DOI QR코드

DOI QR Code

Design and stress analysis of femur bone implant with composite plates

  • Ramakrishna, S. (Department of Mechanical Engineering, Gayatri Vidya Parishad College of Engineering (Autonomous)) ;
  • Pavani, B. (Department of Mechanical Engineering, Gayatri Vidya Parishad College of Engineering (Autonomous))
  • Received : 2019.05.04
  • Accepted : 2020.08.18
  • Published : 2020.03.25

Abstract

Development of lightweight implant plates are important to reduce the stress shielding effect for a prosthesis of femur bone fractures. Stainless steel (SS-316L) is a widely used material for making implants. Stress shielding effect and other issues arise due to the difference in mechanical properties of stainless steel when compared with bone. To overcome these issues, composite materials seem to be a better alternative solution. The comparison is made between two biocompatible composite materials, namely Ti-hydroxyapatite and Ti-polypropylene. "Titanium (Ti)" is fiber material while "hydroxyapatite" and "polypropylene" are matrix materials. These two composites have Young's modulus closer to the bone than stainless steel. Besides the variety of bones, present paper constrained to femur bone analysis only. Being heaviest and longest, the femur is the most likely to fail among all bone failures in human. Modelling of the femur bone, screws, implant and assembly was carried out using CATIA and static analysis was carried out using ANSYS. The femur bone assembly was analyzed for forces during daily activities. Ti-hydroxyapatite and Ti-polypropylene composite implants induced more stress in composite implant plate, results less stress induced in bone leading to a reduction in shielding effect than stainless steel implant plate thus ensuring safety and quick healing for the patient.

Keywords

References

  1. Arifin, A., Sulong, A.B., Muhamad, N., Syarif, J. and Ramli, M.I. (2014), "Material processing of hydroxyapatite and titanium alloy (HA/Ti) composite as implant materials using powder metallurgy", Mater. Des., 55, 165-175. https://doi.org/10.1016/j.matdes.2013.09.045.
  2. Bergmann, G., Graichen, F. and Rohlmann, A. (1995), "Is staircase walking a risk for the fixation of hip implants", J. Biomech., 28(5), 535-553. https://doi.org/10.1016/0021-9290(94)00105-D.
  3. Bola, A.M., Ramos, A. and Simoes, J.A. (2016), "Sensitivity analysis for finite element modeling of humeral bone and cartilage", Biomater. Biomech. Bioeng., 3(2), 71-84. https://doi.org/10.12989/bme.2016.3.2.071.
  4. Florea, R.M. and Carcea, I. (2012), "Polymer matrix composites - routes and properties", Int. J. Mod. Manuf. Technol., 4(1), 59-64.
  5. Madeti, B.K., Chalamalasetti, S.R. and Bollapragada, S.S.S.R. (2015), "Biomechanics of knee joint: Review", Front. Mech. Eng., 10(2), 176-186. https://doi.org/10.1007/s11465-014-0306-x.
  6. Madeti, B.K., Chalamalasetti, S.R. and Priya Gugulothu, S. (2018a), "Buckling and failure analysis of bonesin hip joint", J. Mech. Med. Biol., 18(5), 1850052. https://doi.org/10.1142/S0219519418500525.
  7. Madeti, B.K., Chalamalasetti, S.R. and Bollapragada, S.S.S.R. (2018b), "Force evaluation and stress distribution at possible weight and structure of femur bone in pelvis frame while standing", Int. J. Med. Eng. Inform., 10(3), 235-251. https://doi.org/10.1504/IJMEI.2018.093353.
  8. Mallick, P.K. (2007), Fiber-Reinforced Composites: Materials, Manufacturing and Design, CRC Press, New York, U.S.A.
  9. Metan, S.S., Mohankumar, G.C. and Krishna, P. (2016), "Sensitivity analysis of shoulder joint muscles by using the FEM model", Biomater. Biomech. Bioeng., 3(2), 115-127. https://doi.org/10.12989/bme.2016.3.2.115.
  10. Mohammed, M.T. (2015), "Mechanical and wear properties of HPT-biomedical titanium: A review", Biomater. Biomech. Bioeng., 2(3), 185-196. http://dx.doi.org/10.12989/bme.2015.2.3.185.
  11. Niinomi, M. and Nakai, M. (2011), "Titanium-based biomaterials for preventing stress shielding between implant devices and bone", Int. J. Biomater., 15, 836587. https://doi.org/10.1155/2011/836587.
  12. Petkovic, D., Radenkovic, G. and Mitkovic, M. (2012), "Fractographic investigation of failure in stainless steel orthopedic plates", Facta Univ. Ser. Mech. Eng., 10(1), 7-14.
  13. Ridzwan, M.I.Z., Shuib, S., Hassan, A.Y., Shokri, A.A. and Ibrahim, M.M. (2007), "Problem of stress shielding and improvement to the hip implant designs", J. Med. Sci., 7(3), 460-467. https://doi.org/10.3923/jms.2007.460.467
  14. Shireesha, Y., Ramana, S.V. and Rao, P.G. (2013), "Modelling and static analysis of femur bone by using different implant materials", IOSR J. Mech. Civ. Eng., 7(4), 82-91. https://doi.org/10.9790/1684-0748291
  15. Sopyana, I., Melb, M., Rameshc, S. and Khalidd, K.A. (2007), "Porous hydroxyapatite for artificial bone applications", Sci. Technol. Adv. Mater., 8(1-2), 116-123. https://doi.org/10.1016/j.stam.2006.11.017.
  16. Stronach, B.M., Roach, M.D. and John, K.R.S. (2016), "Failure of Emperion modular femoral stem with implant analysis", Arthroplast. Today, 2(1), 11-14. https://doi.org/10.1016/j.artd.2015.11.004.
  17. Tsouknidas, A., Anagnostidis, K., Maliaris, G. and Michailidis, N. (2012), "Fracture risk in the femoral hip region: A finite element analysis supported experimental approach", J. Biomech., 45(11), 1959-1964. https://doi.org/10.1016/j.jbiomech.2012.05.011.
  18. Wang, M. (2016), "Optimization of femoral prosthesis based on comprehensive evaluation of structure and material properties", J. Mech. Med. Biol., 16(8), 5-17. https://doi.org/10.1142/S0219519416400133.
  19. Wirtz, D.C., Schiffers, N., Pandorf, T. and Radermacher, K. (2000), "Critical evaluation of known bone material properties to realize anisotropic FE-simulation of the proximal femur", Biomechanics, 31(10), 1325-1330. https://doi.org/10.1016/S0021-9290(00)00069-5.
  20. Xu, Y., Meng, H., Yin, H., Sun, Z., Peng, J., Xu, X., Guo, Q., Xu, W., Yu, X., Yuan, Z., Xiao, B., Wang, C., Wang, Y., Liu, S., Lu, S., Wang, Z. and Wang, A. (2018), "Quantifying the degradation of degradable implants and bone formation in the femoral condyle using micro-CT 3D reconstruction", Exp. Ther. Med., 15(1), 93-102. https://doi.org/10.3892/etm.2017.5389.