Ex vivo High-resolution Optical Coherence Tomography (OCT) Imaging of Pleural Reaction after Pleurodesis Using Talc

  • Ahn, Yeh-Chan (Department of Biomedical Engineering and Center for Marine-Integrated Biomedical Technology, Pukyong National University) ;
  • Oak, Chulho (Innovative Biomedical Technology Research Center, Kosin University College of Medicine) ;
  • Park, Jung-Eun (Department of Biomedical Engineering and Center for Marine-Integrated Biomedical Technology, Pukyong National University) ;
  • Jung, Min-Jung (Innovative Biomedical Technology Research Center, Kosin University College of Medicine) ;
  • Kim, Jae-Hun (Department of Internal Medicine, Kosin University College of Medicine) ;
  • Lee, Hae-Young (Department of Chest Surgery, Kosin University College of Medicine) ;
  • Kim, Sung Won (Department of Internal Medicine, Kosin University College of Medicine) ;
  • Park, Eun-Kee (Innovative Biomedical Technology Research Center, Kosin University College of Medicine) ;
  • Jung, Maan Hong (Department of Chest Surgery, Kosin University College of Medicine)
  • 투고 : 2016.06.03
  • 심사 : 2016.09.08
  • 발행 : 2016.10.25

초록

The pleura is known as an end target organ of exposure to toxic environmental materials such as fine particulate matter and asbestos. Moreover, long-term exposure to hazardous materials can eventually lead to fatal lung disease such as diffuse pleural fibrosis or mesothelioma. Chest computed tomography (CT) and ultrasound are gold standard imaging modalities for detection of advanced pleural disease. However, a diagnostic tool for early detection of pleural reaction has not been developed yet due to difficulties in imaging ultra-fine structure of the pleura. Optical coherence tomography (OCT), which provides cross-sectional images of micro tissue structures at a resolution of 2-10 μm, can image the mesothelium with a thickness of ~100 μm and therefore enables investigation of the early pleural reaction. In this study, we induced the early pleural reaction according to a time sequence after pleurodesis using talc, which has been widely used in the clinical field. The pleural reaction in talc grouped according to the time sequence (1st, 2nd, 4th weeks) showed a significant thickening (average thickness: 45 ± 7.5 μm, 80 ± 10.7 μm, 90 ± 12.5 μm), while the pleural reaction in sham and normal groups showed pleural change from normal to minimal thickening (average thickness: 16 ± 5.5 μm, 17 ± 4.5 μm, 15 ± 6.5 μm, and 12 ± 7.5 μm, 13 ± 2.5 μm, 12 ± 3.5 μm). The measurement of pleural reaction by pathologic examinations was well-matched with the measurement by OCT images. This is the first study for measuring the thickness of pleural reactions using a biophotonic modality such as OCT. Our results showed that OCT can be useful for evaluating the early pleural reaction.

키워드

I. INTRODUCTION

It is widely accepted that chronic exposure to hazardous airborne particles is closely associated with chronic lung disease [1, 2]. The World Health Organization has identified fine particulate matters as 1st-grade carcinogens of lung cancer [3]. Similar to other well-known carcinogens such as asbestos, long-term exposure to fine particulate matter can eventually lead to development of fatal airway diseases such as pleural fibrosis and pleural malignancy, which have been increasing in incidence worldwide [4, 5]. Therefore, the development of a diagnostic imaging modality for such exposure has been an essential issue in health care services [6]. Computed tomography (CT) and ultrasound (US) are generally used to detect pleural disease [7, 8]. However, the diagnostic yield of CT and US is not high due to limited resolution [9, 10]. Due to resolution limits, the ultra-fine structure of the pleural layer has been an area of uncertainty for visualization using conventional methods [11, 12]. Recently, optical coherence tomography (OCT) has been developed, which is an imaging modality with non-invasive, high-speed, high-resolution capabilities [13]. In spite of a limited imaging depth, OCT provides real-time cross-sectional images with a micrometer scale. Several OCT studies related to pulmonology have been published [14]. Talc has been commonly used in the chemical, textile, cosmetic, construction industries, and even in the health care industry [15]. There have been many reports regarding potential hazard of industrial talc to induce asbestos-related diseases, including diffuse pleural thickening, interstitial fibrosis, and mesothelioma. In this study [16], we used an animal model of pleurodesis and imaged the pleural reaction induced by talc. Pleurodesis is a medical method used to induce the pleural reaction by infusing a sclerosing agent into the pleural cavity (Fig. 1). Talc is the most commonly used sclerosing agent, with many reports of its effectiveness [17, 18]. The main target of the sclerosing agent is the pleural surface, a regenerative layer for the early pleural reaction after pleurodesis. To the best of our knowledge, this is the first study testing the feasibility of OCT for evaluation of the early pleural reaction based on the time sequence after pleurodesis. In addition, we performed ultrasound examinations for pleural reaction at 28 days after pleurodesis.

