Calibration of cylindrical NaI(Tl) gamma-ray detector intended for truncated conical radioactive source

  • Badawi, Mohamed S. (Department of Physics, Faculty of Science, Beirut Arab University) ;
  • Thabet, Abouzeid A. (Department of Biomedical Equipment Technology, Faculty of Applied Health Sciences Technology, Pharos University in Alexandria)
  • Received : 2021.04.23
  • Accepted : 2021.10.12
  • Published : 2022.04.25

Abstract

The computation of the solid angle and the detector efficiency is considering to be one of the most important factors during the measuring process for the radioactivity, especially the cylindrical γ-ray NaI(Tl) detectors nowadays have applications in several fields such as industry, hazardous for health, the gamma-ray radiation detectors grow to be the main essential instruments in radiation protection sector. In the present work, a generic numerical simulation method (NSM) for calculating the efficiency of the γ-ray spectrometry setup is established. The formulas are suitable for any type of source-to-detector shape and can be valuable to determine the full-energy peak and the total efficiencies and P/T ratio of cylindrical γ-ray NaI(Tl) detector setup concerning the truncated conical radioactive source. This methodology is based on estimate the path length of γ-ray radiation inside the detector active medium, inside the source itself, and the self-attenuation correction factors, which typically use to correct the sample attenuation of the original geometry source. The calculations can be completed in general by using extra reasonable and complicate analytical and numerical techniques than the standard models; especially the effective solid angle, and the detector efficiency have to be calculated in case of the truncated conical radioactive source studied condition. Moreover, the (NSM) can be used for the straight calculations of the γ-ray detector efficiency after the computation of improvement that need in the case of γ-γ coincidence summing (CS). The (NSM) confirmation of the development created by the efficiency transfer method has been achieved by comparing the results of the measuring truncated conical radioactive source with certified nuclide activities with the γ-ray NaI(Tl) detector, and a good agreement was obtained after corrections of (CS). The methodology can be unlimited to find the theoretical efficiencies and modifications equivalent to any geometry by essential sufficiently the physical selective considered situation.

Keywords

Acknowledgement

Dr. Mohamed. S. Badawi as one of the authors would like to show appreciation to the Radiation Physics Laboratory (RPL), Physics Department, Faculty of Science, Alexandria University, 21511 Alexandria, Egypt for letting him obtain the measurements by himself.

