we fabricated an advanced NPIS prototype. Third, we carried out a performance test by the accreditation institute and an empirical evaluation at the communicable diseases trauma center, to locate the improvement pain points of the prototype. Finally, we draw a conclusion on quality improvement points to be considered during the production stage. Results: The suggested, NPIS outperforms the existing one in terms of communicable disease controls, patient treatments and the safety during the patient transfer, and the ease of use. Additionally, the suggested NPIS turns out to be a proper alternative to satisfy the requirements of long range transfer of patients with communicable disease. Conclusion: Authors developed a quality improved, NPIS prototype. There are several points that still remain to be improved. Additionally, we expect that the concurrent use of the suggested NPIS with the existing one will fortify the preparedness capabilities on reacting to the communicable diseases.

Keywords

Acknowledgement

이 연구는 2023년도 산업통상자원부 및 한국산업기술평가관리원(KEIT) 연구비 지원에 의한 연구임('20015023').

References

  1. Chen, Y., Zheng, J., Wu, D., Zhang, Y., & Lin, Y. 2020. Application of the PDCA cycle for standardized nursing management in a COVID-19 intensive care unit. Annals of Palliative Medicine 9(3):1198-1205.  https://doi.org/10.21037/apm-20-1084
  2. ECO Zlin Co.,Ltd. PRODUCT Bio-Bag EBV-30/40 IN-CH Datasheet. Available from: https://www.egozlin.cz/en/products/cbrn-protection/patient-isolation 
  3. Han, S. H., Kim, S. R., Cha, K. S., Son, H. J., Shin, M. J., Choi, J. R., & Choi, J. Y. (2020). Infection control during surgery for patients with confirmed or suspected coronavirus disease-19. Korean Journal of Healthcare-Associated Infection Control and Prevention 25(1):4-10.  https://doi.org/10.14192/kjicp.2020.25.1.4
  4. Huang, Y., Huang, Y., Yi, L., Pan, W., & Chen, Y. 2023. A FOCUS-PDCA quality improvement model for reducing the distribution defect rate of sterile packages. Scientific Reports 13(1):15016. 
  5. Jeong, J. W., Kwon, H. M., Hong, S. H., & Lee, M. K. 2020. Estimation of Reproduction Number for COVID-19 in Korea. Journal of Korean Society for Quality Management 48(3):493-498.  https://doi.org/10.7469/JKSQM.2020.48.3.493
  6. Kim, D. H., 2018. Economic analysis of negative pressure isolation room installation - Real option application. Health Insurance Review & Assessement service(HIRA) Policy Trends 12(4):61-76. 
  7. Kim, K. H., Hong, K. J., Haam, S. H., & Choi, J. W. 2018. Efficacy and Usability of Patient Isolation Transport Module for CBRN Disaster: A Manikin Simulation Study. Fire Science and Engineering 32(3):116-122.  https://doi.org/10.7731/KIFSE.2018.32.3.116
  8. Kim, S. C., Kong, S. Y., Park, G. J., Lee, J. H., Lee, J. K., Lee, M. S., & Han, H. S. 2021. Effectiveness of negative pressure isolation stretcher and rooms for SARS-CoV-2 nosocomial infection control and maintenance of South Korean emergency department capacity. The American Journal of Emergency Medicine 45:483-489.  https://doi.org/10.1016/j.ajem.2020.09.081
  9. Kim Sejin. 2022. A Study on effective prediction and response measures for new infectious disease: Focusing on expert Delphi survey. Korea Institute of Public Administration Research 40:56-63. 
  10. Korea Disease Control and Prevention Agency. 2023. Guidelines for responding to class 1 infectious diseases, animal influenza, and human infectious diseases. 
  11. Li, Y., Wang, H., & Jiao, J. 2020. The application of strong matrix management and PDCA cycle in the management of severe COVID-19 patients. Critical Care 24:1-4.  https://doi.org/10.1186/s13054-019-2683-3
  12. Liu, Q., Zhang, Q., Chen, J., & Wang, T. 2022. Research on the effect of FOCUS-PDCA quality management on improving the maintenance and use of monitors in hospitals. Frontiers in Computing and Intelligent Systems 1(1):62-65.  https://doi.org/10.54097/fcis.v1i1.1463
  13. Lyng, J. W. 2021. Development of a Negative Pressure Isolation System for Containment, Filtration, and Disinfection of Airborne Diseases for Use in Hospitals, Ambulances, and Alternate Care Settings. In Frontiers in Biomedical Devices. American Society of Mechanical Engineers 84812:V001T04A008. 
  14. Maruta, R. 2012. Maximizing knowledge work productivity: a time constrained and activity visualized PDCA cycle. Knowledge and Process Management 19(4):203-214.  https://doi.org/10.1002/kpm.1396
  15. Mehand, M. S., Al-Shorbaji, F., Millett, P., & Murgue, B. 2018. The WHO R&D Blueprint: 2018 review of emerging infectious diseases requiring urgent research and development efforts. Antiviral Research 159:63-67.  https://doi.org/10.1016/j.antiviral.2018.09.009
  16. Park, D. J., Kim, H. G., & Yun, W. Y. 2007. Research Trend and Future of ISO 9000 Quality Management System: Literature Review. Journal of Korean Society for Quality Management 35(3):1-22. 
  17. Pena, M., Neu, D. T., Feng, H. A., Hammond, D. R., Mead, K. R., & Banerjee, R. K. 2023. Use of a Negative Pressure Containment Pod Within Ambulance-Workspace During Pandemic Response. Journal of Medical Devices 17(1):011009. 
  18. Ramdan, M. I., Abd Rahim, I., Ab Rahman, N. H. N., Hawary, A. F., Ismail, M. A., Selvan, M., ... & Pink, L. W. 2021. Development Of an Affordable Negative-Pressure Full-Body Isolation Pod for Covid-19 Patient Transportation. Journal of Advanced Research in Fluid Mechanics and Thermal Sciences 88(3):137-144.  https://doi.org/10.37934/arfmts.88.3.137144
  19. Realyvasquez-Vargas, A., Arredondo-Soto, K. C., Carrillo-Gutierrez, T., & Ravelo, G. 2018. Applying the Plan-Do-Check-Act (PDCA) cycle to reduce the defects in the manufacturing industry. A case study. Applied Sciences 8(11):2181. 
  20. Saxena, S., Ramer, L., & Shulman, I. A. 2004. A comprehensive assessment program to improve blood-administering practices using the FOCUS-PDCA model. Transfusion 44(9):1350-1356.  https://doi.org/10.1111/j.1537-2995.2004.03117.x
  21. Yoo. H. J. 1998. Total quality management Activities Evaluation (TAE) Model by the traditional scoring system and the efficiency measuring system. Journal of Korean Society for Quality Management 26(3):100-107. 
  22. Wei, W., Wang, S., Wang, H., & Quan, H. 2020. The application of 6S and PDCA management strategies in the nursing of COVID-19 patients. Critical Care 24:1-4.  https://doi.org/10.1186/s13054-019-2683-3
  23. Wosem Co.,Ltd. PRODUCT, WS-02 Datasheet. Available from: http://www.wosem.co.kr/bbs/board.php?bo_table=product1&wr_id=21) 
  24. Xie, C., Zhang, J., Luo, J., Jian, M., Zhao, T., Wang, J., ... & Shi, Y. 2022. FOCUS-PDCA can effectively optimize the critical value of test items. Journal of Medical Biochemistry 41(3):347-354. https://doi.org/10.5937/jomb0-34958