On the Use Factor Analysis and Adequacy Evaluation of CyberKnife Shielding Design Using Clinical Data

  • Cho, Yu Ra (Department of CyberKnife Center, Korea Institute of Radiological & Medical Sciences) ;
  • Jung, Haijo (Radiation Safety Office, Korea Institute of Radiological & Medical Sciences) ;
  • Lee, Dong Han (Department of CyberKnife Center, Korea Institute of Radiological & Medical Sciences)
  • Received : 2018.10.26
  • Accepted : 2018.12.18
  • Published : 2018.12.31

Abstract

Although the current internationally recommended standard for the use factor (U) applied to CyberKnife is 0.05 (5%), the CyberKnife shielding standard is applied more stringently. This study, based on clinical data, was aimed at examining the appropriateness of existing shielding guidelines. Sixty patients treated with G4 CyberKnife were selected. The patients were divided into two groups, according to whether they underwent skull or spine tracking. Based on the results, the use factors for each wall ranged from 0.028 (2.8%) to 0.031 (3.1%) for the intracranial treatment and 0.020 (2.0%) to 0.022 (2.2%) for the body treatment. Excessive barrier thickness resulted in inefficient use of space and higher cost to the institutions. Furthermore, because the use factor is influenced by the position of the robot, the use factor determined based on the clinical data of this study would facilitate more reasonable treatment room design.

Keywords

References

  1. NCRP Report 44. Radiation Protection for Particle Accelerator Facilities. Bethesda 2003;78-82.
  2. NCRP Report 49. Structural shielding design and evaluation for medical use of x-rays and gamma rays of energies up to 10 MeV ; 1976.
  3. McGinley PH. Shielding techniques for radiation oncology facilities, 2nd edition. Madison : Medical Physics Publishing, 2002.
  4. NCRP Report 151. In Structural shielding design and evaluation for megavoltage x-and gamma-ray radiotherapy facilities ; 2010.
  5. Henzen D, Schmidhalter D, Zanella CC, et al. Evaluation of clinically applied treatment beams with respect to bunker shielding parameters for a CyberKnife M6. J. Appl. Clin. Med. Phys. 2018;19(1):243-249.
  6. Purwar AK, Lee MY, Rogers JE. Tenth-value layer measurements of leakage radiation and secondary barrier shielding calculations for the CyberKnife robotic radiosurgery system. Accuray White paper. Sunnyvale, CA, Accuray ; 2009.
  7. Rodgers JE. CyberKnife treatment room design and radiation protection, Chapter 5. CyberKnife Society Press ; 2005.
  8. Rodgers JE. TVLs of common shielding materials for a robotically mounted stereotactic radiosurgery machine. J. Applied. Clin. Med. Phys. 2005;157-164.
  9. Yang J, Wu W, Hasson B. An evaluation of the use factor for instantaneous dos rates for CyberKnife shielding design. AAPM Annual Meeting, Scottsdale, AZ, 2007; 27-30.
  10. Yang J, Feng J. Radiation shielding evaluation based on five years of data from a busy CyberKnife center. J. Appl. Clin. Med. Phys. 2014;15:313-322. https://doi.org/10.1120/jacmp.v15i6.4575
  11. Echner GG, Kilby W, Lee M. The design, physical properties and clinical utility of an iris collimator for robotic radiosurgery. Phys. Med. Biol. 2009;54(18):5359-5380. https://doi.org/10.1088/0031-9155/54/18/001
  12. Rogers JE. Radiation therapy vault shielding calculation methods when IMRT and TBI procedure contribute. J. Appl. Clin. Med. Phys. 2002;2:57-64.
  13. Diebel J. Representing Attitude : Euler Angles, Unit Quaternions, and Rotation Vectors. Matrix. 2006;58:1-35.
  14. Eric W. Weisstein. Euler angles. From MathWorld-A Wolfram Web Resource. http://mathworld.wolfram.com/Cayley-KleinParameter.html ; 2009.
  15. Arend L. Quaternions, finite rotation and euler parameters ; 2002.
  16. Asmerom G, Bourne D, Chappelow J et al. The design and physical characterization of a multileaf collimator for robotic radiosurgery. Biomed. Physics. Eng. Express. 2016;2:17003. https://doi.org/10.1088/2057-1976/2/1/017003
  17. Sheridan ME, Martin M, Khalil S et al. Protecting a CT simulator room to accommodate a CyberKnife facility. Radiat. Prot. Dosimetry. 2015;165:468-471. https://doi.org/10.1093/rpd/ncv047