A retrospective randomized study of success rates according to abutment selection in DENTIS submerged implant with an internal hex connection

DENTIS 내부연결형 서브머지드 임플란트에서 지대주 선택에 따른 성공률의 후향적 연구

  • Kim, Eun-Hee (Department of Prosthodontics, Dong-a University) ;
  • Lee, Joeng-Eun (Department of Prosthodontics, Dong-a University) ;
  • Hwang, Hee-Seong (Department of Prosthodontics, Dong-a University) ;
  • Kim, Chul-Hoon (Department of Oral and Maxillofacial Surgery, Implant Clinics, Dong-a University) ;
  • Kim, Jung-Han (Department of Oral and Maxillofacial Surgery, Implant Clinics, Dong-a University) ;
  • Kim, Bok-Joo (Department of Oral and Maxillofacial Surgery, Implant Clinics, Dong-a University)
  • 김은희 (동아대학교의료원 치과보철과) ;
  • 이정은 (동아대학교의료원 치과보철과) ;
  • 황희성 (동아대학교의료원 치과보철과) ;
  • 김철훈 (동아대학교의료원 구강악안면외과) ;
  • 김정한 (동아대학교의료원 구강악안면외과) ;
  • 김복주 (동아대학교의료원 구강악안면외과)
  • Received : 2018.05.10
  • Accepted : 2018.08.28
  • Published : 2018.10.31

Abstract

PURPOSE. The purpose of this study is to determine the efficacy of the DENTIS submerged-type implant with an internal hex connection and to build corresponding abutment-selection criteria. MATERIAL AND METHODS. A total of 204 patients received submerged implant fixtures with an internal hex connection at the Dong-A University Hospital Dental clinic in Busan from January 2013 and May 2016. Three specific abutments, UCLA abutments, customized abutments, ready-made abutments, were randomly selected. Implant success was defined as the basis of the International Congress of Oral Implantologists(ICOI, 2007) criteria. The relationship between the implant success rate and the abutment factor was analyzed using the Kruskal-Wallis test(P<.05). RESULTS. A total of 508 implants were placed in 204 patients. After a mean observation period of 38.6 months, 493 out of 508 implants were in normal function, yielding an overall success rate of 97.05%. A total of 15 implants failed: 10 in the maxillary molar area, 4 in the mandibular molar area, and 1 in the mandibular incisal area. All of the implant failures occurred in a single-implant prosthesis, especially high in the maxillary molar area. The Kruskal-Wallis analysis showed that abutment selection has no significant correlation with implant failure(P>.05). CONCLUSION. DENTIS submerged implants with an internal hex connection showed predictable results with a success rate of 97.05%. It is no influence on the success rate in the selection of submerged implant abutment with an internal hex connection.

