A Study on the Photoneutron Dose Estimation in Flattening Filter Mode and Flattening Filter Free Mode for Medical Linear Accelerator

의료용 선형가속기의 Flattening Filter Mode와 Flattening Filter Free Mode 간에 광중성자 선량 평가

  • Yang, Oh Nam (Department of Radiology, Mokpo Science University) ;
  • Lim, Cheong Hwan (Department of Radiological Science, Hanseo University)
  • 양오남 (목포과학대학교 방사선과) ;
  • 임청환 (한서대학교 방사선학과)
  • Received : 2017.04.27
  • Accepted : 2017.06.05
  • Published : 2017.06.30

Abstract

In this study, the generation of photoneutrons between the 10 MV FF mode and the FFF mode was evaluated and the amount of photoneutrons generated by the 10 MV and 15 MV energy changes in the FFF mode was evaluated. The generated neutrons were evaluated at 13 measurement points and the KTEPC was used to collect the generated neutrons. 10 MV FF mode was measured at 10 MV FF mode and FFF mode at all measurement points. In the superior direction, 0.455mSv and 0.152mSv were the largest, and more than 33% optical neutron was generated in FF. 10 MV in FFF mode, 15 MV in 15 MV, and 0.402 mSv in the direction of Superior, and 6.9% in the direction.

본 연구에서는 최신형 선형가속기 TrueBeam을 대상으로 10 MV FF mode와 FFF mode간 광중성자의 발생량을 평가하였고, FFF mode에서 10 MV, 15 MV 에너지 변화에 따라 발생되는 광중성자의 발생량을 평가하였다. 발생된 광중성자는 13개 측정점에서 평가하였으며, 발생된 광중성자 수집을 위해서는 한국천문연구원에서 개발한 조직등가비례계수기(KTEPC)를 이용하였다. 10 MV FF mode와 FFF mode 간 측정결과에서 10 MV FF mode가 모든 측정 점에서 높게 측정되었으며, Superior 방향에서 0.455 mSv와 0.152 mSv로 가장 많은 차이를 보였으며, FF에서 33% 이상 많은 광중성자가 발생하였다. FFF mode에서 10 MV, 15 MV 에너지에 따라서는 15 MV에서 모든 방향에서 높게 평가되었고, 방향 기준으로는 Superior 방향에서 0.402 mSv로 가장 높게 평가되었으며, 전체 평균 6.9%로 높게 측정되었다.

Keywords

References

  1. Ing H, Nelson WR, Shore RA: Unwanted photon and neutron radiation resulting from collimated photon beams interacting with the body of radiotherapy patients, Med. Phys, 9(1), 27-33, 1982 https://doi.org/10.1118/1.595137
  2. Lakosi L.: Photoneutron interrogation of low-enriched uranium induced by bremsstrahlung from a 4MeV linac", Section B, Beam interactions with materia ls and atoms, Nuclear Instruments & methods in physics research. Vol.266, No.2, 295-300, 2008
  3. Jung jae ho, Kim Hong Suk, Park Yoon Hwan, Lee Jae Ki: Evaluation of dose by photoneutron generated from medical linear accelerator, Proceedings of the Korean Radioactive Waste Society, 01, 296-296, 2004
  4. McGinley P. H.: Photoneutron Production in the Primary Barriers of Medical Accelerator Rooms, Health physics, 62(4), 359-362,1992 https://doi.org/10.1097/00004032-199204000-00011
  5. Nath R, Epp ER, Laughlin JS, Swanson WP, Bond VP: Neutrons from high-energy X-ray medical accelerators: an estimate of risk to the radiotherapy patient, Medical Physics, 11(3), 231-241, 1984 https://doi.org/10.1118/1.595497
  6. Fatemeh Torabi, S. Farhad Masoudi, Faezeh Rahmani: Photoneutron production by a 25MeV electron linac for BNCT application, Annals of Nuclear Energy, 54, 192-196, 2013 https://doi.org/10.1016/j.anucene.2012.11.001
  7. Difilippo, F., Papiez, L., Moskvin, V., et al.: Contamination dose from photoneutron processes in bodily tissues during therapeutic radiation delivery, Medical physics, 30(10), 2849-2854, 2003 https://doi.org/10.1118/1.1612947
  8. Eun-Tae Park: Evaluation of Photoneutron Dose in Radiotherapy Room Using MCNPX, The Journal of the Korea Contents Association, 15(6), 283-289, 2015 https://doi.org/10.5392/JKCA.2015.15.06.283
  9. ICRP 74: Conversion coefficients for use I nradiological protection against external radiation, International Commission on Radiological Protection, ICRP Publication 74, 1996
  10. ICRP 60: 1990 Recommendations of the Internaitonal Commission on Radiological Protection, Internaional Commission on Radiological Protectin, ICRP Publicaton 60, 1991
  11. Cho Woong, Park Jeong-Hoon, Jung Won-Gyun Suh Tae-Suk: Development of a Beam Source Modeling Technique for a Flattening Filter Free (FFF) Beam, Machine Learning and Applications and Workshops, 10th International Conference,.2, 215-219, 2011
  12. Chang Si-Young, Kim In-Gyoo: Developments in Radiation Health Science and Their Impact on Radiation Protection, Journal of radiation protection and research, 23(3), 185-196, 1998
  13. Radiation Protection & Safety : Chung-ku Publisher, 185, 2014