FIG. 1.Illustration of (a) normal chest x-ray, (b) pleurodesis, which is a medical procedure in which the pleural space is artificially obliterated by infusion of sclerosing agents, and (c) abnormal chest x-ray with pleural reaction, which is the pleural thickening caused by inflammatory response of mesothelial cell after pleurodesis.

 

II. METHODS

2.1. Animal Preparation

All experiments were performed in accordance with the Guide for the Care and Use of Laboratory Animals (DHEW publication NIH 85-23, revised 2010, Office of Science and Health Reports, DRR/NIH, Bethesda, MD). The study was approved by the Animal Care and Use Committee at Kosin University College of Medicine. Male New Zealand white rabbits weighing 3.0 to 3.5 kg were used in this study. The initial intramuscular anesthesia was performed with ketamine 5 mg/kg and xylazine 0.8 mg/kg. We maintained the anesthetic depth by injection of 10 mg/h ketamine and 3 mg/h xylazine. Oxygen saturation was monitored with a pulse oximeter in the ear.

2.2. Pleurodesis

Figure 1 describes a brief concept of pleurodesis and Fig. 2 shows a procedure of pleurodesis of this study. The rabbits were fixed on the operation table in a lateral position for preparation and image acquisition. The right chest was shaved, and a 1.0-cm skin incision was made midway between the spine and the scapular tip. A 16-gauge plastic catheter was placed percutaneously into the right pleural space. To compare the effects of the agents used for pleurodesis, rabbits were divided into three groups treated intrapleurally: normal (control group, n = 5), 400 mg/kg talc (talc group, n = 15 (day 7, N = 5; day 14, N = 5; and day 28, N = 5)), and 10 ml normal saline (sham control group, n = 15 (day 7, N = 5; day 14, N=5; and day 28, N = 5)). Before starting pleurodesis procedures, we selected 400 mg/kg of talc because it was the lowest dose that produced recognizable changes in the previous study [19]. The talc slurry was instilled into the right thoracic cavity through the 16-gauge plastic catheter.

FIG. 2.Illustration of animal experiment sequence. (a) Step 1; the right chest was shaved and a 1.0 cm skin incision was made midway between the spine and the scapular tip, (b) Step 2; a 16-gauge plastic catheter was placed percutaneously into the right pleural space, (c) Step 3; talc 1 mg/ml was installed into pleural cavity, and (d) Step 4; resection of chest wall after the rabbit was euthanized with CO2 gas, and OCT imaging was performed. (e) chest x-ray after procedure which showed the 16-gauge plastic catheter (arrows) in the thoracic cavity.

2.3. Follow-up after Pleurodesis

After the procedure, the rabbits were monitored daily for weight-loss, inappetance, moribund state, infection, and organ system dysfunction. All efforts were made to minimize suffering.

Animals were euthanized using CO2 gas after an intramuscular injection of ketamine at day 7 (N = 5), day 14th (N = 5), or day 28 (N = 5) after pleurodesis, according to group assignment, and the pleural tissue was analyzed using microscopic, OCT, and pathologic examinations.

2.4. OCT System

Two lab-made spectral-domain optical coherence tomography (SD-OCT) systems were used to measure thickness of pleura from the basement membrane to the top of granulation in real-time. We used an 850 nm OCT system with a better resolution for imaging the early pleural reaction at the first week, while using a 1310 nm OCT system with a better penetration for imaging the advanced pleural reaction at the second and fourth weeks.

