Set-up errors in head and neck cancer treated with IMRT technique assessed by cone-beam computed tomography: a feasible protocol

  • Received : 2017.10.18
  • Accepted : 2017.12.18
  • Published : 2018.03.31

Abstract

Purpose: To investigate set-up errors, suggest the adequate planning target volume (PTV) margin and image-guided radiotherapy frequency in head and neck (H&N) cancer treated with intensity-modulated radiotherapy (IMRT) assessed by kV cone-beam computed tomography (CBCT). Methods: We analyzed 360 CBCTs in 60 patients with H&N cancer treated with IMRT. The target delineation was contoured according to ICRU62. PTVs were generated by adding a 3-5 mm margin in all directions to the respective clinical target volumes. The kV CBCT images were obtained at first three days of irradiation and weekly thereafter. The overall mean displacement, range, systematic (${\Sigma}$) and random (${\sigma}$) errors were calculated. Adequate PTV margins were calculated according to the van Herk formula ($2.5{\Sigma}+0.7r$). Results: The mean of set-up errors was less than 2 mm in any direction. The overall frequency of set-up displacements greater than 3 mm was 3.9% in medial-lateral (ML) direction, 8% in superior-inferior (SI) direction, and 15.5% in anterior-posterior (AP) direction. The range of translations shifts was 0-9 mm in ML direction, 0-5 mm in SI direction and 0-10 mm in AP direction, respectively. After systematic set-up errors correction, the adequate margin to overcome the problem of set-up errors was found to be less than 3 mm. Conclusion: Image-guided kV CBCT was effective for the evaluation of set-up accuracy in H&N cancer. The kV CBCT at first three fractions and followed-by weekly appears adequate for reducing significantly set-up errors in H&N cancer treated with IMRT technique. Finally, 3-5 mm PTV margins appear adequate and safe to overcome the problem of set-up errors.

