Evaluation of Image Quality and Stability of Radiation Output according to Change in Tube Voltage and Sensitivity when Abdomen and Pelvis Examination of Digital Radiography (DR)

디지털 방사선 시스템(DR)의 복부와 골반부 검사 시 관전압과 감도 변화에 따른 영상 화질과 방사선 출력의 안정성 평가

  • 황준호 (경희대학교병원 영상의학과) ;
  • 양형진 (고려대학교 의용과학대학원 의학물리학과) ;
  • 최지안 (경희대학교병원 영상의학과) ;
  • 이경배 (경희대학교병원 영상의학과)
  • Received : 2019.09.17
  • Accepted : 2019.10.15
  • Published : 2019.12.28

Abstract

The purpose of this study is to find the optimal method for clinical application by analyzing image quality and radiation output according to parameter combination when using the Automatic Exposure Control (AEC). The experimental method combines 70, 81 kVp with sensitivity S200, S400, S800 and S1000 of the Automatic Exposure Control for Entrance Surface Dose (ESD), current volume, Signal to Noise Ratio (SNR), Contrast to Noise Ratio (CNR), Time-to-Radiation Dose Curve in abdomen and pelvis. And then, image quality and radiation output stability were evaluated. As a results, Entrance Surface Dose, current volume, Signal to Noise Ratio, Contrast to Noise Ratio decreased as the tube voltage and sensitivity were set higher. In addition, the higher tube voltage and sensitivity, the Time-to-Radiation Dose Curve showed a poor output stability. In conclusion, the higher the combination of tube voltage and sensitivity in the use of Automatic Exposure Control, the more problems can be seen in image quality and stability of the radiation output. Therefore, a relatively low combination of tube voltage and sensitivity showed that the image quality and radiation output stability could be optimized by minimizing the error range that would occur when the detector recognized a combination of parameters.

본 연구는 자동노출제어장치 사용 시 파라미터 조합에 따른 영상 화질과 방사선 출력을 분석하여 임상에 적용할 수 있는 최적의 방법을 모색하고자 하였다. 실험방법은 관전압 70, 81 kVp와 자동노출제어장치(Automatic Exposure Control, AEC)의 감도 S200, S400, S800, S1000을 조합해서 복부와 골반부의 입사표면선량, 관전류량, 신호 대 잡음비(Signal to Noise Ratio, SNR), 대조도 대 잡음비(Contrast to Noise Ratio, CNR), 시간-방사선량 곡선을 구하였다. 그 후 영상 화질과 출력의 안정성을 평가하였다. 그 결과 입사표면선량, 관전류량, 신호 대 잡음비, 대조도 대 잡음비는 관전압과 감도가 높게 설정될수록 감소하였다. 또한 관전압과 감도가 높게 설정될수록 시간-방사선량 곡선은 출력의 안정성이 줄어드는 양상을 보였다. 결론적으로 복부와 골반부 검사 시 관전압과 감도를 높게 조합할수록 검출기는 영상 화질과 방사선 출력을 정상적으로 재현해내지 못하였다. 따라서 비교적 낮은 관전압과 감도를 조합하여 검출기가 파라미터의 조합을 인식할 때 발생하는 오차 범위를 최소화해야 영상 화질과 방사선 출력의 안정성을 최적화할 수 있다는 것을 알 수 있었다.