Figure 3 shows a schematic diagram of our SD-OCT based on a Michelson interferometer. A broadband light source sent a beam to a fiber based 2-by-2 beam splitter through optical fiber and the split beams were directed to reference arm and sample arm. The light source has a center wavelength of 850 nm (BroadLighter D855, Superlum, Ireland) or 1310 nm (SLED Butterfly, EXALOS, Schlieren, Switzerland) and a full width at half maximum (FWHM) of 100 nm or 80 nm, respectively. After that, beams were reflected at reference mirror and samples, respectively. The reflected beams were coupled into each optical fiber and delivered to the 2-by-2 beam splitter. The merged beam was diffracted by a grating (1800 lines/mm for the 850 nm OCT system or 1145 lines/mm for the 1310 nm OCT system, Wasatch Photonics, NC, USA), depending on wavelength. The diffracted interference pattern was detected by a CMOS camera with 4096 pixels (Sprint spL4096-140km, Basler, PA, USA) at a line rate of 140 kHz for the 850 nm OCT system or an InGaAs camera with 1024 pixels (SU1024-LDH2, Goodrich, Princeton NJ, USA) at a line rate of 92 kHz for the 1310 nm OCT system. A scan unit with two galvanometers (6220H, Cambridge Technology, MA, USA) allowed to scan specimens in two axes with 5 by 5 mm2 scanning area. To obtain specifications of the OCT system, we measured the point spread function. The 850 nm OCT system has a depth resolution of 3 μm in air, a roll-off distance of 1 mm at 12 dB power drop, and a dynamic range of 103 dB. The 1310 nm OCT system has a depth resolution of 11.629 μm in air, a roll-off distance of 1 mm at 5.65 dB power drop, and a dynamic range of 106 dB.

FIG. 3.Schematic diagram of SD-OCT system, with 850 nm or 1310 nm broadband source.

FIG. 4.Normal chest wall and parietal pleura; (a) gross image, (b and e) 1310 nm OCT images; yellow arrow means the top surface of granulation and yellow head means the basement membrane at the pleural layer, (c and f) pathology (H-E stain) magnified to 40 times, and (d) thickness measured at A,B and C and arithmetically averaged.

2.5. US System

We used a commercial US system with a linear transducer (10 M Hz) (Sonix Touch, Ultrasonix, British Columbia, Canada). The US system provided 93 μm of spatial resolution for x and y directions.

2.6. Pathologic Examination

Ex vivo samples were selected from the parietal pleura including apex, mediastinum, and diaphragm. Pleural thicknesses at these three points were measured and mean thicknesses were calculated. Lung tissue (including pleura, alveoli, intercostal muscle, fascia, skin) were cut into 6×6 cm2 pieces, fixed in 10% neutral buffered formalin (NBF), and embedded in paraffin. Serial sections (4 μm thick) were stained with hematoxylin and eosin (H&E) and examined by microscopy.

2.7. Statistical Analysis

Data were presented as mean ± standard deviation (SD). Statistical analysis was conducted using SPSS, version 17.0 (SPSS Inc., Chicago, IL, USA). A two-tailed p-value <0.05 was considered statistically significant.

 

III. RESULTS AND DISCUSSION

The gross examinations of granulation showed scattered nodular lesions in the parietal pleural surface and the nodular lesion increased in size for 4 weeks as indicated by blue arrows in Figs. 5(a), 5(e), 6(a), 6(d), 7(a), and 7(e). The distance from the basement membrane to the top of granulation using OCT examination showed gradual increase according to time (Figs. 5(b), 5(f), 6(b), 6(e), 7(b), and 7(f)). The pathologic examination showed infiltration of inflammatory cells surrounding talc particles (Fig. 6(f)). The pathologic findings were well matched with OCT images (Figs. 5(d), 5(h), 7(c), and 7(g)). The change of pleural reaction in the talc group at the 1st and 2nd weeks showed the significant thickening (average thickness: 45 ± 7.5 μm, 80 ± 10.7 μm), while the pleural reaction in sham and normal groups showed pleural change from normal to minimal thickening (average thickness: 16 ± 5.5 μm, 17 ± 4.5 μm and 12 ± 7.5 μm, 13 ± 2.5 μm. p=0.03, p=0.03 respectively). (Figs. 5, 6 and Table 1).

