Abstract
In the stripping scan square wave voltammetry (SV+SWV) polarography that is often used to analyze the concentration of heavy metals in water, we must measure the point where the faradic current that flows by the pure oxidation-reduction reaction at the electrode is greater than the capacitive current, the frequency cannot be too high. Therefore we wanted to find the frequency range that can be measured. In order to do this, we came up with a method to analyze the signal using FPGA Soc. With this method, the frequency of the square wave was increased from 10Hz to 400Hz by 10Hz, and the measuring time of the square wave was changed from 96.695% to 96.765% by 0.005% while 1600 experiments were conducted. As a result, the frequency of the square wave maintained a stable area of potential-current within 320Hz and it was possible to measure the potential-current signal when calculating the measuring time within the frequency range of 96.7155%.