Keywords

References

  1. Adell R, Lekholm U, Rockler B, Branemark PI. A 15-year study of osseointegrated implants in the treatment of the edentulous jaw. Int. J Oral Surg 1981;10:387-416. https://doi.org/10.1016/S0300-9785(81)80077-4
  2. Maeda Y, Satoh T, Sogo M. In vitro differences of stress concentrations for internal and external hex implant-abutment connections: a short communication. J Oral Rehab 2006;33:75-78. https://doi.org/10.1111/j.1365-2842.2006.01545.x
  3. Oh TJ, Yoon J, Misch CE, Wang HL. The cause of early implant bone loss : myth or science? J Periodontol 2002;73:322-333. https://doi.org/10.1902/jop.2002.73.3.322
  4. Stephen T, Chen Ivan D, Darby Eric C, Reynolds. A prospective clinical study of non-submerged immediate implants:clinical outcomes and esthetic results. Clin. Oral Impl. Res 2007;18:552-562. https://doi.org/10.1111/j.1600-0501.2007.01388.x
  5. Lewis SG, Beumer J, Perri GR, Hornburg WP. Single tooth implant supported restorations. Int J Oral Maxillofac Implants 1988;3:25-30
  6. Lewis SG, Beumer J, Perri GR, Hornburg WP. The "UCLA" abutment. Int J Oral Maxillofac Implants 1988;3:183-189.
  7. Lewis SG, Llamas D, Avera S. The UCLA abutment: a four-year review. J Prosthet Dent 1992;67:509-515. https://doi.org/10.1016/0022-3913(92)90082-L
  8. Misch CE, Perel ML, Wang HL, Sammartino G, Galindo-Moreno P, Trisi P, Steigmann M, Rebaudi A, Palti A, Pikos MA, Schwartz-Arad D, Choukroun J, Perez JL, Marenzi G, Valavanis DK. Implant success, survival, and failure: The International Congress of Oral Implantologists(ICOI) Pisa Consensus Conference. Implant Dentistry 2008;17:5-11. https://doi.org/10.1097/ID.0b013e3181676059
  9. Becker J, Ferrari D, Herten M, Kirsch A, Schaer A, Schwarz F. Influence of platform switching on crestal bone changes at non-submerged titanium implants: a histomorphometrical study in dogs. J Clin Periodontol 2007; 34:1089-1096. https://doi.org/10.1111/j.1600-051X.2007.01155.x
  10. Balfour A, O'Brien GR. Comparative study of antirotational single tooth abutments. J Prosthet Dent 1995;73:36-43. https://doi.org/10.1016/S0022-3913(05)80270-7
  11. Chang HS, Chen YC, Hsieh YD, Hsu ML. Stress distribution of two commercial dental implant systems: A three-dimensional finite element analysis. J Dent Sci 2013;8: 261-271. https://doi.org/10.1016/j.jds.2012.04.006
  12. Khraisat A, Stegaroiu R, Nomura S, Miyakawa O. Fatigue resistance of two implant/abutment joint design. J Prosthet Dent 2002;88:604-10. https://doi.org/10.1067/mpr.2002.129384
  13. Steinebrunner L, Wolfart S, Ludwig K, Kern M. Implant-abutment interface design affects fatigue and fracture strength of implants. Clin Oral Impl Res 2008;29:1276-1284.
  14. Adell R, Eriksson B, Lekholm U, Branemark PI, Jemt T. A long-term follow-up study of osseointegrated implants in the treatment of totally edentulous jaws. Int J Oral Maxillofac Implants 1990;5:347-359.
  15. Buser D, Mericske-Stern R, Bernard JP, Behneke N, Hirt HP, Belser UC, Lang NP. Long-term evaluation of non-submerged ITI implants. Clin Oral Impl Res 1997;8:161-172. https://doi.org/10.1034/j.1600-0501.1997.080302.x
  16. Isidor F. Loss of osseointegration caused by occlusal load of oral implants: A clinical and radiographic study in monkeys. Clin Oral Impl Res 1996;7:143-152. https://doi.org/10.1034/j.1600-0501.1996.070208.x
  17. Rangert B, Krogh PHJ, Langer B, Roekel NV. Bending overload and implant fracture: A retrospective clinical analysis. Int J Oral Maxillofac Implants 1995;10:326-334.
  18. Goodacre CJ, Kan JYK, Rungcharassaeng K. Clinical complications of osseointegrated implants. J Prosthet Dent 1999;81:537-552. https://doi.org/10.1016/S0022-3913(99)70208-8
  19. Alsaadi G, Quirynen M, Koma′rek A, van Steenberghe D. Impact of local and systemic factors on the incidence of oral implant failures, up to abutment connection. J Clin Periodontol 2007;34:610-7. https://doi.org/10.1111/j.1600-051X.2007.01077.x
  20. Moy PK, Medina D, Shetty V, Aghaloo TL. Dental implant failure rates and associated risk factors. Int J Oral Maxillofac Implants 2005;20:569-77.
  21. Chrcanovic BR, Kisch J, Albrektsson T, Wennerberg A. Factors influencing early dental implant failures. J Dent Res 2016;95:995-1002. https://doi.org/10.1177/0022034516646098
  22. Chung CH, Son MK. The classification and comparison of implant prosthesis according to types of retention. Part I: Screw retained prosthesis vs cement retained prosthesis. J Korean Acad Oral Maxillofac Implantology 2010;14:138-151.
  23. Hebel KS, Gajjar RC. Cement-retained versus screw-retained implant restoration: Achieving optimal occlusion and esthetics in implant dentistry. J Prosthet Dent 1997;77:28-35. https://doi.org/10.1016/S0022-3913(97)70203-8
  24. Kallus T, Bessing C. Loose gold screws frequently occur in full arch fixed prostheses supported by osseointegrated implants after 5 years. Int J Oral Maxillofac Implants 1994;9:169-178.
  25. Carlson B, Carlsson GE. Prosthodontic complications in osseointegrated dental implant treatment. Int J Oral Maxillofac Implants 1994;9:90-94.
  26. Jemt T, Linden B, Lekholm U. Failures and complications in 127 consecutively placed fixed partial prostheses supported by Branemark implants: From prosthetic treatment to first annual checkup. Int J Oral Maxillofac Implants 1992;7:40-44.
  27. Park SD, Lee Y, Kim YL, Yu SH, Bae JM, Cho HW. Microleakage of different sealing materials in access holes of international connection implant systems. J Prosthet Dent 2012;108:173-180. https://doi.org/10.1016/S0022-3913(12)60143-7
  28. Byrne D, Houston F, Cleary R, Claffey N. The fit of cast and premachined implant abutments. J Prosthet Dent 1998;80:184-192. https://doi.org/10.1016/S0022-3913(98)70108-8
  29. Weng D, Jacobson Z, Tarnow D, Hurzeler MB, Faehn O, Sanavi F, Barkvoll P, Stach RM. A prospective multicenter clinical trial of 3i machinedsurface implants: Results after 6 years of followup. Int J Oral Maxillofac Implants 2003;18:417-423.
  30. Herrmann I, Lekholm U, Holm S, Kultje C. Evaluation of patient and implant characteristics as potential prognostic factors for oral implant failure. Int J Oral Maxillofac Implants 2005;20:220-230.
  31. Henriksson K, Jemt T. Evaluation of custom-made procera ceramic abutments for single-implant tooth replacement: a prospective 1-year follow-up study. Int J Prosthodont 2003;16:626-30.
  32. Kim HS. Fabrication of custom abutment using dental CAD/CAM system. J Korean Dent Assoc 2012;50:118-25.
  33. Wu T, Liao W, Dai N, Tang C. Design of a custom angled abutment for dental implants using computer-aided design and nonlinear finite element analysis. J Biomech 2010;43:1941-6. https://doi.org/10.1016/j.jbiomech.2010.03.017
  34. Sailer I, Zembic A, Jung RE, Siegenthaler D, Holderegger C, Hammerle CH. Randomized controlled clinical trial of customized zirconia and titanium implant abutments for canine and posterior single-tooth implant reconstructions: preliminary results at 1 year of function. Clin Oral Implants Res 2009;20:219-25. https://doi.org/10.1111/j.1600-0501.2008.01636.x
  35. Baldassarri M, Hjerppe J, Romeo D, Fickl S, Thompson VP, Stappert CF. Marginal accuracy of three implant-ceramic abutment configurations. Int J Oral Maxillofac Implants 2012;27:537-43.
  36. Lee CJ, Yang SE, Kim SG. Evaluation of reverse torque value of abutment screws on CAD/CAM custom-made implant abutments. J Korean Acad Prosthodont 2012;50:128- https://doi.org/10.4047/jkap.2012.50.2.128