FIG. 5.Chest wall and mild pleural thickening a week later after pleurodesis; (a and e) gross images, (b and f) 850 nm OCT images; the region between the red arrows shows the area of interest, (c) pathology magnified to 1.25 times, (d, g and h) pathology magnified to 100 times.

FIG. 6.Chest wall and moderate pleural thickening two weeks later after pleurodesis; (a and d) gross images, (b and e) 1310 nm OCT images; the region between the red arrows shows the area of interest, (c) pathology magnified to 100 times, and (f) pathology magnified to 100 times (the green arrows show typical talc particles).

FIG. 7.Chest wall and moderate pleural thickening four weeks later after pleurodesis; (a and e) gross images, (b and f) 1310 nm OCT images; the region between the red arrows shows the area of interest, (c and g) pathology magnified to 40 times, and (d and h) 10 M Hz US images and the region between the red arrows shows the area of interest.

TABLE 1.Measured pleural thickness by pathology and ex vivo OCT image.

With regard to the early pleural reaction, Adamson et al. reported that the peak blood level of early mesothelial proliferation occurred in the first 2-4 weeks after asbestos exposure [20]. However, there was no study of early pleural reaction after pleurodesis using imaging modality. In this study, the high-resolution capability of OCT was helpful in showing the time sequence of events in the delicately changing pleural space. OCT was able to detect pleural changes even between the 1st and 2nd weeks. There was no significant change in the thickness of pleural reaction in talc group at the 2nd and 4th weeks (average thickness: 80 ± 10.7 μm, 90 ± 12.5 μm), and the pleural reaction in sham and normal groups also showed no significant pleural change from normal to minimal thickening (average thickness: 17 ± 4.5 μm, 15 ± 6.5 μm, and 13 ± 2.5 μm, 12 ± 3.5 μm)(Figs. 6, 7 and Table 1). In the study of Kennedy et al., it was reported that pleural adhesions formed shortly after talc administration and did not increase over time after initial pleural reaction. Our results from the pleural thickness between the 2nd and 4th weeks were well correlated with findings from previous studies [21].

In our study, the measurement of pleural reaction by pathologic examinations were well matched with the measurement by OCT images. In comparison with previous study using pathologic examination, Muta et al. [19] reported similar result in the thickness of pleural reaction measured on the 30th day after talc exposure in rats. Therefore, OCT has a potential to provide in vivo imaging of mesothelium based on thoracoscopic examination and the detection of pleural lesions in an early stage. Therefore further in vivo study using real-time OCT would be warranted in an animal model. Brenner group firstly demonstrated the in vivo thoracoscopic imaging capabilities of three-dimensional OCT systems with improved forward-scanning rigid GRIN lens rod probes for detecting pleural cancer [22]. Clinically, mesothelioma and pleural plaques occur exclusively in parietal mesothelium, not in the visceral mesothelium [23].

In the present study, US examination visualized localized nodular pleural reaction after talc pleurodesis but could not measure the thickness of pleural reaction, while OCT could show thickening of pleural layers (Fig. 7). In study of fusion image using intra-vascular ultrasound (IVUS) and OCT image for detecting atherosclerosis, multimodal fusion imaging showed a complimentary role from the view of localization of lesion with IVUS and specification of lesion with OCT [24, 25]. Therefore, OCT might have a potential complementary role in the early detection of pleural lesions when used in combination with other diagnostic approaches such as endoscopic ultrasound. Our study contributes to the knowledge of the biologic behavior of mesothelium in the carcinogenesis of pleura. Additionally, further developments in imaging technologies such as ultrahigh resolution imaging are expected to improve the diagnostic accuracy and ability to characterize pathologic events occurring after injury. In this study, two kinds of OCT (850 and 1310 nm) were used for visualizing the pleural reaction from early response at the 1st week to late responses at the 2nd and 4th weeks. The 850 nm OCT system had a higher depth resolution than the 1310 nm OCT system so that it provided more accurate thickness estimation in an early stage. The 1310 nm OCT system has a deeper imaging depth. Both systems, however, were enough to image the whole thickness of mesothelium.