Keywords

References

  1. Siegel RL, Miller KD, Jemal A. Cancer statistics, 2015. CA Cancer J Clin 2015;65:5-29. https://doi.org/10.3322/caac.21254
  2. Marur S, Forastiere AA. Head and neck cancer: changing epidemiology, diagnosis, and treatment. Mayo Clin Proc 2008;83:489-501. https://doi.org/10.4065/83.4.489
  3. Stambuk HE, Karimi S, Lee N, Patel SG. Oral cavity and oropharynx tumors. Radiol Clin North Am 2007;45:1-20. https://doi.org/10.1016/j.rcl.2006.10.010
  4. National Comprehensive Cancer Network. NCCN Clinical Practice Guidelines in Oncology: head and neck cancer version II.2017 [Internet]. Fort Washington, PA: National Comprehensive Cancer Network; c2018 [cited 2018 Mar 1]. Available from: https://www.nccn.org/professionals/physician_gls/default.aspx.
  5. Merlotti A, Alterio D, Vigna-Taglianti R, et al. Technical guidelines for head and neck cancer IMRT on behalf of the Italian association of radiation oncology: head and neck working group. Radiat Oncol 2014;9:264. https://doi.org/10.1186/s13014-014-0264-9
  6. Pasqualetti F, Bocci G, Mey V, et al. Akt1 rs2498801 is related to survival in head and neck squamous cell cancer treated with radiotherapy. Anticancer Res 2015;35:269-71.
  7. Bucci MK, Bevan A, Roach M 3rd. Advances in radiation therapy: conventional to 3D, to IMRT, to 4D, and beyond. CA Cancer J Clin 2005;55:117-34. https://doi.org/10.3322/canjclin.55.2.117
  8. Forastiere AA, Zhang Q, Weber RS, et al. Long-term results of RTOG 91-11: a comparison of three nonsurgical treatment strategies to preserve the larynx in patients with locally advanced larynx cancer. J Clin Oncol 2013;31:845-52. https://doi.org/10.1200/JCO.2012.43.6097
  9. Lee N, Xia P, Fischbein NJ, Akazawa P, Akazawa C, Quivey JM. Intensity-modulated radiation therapy for head-andneck cancer: the UCSF experience focusing on target volume delineation. Int J Radiat Oncol Biol Phys 2003;57:49-60. https://doi.org/10.1016/S0360-3016(03)00405-X
  10. Ursino S, Faggioni L, Guidoccio F, et al. Role of perfusion CT in the evaluation of functional primary tumour response after radiochemotherapy in head and neck cancer: preliminary findings. Br J Radiol 2016;89:20151070. https://doi.org/10.1259/bjr.20151070
  11. Chao KS, Ozyigit G, Tran BN, Cengiz M, Dempsey JF, Low DA. Patterns of failure in patients receiving definitive and postoperative IMRT for head-and-neck cancer. Int J Radiat Oncol Biol Phys 2003;55:312-21. https://doi.org/10.1016/S0360-3016(02)03940-8
  12. Ursino S, D'Angelo E, Mazzola R, et al. A comparison of swallowing dysfunction after three-dimensional conformal and intensity-modulated radiotherapy: a systematic review by the Italian Head and Neck Radiotherapy Study Group. Strahlenther Onkol 2017;193:877-89. https://doi.org/10.1007/s00066-017-1160-7
  13. Eisbruch A, Kim HM, Terrell JE, Marsh LH, Dawson LA, Ship JA. Xerostomia and its predictors following parotid-sparing irradiation of head-and-neck cancer. Int J Radiat Oncol Biol Phys 2001;50:695-704. https://doi.org/10.1016/S0360-3016(01)01512-7
  14. Chao KS, Deasy JO, Markman J, et al. A prospective study of salivary function sparing in patients with head-and-neck cancers receiving intensity-modulated or three-dimensional radiation therapy: initial results. Int J Radiat Oncol Biol Phys 2001;49:907-16. https://doi.org/10.1016/S0360-3016(00)01441-3
  15. Delishaj D, Ursino S, Lombardo E, et al. Impact of treatment volumes in loco-regional failure of oral cancer in patients treated with IMRT. Radiother Oncol 2016;119:S985.
  16. Parliament MB, Scrimger RA, Anderson SG, et al. Preservation of oral health-related quality of life and salivary flow rates after inverse-planned intensity- modulated radiotherapy (IMRT) for head-and-neck cancer. Int J Radiat Oncol Biol Phys 2004;58:663-73. https://doi.org/10.1016/S0360-3016(03)01571-2
  17. Jaffray DA, Siewerdsen JH, Wong JW, Martinez AA. Flat-panel cone-beam computed tomography for image-guided radiation therapy. Int J Radiat Oncol Biol Phys 2002;53:1337-49. https://doi.org/10.1016/S0360-3016(02)02884-5
  18. Nielsen M, Bertelsen A, Westberg J, Jensen HR, Brink C. Cone beam CT evaluation of patient set-up accuracy as a QA tool. Acta Oncol 2009;48:271-6. https://doi.org/10.1080/02841860802245155
  19. Den RB, Doemer A, Kubicek G, et al. Daily image guidance with cone-beam computed tomography for head-and-neck cancer intensity-modulated radiotherapy: a prospective study. Int J Radiat Oncol Biol Phys 2010;76:1353-9. https://doi.org/10.1016/j.ijrobp.2009.03.059
  20. Korreman S, Rasch C, McNair H, et al. The European Society of Therapeutic Radiology and Oncology-European Institute of Radiotherapy (ESTRO-EIR) report on 3D CT-based in-room image guidance systems: a practical and technical review and guide. Radiother Oncol 2010;94:129-44. https://doi.org/10.1016/j.radonc.2010.01.004
  21. Stroom JC, Heijmen BJ. Geometrical uncertainties, radiotherapy planning margins, and the ICRU-62 report. Radiother Oncol 2002;64:75-83. https://doi.org/10.1016/S0167-8140(02)00140-8
  22. van Herk M. Errors and margins in radiotherapy. Semin Radiat Oncol 2004;14:52-64. https://doi.org/10.1053/j.semradonc.2003.10.003
  23. Amer AM, Mackay RI, Roberts SA, Hendry JH, Williams PC. The required number of treatment imaging days for an effective off-line correction of systematic errors in conformal radiotherapy of prostate cancer: a radiobiological analysis. Radiother Oncol 2001;61:143-50. https://doi.org/10.1016/S0167-8140(01)00440-6