References

  1. Y.J. Wang, B.E. Patt, J.S. Iwanczyk, S.R. Cherry, Y. Shao, High efficiency CsI(Tl)/HgI/sub 2/gamma ray spectrometers, IEEE Trans. Nucl. Sci. 42 (4) (1995) 601-605, https://doi.org/10.1109/23.467906.
  2. S. Rawat, M. Tyagi, G. Anil Kumar, S.C. Gadkari, Efficiency studies on Gd3Ga3Al2O12:Ce scintillators: simulations and measurements, IEEE Trans. Nucl. Sci. 65 (8) (2018) 2109-2113, https://doi.org/10.1109/TNS.2018.2823749.
  3. K. Debertin, R.G. Helmer, Gamma- and X-Ray Spectrometry with Semiconductor Detectors, North-Holland, Netherlands, 1988.
  4. C.M. Salgado, L.E.B. Brandao, R. Schirru, C.M.N.A. Pereira, C.C. Conti, Validation of a NaI(Tl) detector's model developed with MCNP-X code, Prog. Nucl. Energy 59 (2012) 19-25, https://doi.org/10.1016/j.pnuce ne.2012.03.006.
  5. H.D. Tam, N.T.H. Yen, L.B. Tran, H.D. Chuong, T.T. Thanh, Optimization of the Monte Carlo simulation model of NaI(Tl) detector by Geant4 code, Appl. Radiat. Isot. 130 (2017) 75-79, https://doi.org/10.1016/j.apradiso.2017.09.020.
  6. H.-X. Shi, B.-X. Chen, T.-Z. Li, D. Yun, Precise Monte Carlo simulation of gamma-ray response functions for an NaI(Tl) detector, Appl. Radiat. Isot. 57 (4) (2002) 517-524, https://doi.org/10.1016/s0969-8043(02)00140-9.
  7. D. Novkovic, M. Durasevic, A. Kandic, I. Vukanac, B. Seslak, Z. Milosevic, Coincidence summing corrections for point and volume 152 Eu sources, Appl. Radiat. Isot. 107 (2016) 138-144, https://doi.org/10.1016/j.apradiso.2015.10.015.
  8. M.-C. Lepy, T. Altzitzoglou, M.J. Anagnostakis, D. Arnold, M. Capogni, A. Ceccatelli, P. De Felice, R. Dersch, P. Dryak, A. Fazio, L. Ferreux, M. Guardati, J.B. Han, S. Hurtado, K.L. Karfopoulos, S. Klemola, P. Kovar, K.B. Lee, R. Ocone, T. Vidmar, Intercomparison of methods for coincidence summing corrections in gamma-ray spectrometry, Appl. Radiat. Isot. 68 (7-8) (2010) 1407-1412, https://doi.org/10.1016/j.apradiso.2010.01.012.
  9. A.M. El-Khatib, B.A. Salem, M.S. Badawi, M.M. Gouda, A.A. Thabet, M.I. Abbas, Full-Energy peak efficiency of an NaI(Tl) detector with coincidence summing correction showing the effect of the source-to-detector distance, Chin. J. Phys. 55 (2) (2017) 478-489, https://doi.org/10.1016/j.cjph.2016.11.013.
  10. M.S. Badawi, S.I. Jovanovic, A.A. Thabet, A.M. El-Khatib, A.D. Dlabac, B.A. Salem, M.M. Gouda, N.N. Mihaljevic, K.S. Almugren, M.I. Abbas, Calibration of 4π NaI(Tl) detectors with coincidence summing correction using new numerical procedure and ANGLE4 software, AIP Adv. 7 (3) (2017), 035005, https://doi.org/10.1063/1.4978214.
  11. M.S. Badawi, I. Ruskov, M.M. Gouda, A.M. El-Khatib, M.F. Alotiby, M.M. Mohamed, A.A. Thabet, M.I. Abbas, A numerical approach to calculate the full-energy peak efficiency of HPGe well-type detectors using the effective solid angle ratio, J. Instrum. 9 (7) (2014) P07030, https://doi.org/10.1088/1748-0221/9/07/p07030. P07030.
  12. M.M. Gouda, M.S. Badawi, A.M. El-Khatib, M.M. Mohamed, A.A. Thabet, M.I. Abbas, Calibration of well-type NaI(Tl) detector using a point sources measured out the detector well at different axial distances, J. Instrum. 10 (3) (2015) P03022, https://doi.org/10.1088/1748-0221/10/03/p03022. P03022.
  13. A. Hamzawy, D.N. Grozdanov, M.S. Badawi, F.A. Aliyev, A.A. Thabet, M.I. Abbas, I.N. Ruskov, A.M. El-Khatib, Y.N. Kopatch, M.M. Gouda, New numerical simulation method to calibrate the regular hexagonal NaI(Tl) detector with radioactive point sources situated non-axial, Rev. Sci. Instrum. 87 (11) (2016), 115105, https://doi.org/10.1063/1.4966990.
  14. A.A. Thabet, A.D. Dlabac, S.I. Jovanovic, M.S. Badawi, N.N. Mihaljevic, A.M. El-Khatib, M.M. Gouda, M.I. Abbas, Experimental verification of gamma-efficiency calculations for scintillation detectors in ANGLE 4 software, Nucl. Technol. Radiat. Protect. 30 (1) (2015) 35-46, https://doi.org/10.2298/ntrp1501035t.
  15. A. El-Khatib, A.A. Thabet, M.A. Elzaher, M.S. Badawi, B.A. Salem, Study on the effect of the self-attenuation coefficient on γ-ray detector efficiency calculated at low and high energy regions, Nucl. Eng. Technol. 46 (2) (2014) 217-224, https://doi.org/10.5516/NET.04.2013.077.
  16. M.I. Abbas, M.S. Badawi, I.N. Ruskov, A.M. El-Khatib, D.N. Grozdanov, A.A. Thabet, Yu N. Kopatch, M.M. Gouda, V.R. Skoy, Calibration of a single hexagonal NaI(Tl) detector using a new numerical method based on the efficiency transfer method, Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip. 771 (2015) 110-114, https://doi.org/10.1016/j.nima.2014.10.056.
  17. A.M. El-Khatib, M.S. Badawi, A.A. Thabet, S.I. Jovanovic, M.M. Gouda, M.M. Mohamed, A.D. Dlabac, M. Abd-Elzaher, N.N. Mihaljevic, M.I. Abbas, Well-type NaI(Tl) detector efficiency using analytical technique and ANGLE 4 software based on radioactive point sources located out the well cavity, Chin. J. Phys. 54 (3) (2016) 338-346, https://doi.org/10.1016/j.cjph.2016.03.019.
  18. M.M. Gouda, A. Hamzawy, M.S. Badawi, A.M. El-Khatib, A.A. Thabet, M.I. Abbas, Mathematical method to calculate full-energy peak efficiency of detectors based on transfer technique, Indian J. Phys. 90 (2) (2015) 201-210, https://doi.org/10.1007/s12648-015-0737-1.
  19. Canberra Industries, Inc, LabSOCS Calibration Software Model S574, Technical Reference Manual, 2013.
  20. L. Done, L.C. Tugulan, D. Gurau, F. Dragolici, C. Alexandru, Comparison of LabSOCS and GESPECOR codes used in gamma-ray spectrometry, Appl. Radiat. Isot. 109 (2016) 539-543, https://doi.org/10.1016/j.apradiso.2015.11.036.
  21. https://www.canberra.com/literature/isocs/application_notes/ISOCS-LabSOCS-App-Note-C39530.pdf.