Keywords

References

  1. G. Li, T. C. Greene, T. K. Nishino, and C. E. Willis, "Evaluation of Cassette-based Digital Radiography Detectors using Standardized Image Quality Metrics: AAPM TG-150 Draft Image Detector Tests," J. of Applied Clinical Medical Physics, Vol.17, No.5, pp.391-417, 2016. https://doi.org/10.1120/jacmp.v17i5.6008
  2. D. S. Kim and E. Kim, "Noise Power Spectrum of the Fixed Pattern Noise in Digital Radiography Detectors : Noise Power Spectrum of the Fixed Pattern Noise," Med. Phys., Vol.43, No.6, pp.2765-2773, 2016. https://doi.org/10.1118/1.4948691
  3. N. W. Marshall, C. V. Ongeval, and H. Bosmans, "Performance Evaluation of a Retrofit Digital Detector-based Mammography System," Physica Medica : Eur. J. of Med. Phys., Vol.32, No.2, pp.312-322, 2016. https://doi.org/10.1016/j.ejmp.2016.01.002
  4. S. C. Kim, H. K. Lee, Y. S. Lee, and J. H. Cho, "Evaluation of Automatic Exposure Control System Chamber for the Dose Optimization when Examining Pelvic in Digital Radiography," J. of X-ray Science and Technology, Vol.23, No.3, pp.321-330, 2015. https://doi.org/10.3233/XST-150492
  5. B. Louise, F. Ronan, C. Conor, G. Aoife, D. Mark, G. Dermot, O. Geraldine, and D. Anita, "Doses under Automatic Exposure Control (AEC) for Direct Digital Radiographic (DDR) X-ray Systems," Radiation Protection Dosimetry, Vol.147, No.1/2, pp.210-214, 2011. https://doi.org/10.1093/rpd/ncr301
  6. M. Reade, S. Cournane, J. McCavana, and J. Lucey, "Modeling CTDlvol through Analysis of the Localiser Image: Towards Estimating Factors Influencing Automatic Exposure Control Tube Output," Physica Medica : Eur. J. of Med. Phys., Vol.52, pp.172-172, 2018.
  7. ICRP, ICRP Publication 135, 2017.
  8. ICRP, ICRP Publication 93, 2004.
  9. ICRP, ICRP Publication 105, 2007.
  10. S. J. Shepard, J. Wang, M. Flynn, E. Gingold, L. Goldman, K. Krugh, D. L. Leong, E. Mah, K. Ogden, D. Peck, E. Samei and C. E. Willis, Report of AAPM Task Group 116, AAPM, 2009.
  11. R. G. Lane, D. M. Stevens, J. P. Gibbons, L. J. Verhey, K. R. Hogstrom, E. L. Chaney, M. C. Martin, E. E. Klein, K. P. Doppke, B. R. Paliwal. R. E. Wendt III, and M. G. Herman, Report of AAPM Task Group 90, AAPM, 2006.
  12. J. K. Dave, A. K. Jones, R. Fisher, K. Hulme, L. Rill, D. Zamora, A. Woodward, S. Brady, R. D. MacDougall, L. Goldman, S. Lang, D. Peck, B. Apgar, S. J. Shepard, R. Uzenoff, and C. Willis, "Current State of Practice Regarding Digital Radiography Exposure Indicators and Deviation Indices: Report of AAPM Imaging Physics Committee Task Group 232," Med. Phys., Vol.45, No.11, pp.e1146-e1160, 2018.
  13. R. K. Grewal, N. Young, L. Colins, N. Karunnaratne, and N. Sabharwal, "Digital Chest Radiography Image Quality Assessment with Dose Reduction," Australasian Physical & Engineering Sciences in Medicine, Vol.35, No.1, pp.71-80, 2012. https://doi.org/10.1007/s13246-012-0125-5
  14. G. Dougherty, Digital Image Processing for Medical Applications, Cambridge, 2009.
  15. G. Dougherty, Pattern Recognition and Classification, Springer Science & Business Media, 2012.
  16. 최지안, 황준호, 이경배, "인터벤션에서 자동노출제어 장치 이용 시 내장형 면적 선량계와 이동형 면적 선량계의 면적선량 비교를 통한 선량계 오차분석과 교정지침 필요성 연구," 한국콘텐츠학회논문지, 제18권, 제3호, pp.508-515, 2018.
  17. IEC, IEC 61267 : Medical Diagnostic X-ray Equipment - Radiation Conditions for Use in the Determination of Characteristics, 2005.
  18. IEC, IEC 62220-1 : Medical Electrical Equipment - Characteristics of Digital X-ray Imaging Devices - Part 1: Determinations of the Detective Quantum Efficiency, 2003.
  19. IEC, IEC 62494-1 : Medical Electrical Equipment - Exposure Index of Digital X-ray Imaging Systems - Part 1: Definitions and Requirements for General Radiography, 2008.
  20. ISO, ISO 9236-1 : Photography - Sensitometry of Screen/Film Systems for Medical Radiography - Part 1: Determination of Sensitometric Curve Shape, Speed and Average Gradient, 2004.
  21. E. B. Podgorsak, Radiation Physics for Medical Physicists, Springer, 2016.
  22. 식품의약품안전처, 영상의학 검사(일반촬영)에서의 표준 촬영기법, 2014.
  23. 김상현, "디지털 흉부단층합성검사에서 감도와 관전압 변화에 따른 영상 최적화," 한국방사선학회 논문지, 제12권, 제4호, pp.541-547, 2018.
  24. J. T. Smith, A. J. Couture, J. R. Stowell, and D. R. Allee, "Optically Seamless Flexible Electronic Tiles for Ultra Large-Area Digital X-ray Imaging," IEEE Transaction on Components, Packaging, and Manufacturing Technology, Vol.4, No.6, pp.1109-1115, 2014. https://doi.org/10.1109/TCPMT.2014.2316477
  25. 황준호, 이경배, "DR (Digital Radiography)에서 자동노출제어장치의 감도변화에 따른 검출기 성능평가시 흡수선량과 히스토그램을 통한 정량적 분석방법에 관한 연구," 한국콘텐츠학회논문지, 제18권, 제1호, pp.232-240, 2018. https://doi.org/10.5392/JKCA.2018.18.01.232
  26. https://www.iba-dosimetry.com, 2019.06.03.
  27. 보건복지부, 진단용 방사선 발생장치의 안전관리에 관한 규칙, 2015.