DOI QR코드

DOI QR Code

Ultra-low-power Pulse Oximeter with a 32.768 kHz Real Clock

  • Lee, Wonjun (Department of Electrical and Computer Engineering, Korea University) ;
  • Han, Youngsun (Department of Electronic Engineering, Kyungil University) ;
  • Kim, Chulwoo (Department of Electrical and Computer Engineering, Korea University) ;
  • Rieh, Jae-sung (Department of Electrical and Computer Engineering, Korea University) ;
  • Park, Jongsun (Department of Electrical and Computer Engineering, Korea University) ;
  • Park, Jae Young (Department of Urology, Korea University College of Medicine) ;
  • Kim, Seon Wook (Department of Electrical and Computer Engineering, Korea University)
  • Received : 2016.12.30
  • Accepted : 2017.02.23
  • Published : 2017.04.30

Abstract

A conventional pulse oximeter has high power consumption; thus, its mobility is severely limited. In this paper, we discuss the drawbacks of the existing pulse oximeters and propose a new ultra-low-power pulse oximeter that supports wireless data transmission for remotely monitoring vital signs, such as peripheral capillary oxygen saturation (SpO2) and beats per minute (BPM). We could notably reduce power consumption by using a low-frequency single clock in all well-customized modules. Also, our device is publicly certified, and thus, possibly engaged in clinical trials for commercial use.

Keywords

References

  1. Tavakoli, Maziar, Lorenzo Turicchia, and Rahul Sarpeshkar. "An ultra-low-power pulse oximeter implemented with an energy-efficient transimpedance amplifier." IEEE Transactions on Biomedical Circuits and Systems, vol. 4, no.1, pp.27-38, 2010. https://doi.org/10.1109/TBCAS.2009.2033035
  2. P. R. Metri and A. Desai, "Low power low cost single chip pulse oximeter design using msp430," INTERNATIONAL JOURNAL FOR RESEARCH IN EMERGING SCIENCE AND TECHNOLOGY, vol. 2, May 2015.Node.
  3. E. S. Winokur, T. O'Dwyer, and C. G. Sodini, "A low-power, dual-wavelength photoplethysmogram (ppg) soc with static and time-varying interferer removal," IEEE transactions on biomedical circuits and systems, vol. 9, no. 4, pp. 581-589, 2015. https://doi.org/10.1109/TBCAS.2014.2358673
  4. General Technical Document Review and Clinic Test Review, Korea Testing Laboratory. [Online].
  5. General Technical Document Review and Clinic Test Review, Ministry of Food and Drug Safety. [Online].
  6. S. Bagha, L. Shaw, "A Real Time Analysis of PPG Signal for Measurement of SpO2 and Pulse Rate", International Journal of Computer Application, vol. 36, no. 11, pp. 45-50, Dec. 2011.
  7. H. Lee, S. Park, C. Lim, and C. Kim, "A 100-nw 9.1-enob 20-ks/s sar adc for portable pulse oximeter." IEEE Trans. on Circuits and Systems, vol. 62-II, no. 4, pp. 357-361, April 2015.
  8. DATREND Systems Inc. SpO2 simulator