The Advent of Cuffless Mobile Device Blood Pressure Measurement: Remaining Challenges and Pitfalls

  • Hae-Young Lee (Department of Internal Medicine, Seoul National University Hospital) ;
  • Thilo Burkard (Medical Outpatient and Hypertension Clinic, ESH Hypertension Centre of Excellence, University Hospital Basel)
  • Received : 2021.12.14
  • Accepted : 2021.12.28
  • Published : 2022.03.01

Abstract

Blood pressure measurement (BPM) is an essential part of medical examination, and therefore accuracy of BPM devices is crucial. Over the past few years, there has been a rise in new BPM techniques using photoplethysmographic (PPG) signals and complex algorithms for blood pressure estimation. Especially the combination of a mobile device or a smartphone with a camera using PPG may potentially revolutionize BPM in the future. The first-ever BPM application to be approved as a medical device was one by the Korean Ministry of Food and Drug Safety in 2020, despite the lack of robust scientific evidence proving its validity. While the prospect of using these novel BPM devices is an opportunity, there are also some critical issues around calibration and utility in different patient groups that need to be resolved before they can be incorporated into daily clinical practice.

Keywords

References

  1. Stergiou GS, O'Brien E, Myers M, et al. STRIDE BP international initiative for accurate blood pressure measurement: Systematic review of published validation studies of blood pressure measuring devices. J Clin Hypertens (Greenwich) 2019;21:1616-22. https://doi.org/10.1111/jch.13710
  2. Vischer AS, Burkard T. Principles of blood pressure measurement - Current techniques, office vs ambulatory blood pressure measurement. Adv Exp Med Biol 2017;956:85-96. https://doi.org/10.1007/5584_2016_49
  3. Williams B, Mancia G, Spiering W, et al. 2018 ESC/ESH guidelines for the management of arterial hypertension. Eur Heart J 2018;39:3021-104. https://doi.org/10.1093/eurheartj/ehy339
  4. Vaidyanathan S, Jermany J, Yeh C, Bizot MN, Camisasca R. Aliskiren, a novel orally effective renin inhibitor, exhibits similar pharmacokinetics and pharmacodynamics in Japanese and Caucasian subjects. Br J Clin Pharmacol 2006;62:690-8. https://doi.org/10.1111/j.1365-2125.2006.02696.x
  5. Samsung Electronics. Samsung announces blood pressure monitoring application for Galaxy Watch devices [Internet]. Suwon: Samsung; 2020 [cited 2020 September 16]. Available from: https://news.samsung.com/global/samsung-announces-blood-pressure-monitoring-application-for-galaxy-watch-devices?utm_source=ext_newsletter&utm_medium=email.
  6. Picone DS, Deshpande RA, Schultz MG, et al. Nonvalidated home blood pressure devices dominate the online marketplace in Australia: major implications for cardiovascular risk management. Hypertension 2020;75:1593-9. https://doi.org/10.1161/HYPERTENSIONAHA.120.14719
  7. Dey J, Gaurav A, Tiwari VN. InstaBP: cuff-less blood pressure monitoring on smartphone using single PPG sensor. Annu Int Conf IEEE Eng Med Biol Soc 2018;2018:5002-5.
  8. Matsumura K, Rolfe P, Toda S, Yamakoshi T. Cuffless blood pressure estimation using only a smartphone. Sci Rep 2018;8:7298.
  9. Chandrasekhar A, Natarajan K, Yavarimanesh M, Mukkamala R. An iPhone application for blood pressure monitoring via the oscillometric finger pressing method. Sci Rep 2018;8:13136.
  10. Luo H, Yang D, Barszczyk A, et al. Smartphone-based blood pressure measurement using transdermal optical imaging technology. Circ Cardiovasc Imaging 2019;12:e008857.
  11. Vybornova A, Polychronopoulou E, Wurzner-Ghajarzadeh A, Fallet S, Sola J, Wuerzner G. Blood pressure from the optical Aktiia Bracelet: a 1-month validation study using an extended ISO81060-2 protocol adapted for a cuffless wrist device. Blood Press Monit 2021;26:305-11. https://doi.org/10.1097/MBP.0000000000000531
