Fig. 1. Chest model phantom (LungMan, Kyoto Kagaku): (Left) Imaging of Radiography (Right) Photographs taken in this study.
Fig. 2. Pulsatile coronary artery phantom (Mocomo, Fuyo Corporation): (Left) Coronary angiography device and phantom, (Right) Coronary anastomosis.
Fig. 3. ROI setting for image quality evaluation in chest simulated phantom images.
Fig. 4. Setting the region of interest for image quality assessment in pulsed coronary phantom projection.
Fig. 5. Magnification mode and change in frame rate per second: (top) 7.5 f/s, (middle) 15 f/s (bottom) 30 f/s.
Table 1. DAP and reference air kerma according to varying frame rate and zooming modes of Cine acquisition
Table 2. DAP and reference air kerma according to varying frame rate and zooming modes of Cine acquisition of pulsatile coronary phantom
Table 3. Signal intensity, standard deviation, SNR, and CNR according to varying frame rate and zooming modes of Cine acquisition
Table 4. Signal intensity, standard deviation, SNR, and CNR according to varying frame rate and zooming modes of Cine acquisition
참고문헌
- S. C. Ma, H. G. Kim, B. S. Park, H. G. Goo, S. J You, S. I. Baek, Vascular and Interventional Radiology, Dai-Hak co., pp. 101-107, 2008.
- Pantos I, Patatoukas G, Katritsis DG, Efstathopoulos E, "Patient Radiation Doses in Interventional Cardiology Procedures," Current Cardiology Reviews, Vol. 5 No. 1, pp. 1-11, 2009. https://doi.org/10.2174/157340309787048059
- Picano E, Vano E, "The Radiation Issue in Cardiology: The Time for Aaction is Now," Cardiovascular Ultrasound, 2011.
- Wilde P, Pitcher EM, Slack K, "Radiation Hazards for the Patient in Cardiological Procedures," Heart, Vol. 85, No. 2, pp. 127-130, 2001. https://doi.org/10.1136/heart.85.2.127
- Neofotistou V, Vano E, Padovani R, "Preliminary reference levels in interventional cardiology," European Radiology, Vol. 13, No. 10, pp. 2259-2263, 2003. https://doi.org/10.1007/s00330-003-1831-x
- Pyne C, Gadey G, Jeon C, Piemonte T, Waxman S, and Resnic F, "Effect of Reduction of the Pulse Rates of Fluoroscopy and Cine-Acquisition on X-ray Dose and Angiographic Image Quality during invasive cardiovascular procedures," Circulation Cardiovascular Intervention, Vol. 7, No. 4, pp. 441-446, 2014. https://doi.org/10.1161/CIRCINTERVENTIONS.114.001479
- "National Council on Radiation Protection and Measurements, Implementation of the Principle of as low as reasonably achievable for medical and dental personnel," Bethesda:NCRP 1990; NCRP Report, No. 107, 1990.
- J. I. Choi, D. G. Na, H. H. Kim, Y. M. Shin, K. J. Ahn, J. Y. Lee, "Quality Control of Medical Imaging," Korean Journal of Radiology, Vol. 50, No. 5, pp. 317-331, 2004. https://doi.org/10.3348/jkrs.2004.50.5.317
- K. Y. Eun, K. J. Hyeon, K. J. Soo, K. C. Min, "Comparison of computed tomography (CT) dose and image quality - Dynamic CT and Multi Detector CT center," Journal of the Korean Society of Cardio-Vascular Interventional Technology, Vol. 17, No. 1, pp. 163-172, 2014.