Abstract
In the case of nuclear medicine practitioners in medical institutions, a wide range of exposure dose to individual workers can be found, depending on the type of source, the amount of radioactivity, and the use of shielding devices in handling radioactive isotopes. In this regard, this study evaluated the organ dose on practitioners as well as the dose reduction effect of the L-block shielding device in handling the diagnostic radiation source through the simulation based on the Monte Carlo method. As a result, the distribution of organ dose was found to be higher as the position of the radiation source was closer to the handling position of a practitioner, and the effective dose distribution was different according to the ICRP tissue weight. Furthermore, the dose reduction effect according to the L-block thickness tended to decrease, which showed the exponential distribution, as the shielding thickness increased. The dose reduction effect according to each radiation source showed a low shielding effect in proportion to the emitted gamma ray energy level.
의료기관 내 핵의학 종사자는 방사성동위원소 취급 시 사용하는 선원의 종류, 방사능량, 차폐기구의 사용 여부에 따라 종사자 개인별로 광범위한 피폭선량을 나타낼 수 있다. 이에 본 연구에서는 몬테카를로 기법을 기반으로 한 모의실험을 통해 진단용 방사선원 취급 시 종사자의 장기별 선량평가와 L-block 차폐기구 사용에 따른 선량감소효과를 분석하였다. 그 결과, 방사선원의 취급 위치에 근접할수록 높은 장기선량 분포를 나타내었고, ICRP 조직가중치에 따라 유효선량 분포가 상이한 양상을 보였다. 또한 L-block 두께에 따른 선량감소효과는 차폐두께 증가에 따라 지수함수 분포로 감소되는 경향을 나타내었으며, 방사선원별 선량감소효과는 방출하는 감마선 에너지에 비례하여 낮은 차폐효과를 보였다.