  24. Delishaj D, Ursino S, Lombardo E, et al. OC-0274: Analysis of set-up errors in head and neck cancer treated with IMRT technique assessed by CBCT. Radiother Oncol 2016;119(Suppl 1):S127-S128.
  25. Bortfeld T, van Herk M, Jiang SB. When should systematic patient positioning errors in radiotherapy be corrected? Phys Med Biol 2002;47:N297-302. https://doi.org/10.1088/0031-9155/47/23/401
  26. Zeidan OA, Langen KM, Meeks SL, et al. Evaluation of imageguidance protocols in the treatment of head and neck cancers. Int J Radiat Oncol Biol Phys 2007;67:670-7. https://doi.org/10.1016/j.ijrobp.2006.09.040
  27. Suzuki M, Nishimura Y, Nakamatsu K, et al. Analysis of interfractional set-up errors and intrafractional organ motions during IMRT for head and neck tumors to define an appropriate planning target volume (PTV)- and planning organs at risk volume (PRV)-margins. Radiother Oncol 2006;78:283-90. https://doi.org/10.1016/j.radonc.2006.03.006
  28. Su J, Chen W, Yang H, et al. Different setup errors assessed by weekly cone-beam computed tomography on different registration in nasopharyngeal carcinoma treated with intensity-modulated radiation therapy. Onco Targets Ther 2015;8:2545-53.
  29. Houghton F, Benson RJ, Tudor GS, et al. An assessment of action levels in imaging strategies in head and neck cancer using TomoTherapy. Are our margins adequate in the absence of image guidance? Clin Oncol (R Coll Radiol) 2009;21:720-7. https://doi.org/10.1016/j.clon.2009.08.005
  30. Dionisi F, Palazzi MF, Bracco F, et al. Set-up errors and planning target volume margins in head and neck cancer radiotherapy: a clinical study of image guidance with on-line cone-beam computed tomography. Int J Clin Oncol 2013;18:418-27. https://doi.org/10.1007/s10147-012-0395-7
  31. Bissonnette JP, Moseley D, White E, Sharpe M, Purdie T, Jaffray DA. Quality assurance for the geometric accuracy of conebeam CT guidance in radiation therapy. Int J Radiat Oncol Biol Phys 2008;71(1 Suppl):S57-61. https://doi.org/10.1016/j.ijrobp.2007.06.086
  32. McKenzie A, van Herk M, Mijnheer B. Margins for geometric uncertainty around organs at risk in radiotherapy. Radiother Oncol 2002;62:299-307. https://doi.org/10.1016/S0167-8140(02)00015-4
  33. Pisani L, Lockman D, Jaffray D, Yan D, Martinez A, Wong J. Setup error in radiotherapy: on-line correction using electronic kilovoltage and megavoltage radiographs. Int J Radiat Oncol Biol Phys 2000;47:825-39. https://doi.org/10.1016/S0360-3016(00)00476-4
  34. Wang J, Bai S, Chen N, et al. The clinical feasibility and effect of online cone beam computer tomography-guided intensitymodulated radiotherapy for nasopharyngeal cancer. Radiother Oncol 2009;90:221-7. https://doi.org/10.1016/j.radonc.2008.08.017
  35. Johansen J, Bertelsen A, Hansen CR, Westberg J, Hansen O, Brink C. Set-up errors in patients undergoing image guided radiation treatment. Relationship to body mass index and weight loss. Acta Oncol 2008;47:1454-8. https://doi.org/10.1080/02841860802256517
  36. Barker JL Jr, Garden AS, Ang KK, et al. Quantification of volumetric and geometric changes occurring during fractionated radiotherapy for head-and-neck cancer using an integrated CT/linear accelerator system. Int J Radiat Oncol Biol Phys 2004;59:960-70. https://doi.org/10.1016/j.ijrobp.2003.12.024
  37. Xu F, Wang J, Bai S, et al. Detection of intrafractional tumour position error in radiotherapy utilizing cone beam computed tomography. Radiother Oncol 2008;89:311-9. https://doi.org/10.1016/j.radonc.2008.08.003
  38. Velec M, Waldron JN, O'Sullivan B, et al. Cone-beam CT assessment of interfraction and intrafraction setup error of two head-and-neck cancer thermoplastic masks. Int J Radiat Oncol Biol Phys 2010;76:949-55. https://doi.org/10.1016/j.ijrobp.2009.07.004
  39. Zhang L, Garden AS, Lo J, et al. Multiple regions-of-interest analysis of setup uncertainties for head-and-neck cancer radiotherapy. Int J Radiat Oncol Biol Phys 2006;64:1559-69. https://doi.org/10.1016/j.ijrobp.2005.12.023
  40. Clarizio M, Zani M, Delishaj D, et al. EP-1619: Comparison between two different commercial thermoplastic mask systems in image-guided radiation therapy. Radiother Oncol 2015;115:S887-S888. https://doi.org/10.1016/S0167-8140(15)41611-1
  41. Osman SO, de Boer HC, Astreinidou E, Gangsaas A, Heijmen BJ, Levendag PC. On-line cone beam CT image guidance for vocal cord tumor targeting. Radiother Oncol 2009;93:8-13. https://doi.org/10.1016/j.radonc.2009.05.015
  42. van Kranen S, van Beek S, Rasch C, van Herk M, Sonke JJ. Setup uncertainties of anatomical sub-regions in head-andneck cancer patients after offline CBCT guidance. Int J Radiat Oncol Biol Phys 2009;73:1566-73. https://doi.org/10.1016/j.ijrobp.2008.11.035

Cited by

  1. Effect of Surgical Mask on Setup Error in Head and Neck Radiotherapy vol.19, pp.None, 2018, https://doi.org/10.1177/1533033820974021
  2. Evaluation of the effect of random setup errors on dose delivery in Intensity Modulated Radiotherapy vol.26, pp.1, 2018, https://doi.org/10.2478/pjmpe-2020-0006
  3. Impact of Cold Weather on Setup Errors in Radiotherapy vol.2021, pp.None, 2021, https://doi.org/10.1155/2021/1463299
  4. Automated algorithm for calculation of setup corrections and planning target volume margins for offline image‐guided radiotherapy protocols vol.22, pp.7, 2021, https://doi.org/10.1002/acm2.13291