 

IV. CONCLUSION

An animal model of pleurodesis for visualizing the pleural reaction using OCT was feasible. Compared with the normal and normal saline groups, the talc group showed significant pleural reaction. Two house-made OCT systems were utilized to image the mesothelial layer in the pleura. Increase in parietal pleural thickness might be a mechanism underlying the effectiveness of talc pleurodesis for pleural effusions including malignant effusions.

참고문헌

  1. C. A. Pope III, R. T. Burnett, M. J. Thun, E. E. Calle, D. Krewski, K. Ito, and G. D. Thurston, “Lung cancer, cardiopulmonary mortality, and long-term exposure to fine particulate air pollution,” JAMA 287(9), 1132-1141 (2002). https://doi.org/10.1001/jama.287.9.1132
  2. V. D. Aaron, R. V. Martin, M. Brauer, R. Kahn, R. Levy, C. Verduzco and P. J. Vileneuve, “Global estimates of ambient fine particulate matter concentrations from satellite-based aerosol optical depth: development and application,” Environ. Health Perspect. 118(6), 847-855 (2010). https://doi.org/10.1289/ehp.0901623
  3. O. Raaschou-Nielsen, R. Beelen, M. Wang, G. Hoek, Z. J. Andersen, B. Hoffmann, and M. Nieuwenhuijsen, “Particulate matter air pollution components and risk for lung cancer,” Environ. Int. 87, 66-73 (2016). https://doi.org/10.1016/j.envint.2015.11.007
  4. M. Bayram and N. D. Bakan, “Environmental exposure to asbestos: from geology to mesothelioma,” Curr. Opin. Pulm. Med. 20(3), 301-307 (2014). https://doi.org/10.1097/MCP.0000000000000053
  5. N. Ndlovu, J. T. Naude, and J. Murray, “Compensation for environmental asbestos-related diseases in South Africa: a neglected issue,” Glob. Health Action 6 19410 (2013). https://doi.org/10.3402/gha.v6i0.19410
  6. J. P. Corcoran, L. Acton, A. Ahmed, R. J. Hallifax, I. Psallidas, J. M. Wrightson, and F. V. Gleeson, “Diagnostic value of radiological imaging pre‐and post‐drainage of pleural effusions,” Respirology 21(2), 392-395 (2016). https://doi.org/10.1111/resp.12675
  7. T. Havelock, R. Teoh, D. Laws, and F. Gleeson, “Pleural procedures and thoracic ultrasound: British Thoracic Society pleural disease guideline,” Thorax 65(Suppl 2), i61-i76 (2010). https://doi.org/10.1136/thx.2010.137026
  8. A. N. Leung, N. L. Müller, and R. R. Miller. “CT in differential diagnosis of diffuse pleural disease,” AJR Am. J. Roentgenol. 154(3), 487-492 (1990). https://doi.org/10.2214/ajr.154.3.2106209
  9. P. C. Yang, K. T. Luh, D. B. Chang, H. D. Wu, C. J. Yu, and S. H. Kuo, “Value of sonography in determining the nature of pleural effusion: analysis of 320 cases,” AJR Am. J. Roentgenol. 159(1), 29-33 (1992). https://doi.org/10.2214/ajr.159.1.1609716
  10. A. C. Friedman, S. B. Fiel, M. S. Fisher, P. D. Radecki, A. S. Lev-Toaff, and D. F. Caroline, “Asbestos-related pleural disease and asbestosis: a comparison of CT and chest radiography,” AJR Am. J. Roentgenol. 150(2), 269-275 (1988). https://doi.org/10.2214/ajr.150.2.269
  11. S. Standring, Gray’’s anatomy: the anatomical basis of clinical practice. Elsevier Health Sciences (Elsevier Health Sciences, 2015).
