DOI QR코드

DOI QR Code

Early Detection of Infiltration Induced in the Veins of Pig's Ear and Human's Forearm By Using Bioimpedance: Pilot Study

  • Kim, Jaehyung (Research Institute of Nursing Science, Pusan National University) ;
  • Hwang, Youngjun (Medical Science, School of Medicine, Pusan National University) ;
  • Kim, Gunho (Medical Science, School of Medicine, Pusan National University) ;
  • Jeong, Ihn Sook (College of Nursing, Pusan National University) ;
  • Jeon, Gyerok (Dept. of Biomedical Engineering, School of Medicine, Pusan National University)
  • 투고 : 2017.09.11
  • 심사 : 2017.12.22
  • 발행 : 2018.01.31

초록

An early detection of intravenous (IV) infiltration is essential to minimize the injuries during infusion therapy, which is one of the most important tasks for nurses in nursing settings. We report that bioelectrical impedance analysis is useful in the early detection of infiltration at puncture sites. When infiltration was intentionally induced in the vein of a pig's posterior ear, impedance parameters (R, $X_C$, $C_m$) showed significant differences before and after infiltration. In particular, the relative resistance ($R/R_{BI}$) decreased significantly at infiltration and then slowly decreased. This indicates that the vein in pig's ear is thin and the amount of surrounding subcutaneous tissue, and hence the infiltrated solution accumulates slowly after infiltration. However, when infiltration was induced in the vein of human's forearm, the relative resistance at 20 kHz decreased gradually over time. In the $R-X_C$ graph, the positions in the case of infiltration induced in the pig' ear shifted rapidly before and after infiltration, whereas the positions in the case of infiltration induced in the human's forearm moved gradually during infiltration. Our findings suggest that the impedance parameters (R, $R/R_{BI}$, $X_C$, R vs. $X_C$, and $C_m$) are effective indicators to detect the infiltration early in a non-invasive and quantitative manners.

