Patient-Specific Quality Assurance in a Multileaf Collimator-Based CyberKnife System Using the Planar Ion Chamber Array

  • Yoon, Jeongmin (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Lee, Eungman (Department of Radiation Oncology, Yonsei University College of Medicine) ;
  • Park, Kwangwoo (Department of Radiation Oncology, Yonsei University College of Medicine) ;
  • Kim, Jin Sung (Department of Radiation Oncology, Yonsei University College of Medicine) ;
  • Kim, Yong Bae (Department of Radiation Oncology, Yonsei University College of Medicine) ;
  • Lee, Ho (Department of Radiation Oncology, Yonsei University College of Medicine)
  • 투고 : 2018.05.24
  • 심사 : 2018.06.23
  • 발행 : 2018.06.30

초록

This paper describes the clinical use of the dose verification of multileaf collimator (MLC)-based CyberKnife plans by combining the Octavius 1000SRS detector and water-equivalent RW3 slab phantom. The slab phantom consists of 14 plates, each with a thickness of 10 mm. One plate was modified to support tracking by inserting 14 custom-made fiducials on surface holes positioned at the outer region of $10{\times}10cm^2$. The fiducial-inserted plate was placed on the 1000SRS detector and three plates were additionally stacked up to build the reference depth. Below the detector, 10 plates were placed to avoid longer delivery times caused by proximity detection program alerts. The cross-calibration factor prior to phantom delivery was obtained by performing with 200 monitor units (MU) on the field size of $95{\times}92.5mm^2$. After irradiation, the measured dose distribution of the coronal plane was compared with the dose distribution calculated by the MultiPlan treatment planning system. The results were assessed by comparing the absolute dose at the center point of 1000SRS and the 3-D Gamma (${\gamma}$) index using 220 patient-specific quality assurance (QA). The discrepancy between measured and calculated doses at the center point of 1000SRS detector ranged from -3.9% to 8.2%. In the dosimetric comparison using 3-D ${\gamma}$-function (3%/3 mm criteria), the mean passing rates with ${\gamma}$-parameter ${\leq}1$ were $97.4%{\pm}2.4%$. The combination of the 1000SRS detector and RW3 slab phantom can be utilized for dosimetry validation of patient-specific QA in the CyberKnife MLC system, which made it possible to measure absolute dose distributions regardless of tracking mode.

키워드

참고문헌

  1. Blomgren H, Lax I, Naslund I, Svanstrom R. Stereotactic high dose fraction radiation therapy of extracranial tumors using an accelerator. Clinical experience of the first thirty-one patients. Acta Oncol. 1995;34(6):861-70. https://doi.org/10.3109/02841869509127197
  2. Yoon J, Park K, Kim JS, Kim YB, Lee H. Acceptance Testing and Commissioning of Robotic Intensity-Modulated Radiation Therapy M6 System Equipped with $InCise^{TM}$ 2 Multileaf Collimator. Prog Med Phys. 2018;29(1):8-15. https://doi.org/10.14316/pmp.2018.29.1.8
  3. Murai T, Hattori Y, Sugie C, Iwata H, Iwabuchi M, Shibamoto Y. Comparison of multileaf collimator and conventional circular collimator systems in Cyberknife stereotactic radiotherapy. J Radiat Res. 2017;58(5):693-700. https://doi.org/10.1093/jrr/rrw130
  4. Mcguinness CM, Gottschalk AR, Lessard E, et al. Investigating the clinical advantages of a robotic linac equipped with a multileaf collimator in the treatment of brain and prostate cancer patients. J Appl Clin Med Phys. 2015;16(5):284-95. https://doi.org/10.1120/jacmp.v16i5.5502
  5. Dieterich S, Cavedon C, Chuang CF, et al. Report of AAPM TG 135: quality assurance for robotic radiosurgery. Med Phys. 2011;38(6):2914-36. https://doi.org/10.1118/1.3579139
  6. Chong NS, Lee JJS, Kung WH, et al. Patient delivery quality assurance for linac-based IMRT and helical tomotherapy using solid state detectors. Radiat Meas. 2011;46(12):1993-95. https://doi.org/10.1016/j.radmeas.2011.08.004
  7. Markovic M, Stathakis S, Mavroidis P, Jurkovic IA, Papanikolaou N. Characterization of a two-dimensional liquidfilled ion chamber detector array used for verification of the treatments in radiotherapy. Med Phys. 2014;41(5).
  8. Seppenwoolde Y, Berbeco RI, Nishioka S, Shirato H, Heijmen B. Accuracy of tumor motion compensation algorithm from a robotic respiratory tracking system: A simulation study. Med Phys. 2007;34(7):2774-84. https://doi.org/10.1118/1.2739811
  9. Pepin EW, Wu H, Zhang Y, Lord B. Correlation and prediction uncertainties in the cyberknife synchrony respiratory tracking system. Med Phys. 2011;38(7):4036-44. https://doi.org/10.1118/1.3596527
  10. Ezzell GA, Burmeister JW, Dogan N, et al. IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119. Med Phys. 2009;36(11):5359-73. https://doi.org/10.1118/1.3238104
  11. Blanck O, Masi L, Damme MC, et al. Film-based delivery quality assurance for robotic radiosurgery: Commissioning and validation. Phys Med. 2015;31(5):476-83. https://doi.org/10.1016/j.ejmp.2015.05.001
  12. Lin MH, Veltchev I, Koren S, Ma C, Li J. Robotic radiosurgery system patient-specific QA for extracranial treatments using the planar ion chamber array and the cylindrical diode array. J Appl Clin Med Phys. 2015;16(4):290-305. https://doi.org/10.1120/jacmp.v16i4.5486
  13. Kim J, Park K, Yoon J, et al. Feasibility Study of a Custommade Film for End-to-End Quality Assurance Test of Robotic Intensity Modulated Radiation Therapy System. Prog Med Phys. 2016;27(4):189-95. https://doi.org/10.14316/pmp.2016.27.4.189

피인용 문헌

  1. Skin Dose Comparison of CyberKnife and Helical Tomotherapy for Head-and-Neck Stereotactic Body Radiotherapy vol.30, pp.1, 2019, https://doi.org/10.14316/pmp.2019.30.1.1
  2. Development of an x-ray-opaque-marker system for quantitative phantom positioning in patient-specific quality assurance vol.91, pp.None, 2018, https://doi.org/10.1016/j.ejmp.2021.10.017