  12. Association for the Advancement of Medical Instrumentation (US); American National Standards Institute; International Organization for Standardization. AAMI/ANSI/ISO 81060-2:2013, Non-invasive Sphygmomanometers - Part 2: Clinical Investigation of Automated Measurement Type. Arlington (VA): Association for the Advancement of Medical Instrumentation; 2013.
  13. Plante TB, Urrea B, MacFarlane ZT, et al. Validation of the instant blood pressure smartphone App. JAMA Intern Med 2016;176:700-2. https://doi.org/10.1001/jamainternmed.2016.0157
  14. Raichle CJ, Eckstein J, Lapaire O, et al. Performance of a blood pressure smartphone App in pregnant women: the iPARR trial (iPhone App compared with standard RR measurement). Hypertension 2018;71:1164-9. https://doi.org/10.1161/HYPERTENSIONAHA.117.10647
  15. Xing X, Ma Z, Zhang M, Zhou Y, Dong W, Song M. An unobtrusive and calibration-free blood pressure estimation method using photoplethysmography and biometrics. Sci Rep 2019;9:8611.
  16. Stergiou GS, Palatini P, Asmar R, et al. Recommendations and practical guidance for performing and reporting validation studies according to the universal standard for the validation of blood pressure measuring devices by the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO). J Hypertens 2019;37:459-66. https://doi.org/10.1097/HJH.0000000000002039
  17. Stergiou GS, Alpert B, Mieke S, et al. A universal standard for the validation of blood pressure measuring devices: Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Collaboration Statement. J Hypertens 2018;36:472-8. https://doi.org/10.1097/HJH.0000000000001634
  18. Padwal R. Cuffless blood pressure measurement: How did accuracy become an afterthought? Am J Hypertens 2019;32:807-9. https://doi.org/10.1093/ajh/hpz070
  19. Institute of Electrical and Electronics Engineers (IEEE) Engineering in Medicine and Biology Society. IEEE Std 1708-2014. IEEE Standard for Wearable, Cuffless Blood Pressure Measuring Devices. New York (NY): Institute of Electrical and Electronics Engineers, Inc.; 2014.
  20. Parati G, Stergiou GS, Asmar R, et al. European Society of Hypertension guidelines for blood pressure monitoring at home: a summary report of the Second International Consensus Conference on Home Blood Pressure Monitoring. J Hypertens 2008;26:1505-26. https://doi.org/10.1097/HJH.0b013e328308da66
  21. Bilo G, Zorzi C, Ochoa Munera JE, Torlasco C, Giuli V, Parati G. Validation of the Somnotouch-NIBP noninvasive continuous blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010. Blood Press Monit 2015;20:291-4. https://doi.org/10.1097/MBP.0000000000000124
  22. Krisai P, Vischer AS, Kilian L, Meienberg A, Mayr M, Burkard T. Accuracy of 24-hour ambulatory blood pressure monitoring by a novel cuffless device in clinical practice. Heart 2019;105:399-405.
  23. Socrates T, Krisai P, Vischer AS, Meienberg A, Mayr M, Burkard T. Improved agreement and diagnostic accuracy of a cuffless 24-h blood pressure measurement device in clinical practice. Sci Rep 2021;11:1143.
  24. Diaz KM, Tanner RM, Falzon L, et al. Visit-to-visit variability of blood pressure and cardiovascular disease and all-cause mortality: a systematic review and meta-analysis. Hypertension 2014;64:965-82. https://doi.org/10.1161/HYPERTENSIONAHA.114.03903
  25. Jeon YW, Kim HC. Factors associated with awareness, treatment, and control rate of hypertension among Korean young adults aged 30-49 years. Korean Circ J 2020;50:1077-91. https://doi.org/10.4070/kcj.2020.0208
  26. Lee HY, Oh GC, Sohn IS, et al. Suboptimal management status of younger hypertensive population in Korea. Korean Circ J 2021;51:598-606. https://doi.org/10.4070/kcj.2020.0542