Abstract
This study examined the following: accuracy of the exposure conditions in the inverter device and three-phase device; output waveform over the exposure conditions; and average and standard deviation of the output waveform. After assessing whether the dose corresponding to the theoretical dose was presented, the following conclusions were obtained: 1. The accuracy of the tube voltage(kVp) and tube current(mA) exposure time(sec) was within the tolerable level prescribed in Korea's Safety Management Standards. In the error, Inverter device was large the tube voltage and exposure time, the three-phase device was large the tube current. 2. In terms of the output waveform of the exposure conditions and the average and standard deviation of the output waveform, the higher tube voltage and larger tube current resulted in greater standard deviation in pulsation. Moreover, the standard deviation of pulsation was shown to be greater in the inverter device than the three-phase device; there was also greater standard deviation in the inverter device considering the exposure time. 3. Regarding the exposure conditions over the output dose, all linearity showed the coefficient of variation which had an allowable limit of error within 0.05. Although the output dose ratio for the inverter device was 1.00~1.10 times no difference that of the three-phase device, there was almost no difference in dose ratio between the tube currents.