키워드

참고문헌

  1. D. Doellman, L. Hadaway, L.A. Bowe-Geddes, M. Franklin, J. LeDonne, L. Papke-O'Donnell, et al., "Infiltration and Extravasation: Update on Prevention and Management," Journal of Infusion Nursing, Vol. 32, No. 4, pp. 203-211, 2009. https://doi.org/10.1097/NAN.0b013e3181aac042
  2. L. Hadaway, "Infiltration and Extravasation," American J ournal of Nursing, Vol. 107, No. 8, pp. 64-72, 2007.
  3. L. Dougherty, "IV Therapy: Recognizing the Differences Between Infiltration and Extravasation," British Journal of Nursing, Vol. 17, No. 14, pp. 896-901, 2008. https://doi.org/10.12968/bjon.2008.17.14.30656
  4. E. Kagel and G. Rayan, "Intravenous Catheter Complications in the Hand and Forearm," Journal of Trauma-Injury Infection and Critical Care, Vol. 56, No. 1, pp. 123-127, 2004. https://doi.org/10.1097/01.TA.0000058126.72962.74
  5. R. Simona, "A Pediatric Peripheral Intravenous Infiltration Assessment Tool," Journal of Infusion Nursing, Vol. 35, No. 4, pp. 243-248, 2012. https://doi.org/10.1097/NAN.0b013e31825af323
  6. C.M. Rickard, J. Webster, M.C. Wallis, N. Marsh, M.R. McGrail, V. French, et al., "Routine Versus Clinically Indicated Replacement of Peripheral Intravenous Catheters: A Randomised Controlled Equivalence Trial," British Journal of Nursing, Vol. 380, No. 9847, pp. 1066-1074, 2012.
  7. S.S. Dychter, D.A. Gold, D. Carson, and M. Haller, "Intravenous Therapy: A Review of Complications and Economic Considerations of Peripheral Access," Journal of Infusion Nursing, Vol. 35, No. 2, pp. 84-91, 2012. https://doi.org/10.1097/NAN.0b013e31824237ce
  8. Complications of Peripheral I.V. Therapy, Lippincott Nursing Center, http://www.nursingcenter.com/ncblog/february-2015-(1)/complications-of-peripheral-i-v-therapy (accessed Jan., 4, 2017).
  9. C. Lake, C.L. Beecroft, "Extravasation injuries and accidental intra-arterial injection," Critical & Pain, Vol. 10, No. 4, pp. 109-113, 2010.
  10. A. Jambulingam, R. McCrory, L. West, and O. T. Inan, "Non-Invasive, Multi-Modal Sensing of Skin Stretch and Bioimpedance for Detecting Infiltration During Intravenous Therapy," Proceeding of 38th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, pp. 4755-4758, 2016.
  11. L.W. Winchester and N.Y. Chou, Optical Detection of Intravenous Infiltration, http://www.google.com/patents/US7826890 (accessed Dec., 20, 2015).
  12. An Optical Device for Detecting Intravenous Infiltration, http://www.ivteam.com/optical-iv.pdf (accessed Dec., 21, 2016).
  13. Startup Launches Device for Early Detection of IV Tissue Infiltration to Reduce Complications, https://www.fiercebiotech.com/medical-devices/startup-launches-device-for-early-detection-iv-tissue-infiltration-to-reduce (accessed Sept., 7, 2016).
  14. Ultrasound Detection of Peripheral IV Infiltration, https://clinicaltrials.gov/ct2/show/NCT01800552 (accessed Jan., 6, 2017).
  15. U.G. Kyle, I. Bosaeus, A.D. De Lorenzo, P. Deurenberg, M. Elia, J. M. Gomez, et al., "Bio-Electrical Impedance Analysis-Part 1: Review of Principles and Methods," Clinical Nutrition, Vol. 23, No. 5, pp. 1226-1243, 2004. https://doi.org/10.1016/j.clnu.2004.06.004
  16. S.F. Kalil, M.S. Mohktar, and F. Ibrahim, "The Theory and Fundamentals of Bioimpedance Analysis in Clinical Status Monitoring and Diagnosis of Diseases," Sensors, Vol. 14, pp. 10895-10928, 2014. https://doi.org/10.3390/s140610895
  17. S. Berlit, J. Brade, B. Tuschy, E. Foldi, U. Walz-Eschenlohr, H. Leweling, et al., "Whole-Body Versus Segmental Bioelectrical Impedance Analysis in Patients with Edema of the Upper Limb After Breast Cancer Treatment," Anticancer Research, Vol. 33, No. 8, pp. 3403-3406, 2013.
  18. R. Buffa, E. Mereu, O. Comandini, M.E. Ibanez, and E. Marini, "Bioelectrical Impedance Vector Analysis (BIVA) for the Assessment of Two-Compartment Body Composition," European Journal Clinical Nutrition, Vol. 68, No. 11, pp. 1234-1240, 2014. https://doi.org/10.1038/ejcn.2014.170
  19. I.S. Grimnes and O.G. Martinsen, Bioimpedance and Bioelectricity Basics, Academic Press, London, 2015.
  20. S.O. Kasap, Principles of Electrical Engineering Materials and Devices, McGraw-Hill, New York, 1997.
  21. P.P. Urone, College Physics, Brooks/Cole, Pacific Grove, USA, 2001.
  22. J.A. Sargent and F.A. Gotch, Principles and Biophysics of Dialysis. In Replacement of Renal Function by Dialysis, Springer, Dordrecht, Netherland, 1989.
  23. E. Mazzaferro and L.L. Powell, "Fluid Therapy for the Emergent Small Animal Patient: Crystalloids, Colloids, and Albumin Products," Veterinary Clinics of North America: Small Animal Practice, Vol. 43, No. 4, pp. 721-734, 2013. https://doi.org/10.1016/j.cvsm.2013.03.003
  24. F. Zhu, E.F. Leonard, and N.W. Levin, "Body Composition Modeling in the Calf Using an Equivalent Circuit Model of Multi-Frequency Bioimpedance Analysis," Physiological Measurement, Vol. 26, No. 2, pp. S133-S143, 2005. https://doi.org/10.1088/0967-3334/26/2/013
  25. J.H. Kim, I.S. Jeong, S.W. Baik, and G.R. Jeon, "Early Detection of Peripheral Intravenous Infiltration Using Segmental Bioelectrical Impedance: Preliminary Study," Journal of Korea Multimedia Society, Vol. 20, No. 3, pp. 482-490, 2017. https://doi.org/10.9717/kmms.2017.20.3.482
  26. E. Hernandez-Balaguera, E. Lopez-Dolado, and J.L. Polo, "Obtaining Electrical Equivalent Circuits of Biological Tissues Using the Current Interruption Method, Circuit Theory and Fractional Calculus," Royal Society of Chemistry, Vol. 6, pp. 22312-22319, 2016.
  27. U.G. Kyle, I. Bosaeus, A.D. De Lorenzo, P. Deurenberg, M. Elia, J.M. Gomez, et al., "Bioelectrical Impedance Analysis-Part II: Utilization in Clinical Practice," Clinical Nutrition, Vol. 23, pp. 1430-1453, 2004. https://doi.org/10.1016/j.clnu.2004.09.012
  28. J.H. Kim, M.S. Lee, S.W. Baik, G.H. Kim, Y.J. Hwang, G.R. Jeon, "Impedance Parameter Variations at Intravenous (IV) Infiltration Using Bioelectrical Impedance: A Pilot Study," Journal of Korea Multimedia Society, Vol. 20, No. 10, pp. 1678-1688, 2017. https://doi.org/10.9717/KMMS.2017.20.10.1678
  29. B.C. Kim, C.M. Kim, and C.H. Lee, Multi-Channel Impedance Measuring Method and Multi-Channel Impedance Measuring Instrument, WO 2014/035040 A1, Patient PCT/KR2013/005779, Korea, 2013.
  30. J.H. Kim, B.J. Shin, S.W. Baik, and G.R. Jeon, "Early Detection of Intravenous Infiltration Using Multi-Frequency Bioelectrical Impedance Parameters: Pilot Study," Journal of Sensor Science and Technology, Vol. 26, No. 1, pp. 15-23, 2017. https://doi.org/10.5369/JSST.2017.26.1.15
  31. J.H. Kim, Y.J. Hwang, G.H. Kim, B.J. Shin, Y.J. Kim, E.J. Lee, G.R. Jeon, "Comparison of Infiltration Induced in Veins of Rabbit''s Ear and Human's Forearm by Using Bioelectrical Impedance: Pilot Study," Journal of Sensor Science and Technology, Vol. 26, No. 5, pp. 1-8, 2017. https://doi.org/10.5369/JSST.2017.26.1.1
  32. J.H. Kim, B.S. Shin, and G.R. Jeon, "Early Detection of Intravenous Infiltration Using Multi-frequency Bioelectrical Impedance Measurement System: Pilot Study," Journal of Information and Communication Convertgence Engineering, Vol. 15, No. 2, pp. 123-130, 2017.

피인용 문헌