Outcomes Associated with the Off-label Use of Medical Devices in Congenital Heart Disease at a Single Institute

  • Kong, Young Hwa (Department of Pediatrics, Samsung Medical Center, Sungkyunkwan University School of Medicine) ;
  • Song, Jinyoung (Department of Pediatrics, Samsung Medical Center, Sungkyunkwan University School of Medicine) ;
  • Huh, Jun (Department of Pediatrics, Samsung Medical Center, Sungkyunkwan University School of Medicine) ;
  • Kang, I-Seok (Department of Pediatrics, Samsung Medical Center, Sungkyunkwan University School of Medicine)
  • Received : 2016.09.02
  • Accepted : 2016.12.29
  • Published : 2017.07.31

Abstract

Background and Objectives: While the off-label use of implantable medical devices for treating congenital heart disease is not uncommon, the present conditions and outcomes of their off-label use have rarely been described. Therefore, this study evaluated the prevalence and results of using implantable devices off-label to treat congenital heart disease at a single institute. Subjects and Methods: This was a retrospective study based on the medical records of catheter-based interventions for congenital heart disease performed from July 1, 1995 to June 1, 2015. The inclusion criterion was the off-label use of an implantable device. Patient demographic data, procedural success, and follow-up status regarding late complications were investigated, and the results of the off-label use of each device were compared. Results: Off-label use occurred in 144 of 1730 interventions with device implantation, accounting for 8.3% of the interventions. The median patient age and mean body weight were 51.0 months and 16.3 kg, respectively. Immediate and late failures were found in 9 cases, and 3 cases of mortality were not directly related to the devices used. The overall success rate was 93.8%. There were no long-term complications of the off-label use of occlusion devices. No procedural failures resulted from stent implantation, but one case of stent malposition and two cases of stent fracture were identified after procedure completion. Conclusion: In general, the off-label use of implantable devices for treating congenital heart disease is safe and effective.

Keywords

References

  1. Holzer RJ, Sisk M, Phillips A. Hybrid balloon pulmonary valvuloplasty in a 700-g infant: thinking outside the box. Catheter Cardiovasc Interv 2008;72:93-6.
  2. Sutherell JS, Hirsch R, Beekman RH 3rd. Pediatric interventional cardiology in the United States is dependent on the off-label use of medical devices. Congenit Heart Dis 2010;5:2-7. https://doi.org/10.1111/j.1747-0803.2009.00364.x
  3. Baspinar O, Irdem A, Kilinc M. Off-label use of Amplatzer devices in congenital heart disorders during childhood. Acta Cardiol 2013;68:31-5. https://doi.org/10.1080/AC.68.1.2959629
  4. David Y, Hyman WA. Issues associated with off label use of medical devices. Conf Proc IEEE Eng Med Biol Soc 2007;2007:3556-8.
  5. Holzer R, Hijazi Z. The off-versus on-label use of medical devices in interventional cardiovascular medicine?: clarifying the ambiguity between regulatory labeling and clinical decision making, part III: structural heart disease interventions. Catheter Cardiovasc Interv 2008;72:848-52. https://doi.org/10.1002/ccd.21708
  6. McElhinney DB. Beyond indications: postmarket surveillance and the importance of expanded and off-label use of transcatheter devices in structural and congenital interventions. Circ Cardiovasc Interv 2012;5:739-40. https://doi.org/10.1161/CIRCINTERVENTIONS.112.976043
  7. Moore JW, Khan M. Gianturco coil occlusion of patent ductus arteriosus. Curr Interv Cardiol Rep 2001;3:80-5.
  8. Zellers TM, Wylie KD, Moake L. Transcatheter coil occlusion of the small patent ductus arteriosus (<4 mm): improved results with a "multiple coil-no residual shunt" strategy. Catheter Cardiovasc Interv 2000;49:307-13. https://doi.org/10.1002/(SICI)1522-726X(200003)49:3<307::AID-CCD17>3.0.CO;2-M
  9. Cho EH, Song J, Kang IS, et al. Transcatheter closure of small ductus arteriosus with amplatzer vascular plug. Korean J Pediatr 2013;56:396-400. https://doi.org/10.3345/kjp.2013.56.9.396
  10. Cho YK, Chang NK, Ma JS. Successful transcatheter closure of a large patent ductus venosus with the Amplatzer vascular plug II. Pediatr Cardiol 2009;30:540-2. https://doi.org/10.1007/s00246-009-9397-2
  11. Delaney JW, Fletcher SE. Patent ductus arteriosus closure using the $Amplatzer^{(R)}$ vascular plug II for all anatomic variants. Catheter Cardiovasc Interv 2013;81:820-4. https://doi.org/10.1002/ccd.24707
  12. Ng B, Schneider DJ, Hokanson JS. Closure of tubular patent ductus arteriosus in infants and small children with the Amplatzer Vascular Plug II. Congenit Heart Dis 2011;6:64-9. https://doi.org/10.1111/j.1747-0803.2010.00466.x
  13. Kim SJ, Huh J, Song JY, Yang JH, Jun TG, Kang IS. The hybrid perventricular closure of apical muscular ventricular septal defect with Amplatzer duct occluder. Korean J Pediatr 2013;56:176-81. https://doi.org/10.3345/kjp.2013.56.4.176
  14. Lee SM, Song JY, Choi JY, et al. Transcatheter closure of perimembranous ventricular septal defect using Amplatzer ductal occluder. Catheter Cardiovasc Interv 2013;82:1141-6. https://doi.org/10.1002/ccd.24810
  15. Walsh KP, Wilmshurst PT, Morrison WL. Transcatheter closure of patent foramen ovale using the Amplatzer septal occluder to prevent recurrence of neurological decompression illness in divers. Heart 1999;81:257-61. https://doi.org/10.1136/hrt.81.3.257
  16. O'Laughlin MP, Perry SB, Lock JE, Mullins CE. Use of endovascular stents in congenital heart disease. Circulation 1991;83:1923-39. https://doi.org/10.1161/01.CIR.83.6.1923
  17. Hascoet S, Baruteau A, Jalal Z, et al. Stents in paediatric and adult congenital interventional cardiac catheterization. Arch Cardiovasc Dis 2014;107:462-75. https://doi.org/10.1016/j.acvd.2014.06.005
  18. Lee KE, Seo YJ, Kim GB, et al. Complications of cardiac catheterization in structural heart disease. Korean Circ J 2016;46:246-55. https://doi.org/10.4070/kcj.2016.46.2.246
  19. Breinholt JP, Nugent AW, Law MA, Justino H, Mullins CE, Ing FF. Stent fractures in congenital heart disease. Catheter Cardiovasc Interv 2008;72:977-82. https://doi.org/10.1002/ccd.21742
  20. Peng LF, McElhinney DB, Nugent AW, et al. Endovascular stenting of obstructed right ventricle-to-pulmonary artery conduits: a 15-year experience. Circulation 2006;113:2598-605. https://doi.org/10.1161/CIRCULATIONAHA.105.607127
  21. Powell AJ, Lock JE, Keane JF, Perry SB. Prolongation of RV-PA conduit life span by percutaneous stent implantation. Intermediateterm results. Circulation 1995;92:3282-8. https://doi.org/10.1161/01.CIR.92.11.3282

Cited by

  1. Use of ASD closure device for the sealing of false lumen entry in the ascending aorta after dissection Type A surgical repair vol.14, pp.8, 2017, https://doi.org/10.1136/bcr-2021-242304