Evaluation of Source Identification Method Based on Energy-Weighting Level with Portal Monitoring System Using Plastic Scintillator

  • Received : 2020.05.07
  • Accepted : 2020.08.21
  • Published : 2020.09.30

Abstract

Background: Radiation portal monitors (RPMs) involving plastic scintillators installed at the border inspection sites can detect illicit trafficking of radioactive sources in cargo containers within seconds. However, RPMs may generate false alarms because of the naturally occurring radioactive materials. To manage these false alarms, we previously suggested an energy-weighted algorithm that emphasizes the Compton-edge area as an outstanding peak. This study intends to evaluate the identification of radioactive sources using an improved energy-weighted algorithm. Materials and Methods: The algorithm was modified by increasing the energy weighting factor, and different peak combinations of the energy-weighted spectra were tested for source identification. A commercialized RPM system was used to measure the energy-weighted spectra. The RPM comprised two large plastic scintillators with dimensions of 174 × 29 × 7 ㎤ facing each other at a distance of 4.6 m. In addition, the in-house-fabricated signal processing boards were connected to collect the signal converted into a spectrum. Further, the spectra from eight radioactive sources, including special nuclear materials (SNMs), which were set in motion using a linear motion system (LMS) and a cargo truck, were estimated to identify the source identification rate. Results and Discussion: Each energy-weighted spectrum exhibited a specific peak location, although high statistical fluctuation errors could be observed in the spectrum with the increasing source speed. In particular, 137Cs and 60Co in motion were identified completely (100%) at speeds of 5 and 10 km/hr. Further, SNMs, which trigger the RPM alarm, were identified approximately 80% of the time at both the aforementioned speeds. Conclusion: Using the modified energy-weighted algorithm, several characteristics of the energy weighted spectra could be observed when the used sources were in motion and when the geometric efficiency was low. In particular, the discrimination between 60Co and 40K, which triggers false alarms at the primary inspection sites, can be improved using the proposed algorithm.

Keywords

References

  1. Siciliano ER, Ely JH, Kouzes RT, Milbrath BD, Schweppe JE, Stromswold DC. Comparison of PVT and NaI (Tl) scintillators for vehicle portal monitor applications. Nucl Instrum Methods Phys Res A. 2005;550:647-674. https://doi.org/10.1016/j.nima.2005.05.056
  2. Geelhood BD, Ely JH, Hansen RR, Kouzes RT, Schweppe JE, Warner RA. Overview of portal monitoring at border crossings. Proceedings of 2003 IEEE Nuclear Science Symposium. Conference Record (IEEE Cat. No. 03CH37515); 2003 Oct 19-25; Portland, OR. p. 513-517.
  3. Ely J, Kouzes R, Schweppe J, Siciliano E, Strachan D, Weier D. The use of energy windowing to discriminate SNM from NORM in radiation portal monitors. Nucl Instrum Methods Phys Res A. 2006;560:373-387. https://doi.org/10.1016/j.nima.2006.01.053
  4. Stromswold DC, Darkoch JW, Ely JH, Hansen RR, Kouzes RT, Milbrath BD, et al. Field tests of a NaI(Tl)-based vehicle portal monitor at border crossings. Proceedings of the IEEE Symposium Conference Record Nuclear Science; 2004 Oct 16-22; Rome, Italy. p. 196-200.
  5. Keyser RM, Sergent F, Twomey TR, Upp DL. Minimum detectable activity estimates for a germanium-detector based spectroscopic portal monitor. Proceedings of the INMM 47th Annual Meeting Conference; 2006 Jul 16-20; Nashville, TN.
  6. Kangas LJ, Keller PE, Siciliano ER, Kouzes RT, Ely JH. The use of artificial neural networks in PVT-based radiation portal monitors. Nucl Instrum Methods Phys Res A. 2008;587:398-412. https://doi.org/10.1016/j.nima.2008.01.065
  7. Jordan DV, Geelhood BD, Reeder PL, Stephens DL, Craig RA, McIntyre JI. Progress Report on the Advanced Large-Area Plastic Scintillators (ALPS) Project (No. PNNL-14283). Richland, WA: Pacific Northwest National Lab.; 2003.
  8. Shin WG, Lee HC, Choi CI, Park CS, Kim HS, Min CH. A Monte Carlo study of an energy-weighted algorithm for radionuclide analysis with a plastic scintillation detector. Appl Radiat Isot. 2015;101:53-59. https://doi.org/10.1016/j.apradiso.2015.03.014
  9. Lee HC, Shin WG, Park HJ, Yoo DH, Choi CI, Park CS, et al. Validation of energy-weighted algorithm for radiation portal monitor using plastic scintillator. Appl Radiat Isot. 2016;107:160-164. https://doi.org/10.1016/j.apradiso.2015.10.019
  10. Lee HC, Shin WG, Choi HJ, Choi CI, Park CS, Kim HS, Min CH. Radioisotope identification using an energy-weighted algorithm with a proof-of-principle radiation portal monitor based on plastic scintillators. Appl Radiat Isot. 2020;156:109010. https://doi.org/10.1016/j.apradiso.2019.109010
  11. Lee HC, Choi HJ, Koo BT, Choi CI, Park CS, Kim HS, et al. Dynamic radionuclide identification using energy weighted algorithm with commercialized radiation portal monitor based on plastic scintillators. Radiat Phys Chem. 2020;170:108645. https://doi.org/10.1016/j.radphyschem.2019.108645
  12. American National Standards Institute. American National Standard for Performance Criteria for Spectroscopy-Based Portal Monitors Used in Homeland Security. Washington, DC: American National Standards Institute; 2015. IEEE/ANSI N42.38-2015.

Cited by

  1. Pseudo-Gamma Spectroscopy Based on Plastic Scintillation Detectors Using Multitask Learning vol.21, pp.3, 2021, https://doi.org/10.3390/s21030684