  12. Y. C. Ahn, Y. G. Chae, S. S. Hwang, B. K. Chun, M. H. Jung, S. J. Nam, H. Y. Lee, J. M. Chung, C. Oak and E. K. Park, “In vivo Optical Coherence Tomography Imaging of the Mesothelium Using Developed Window Models,” J. Opt. Soc. Korea 19(1), 69-73 (2015). https://doi.org/10.3807/JOSK.2015.19.1.069
  13. D. Huang, E. A. Swanson, C. P. Lin, J. S. Schuman, W. G. Stinson, W. Chang, M. R. Hee, T. Flotte, K. Gregory, and C. A. Puliafito, “Optical coherence tomography,” Science 254, 1178 (1991). https://doi.org/10.1126/science.1957169
  14. R. Hou, T. Le, S. D. Murgu, Z. Chen, and M. Brenner, “Recent advances in optical coherence tomography for the diagnoses of lung disorders,” Expert Rev. Respir. Med. 5(5), 711-724 (2011). https://doi.org/10.1586/ers.11.59
  15. L. Paoletti, S. Caiazza, G. Donelli, and F. Pocchiari, “Evaluation by electron microscopy techniques of asbestos contamination in industrial, cosmetic, and pharmaceutical talcs,” Regul Toxicol Pharmacol 4, 222-35 (1984). https://doi.org/10.1016/0273-2300(84)90022-9
  16. J. Kim, C. Oak, T. Jang, M. Jung, B. Chun, E. K. Park, and K. Takahashi, “Lung Cancer Probably Related to Talc Exposure: a Case Report,” Industrial Health 51(2), 228-231 (2013). https://doi.org/10.2486/indhealth.MS1385
  17. J. R. Campos, F. S. Varga, E. C. Werebe, P. Cardoso, L. R. Teixeira, F. B. Jatene and R. W. Light, “Thoracoscopy talc poudrage: a 15-year experience,” Chest 119, 801-806 (2001). https://doi.org/10.1378/chest.119.3.801
  18. Y. Chen, C. Li, L. Xu, H. Lin, and Y. Cui, “Novel treatment for chylothorax after esophagectomy with 50% glucose pleurodesis,” Ann Vasc Surg. 24 694.e9-694.13 (2010). https://doi.org/10.1016/S0890-5096(10)00414-0
  19. F. Muta, S. Takamori, T. Matsuo, Y. Iwasaki, and K. Yoshiyama, “Changes in the Pleural Cavity by Pleurodesis using Talc or OK-432: An Experimental Study,” Surg. Today 41, 111-114 (2011). https://doi.org/10.1007/s00595-010-4275-3
  20. I. Y. Adamson, J. Bakowska, and D. H. Bowden, “Mesothelial cell proliferation: a nonspecific response to lung injury associated with fibrosis,” Am J Respir Cell Mol Biol 10(3), 253-258 (1994). https://doi.org/10.1165/ajrcmb.10.3.7509611
  21. L. Kennedy, R. A. Harley, S. A. Sahn, and C. Strange, “Talc slurry pleurodesis. Pleural fluid and histologic analysis,” Chest 107, 1707-1712 (1995). https://doi.org/10.1378/chest.107.6.1707
  22. T. Xie, G. Liu, K. Kreuter, S. Mahon, H. Colt, D. Mukai, G. M. Peavy, Z. Chen, and M. Brenner, “In vivo three-dimensional imaging of normal tissue and tumors in the rabbit pleural cavity using endoscopic swept source optical coherence tomography with thoracoscopic guidance,” J. Biomed. Opt. 14(6), 064045 (2009). https://doi.org/10.1117/1.3275478
  23. F. Whitwell, J. Scott, and M. Grimshaw. “Relationship between occupations and asbestosfibre content of the lungs in patients with pleural mesothelioma, lung cancer, and other diseases,” Thorax 32(4), 377-386 (1977). https://doi.org/10.1136/thx.32.4.377
  24. L. Räber, J. Heo, M. Radu, H. Garcia, G. Stefanini, A. Moschovitis, J. Dijkstra, H. Kelbaek, S. Windecker, and P. Serruys. “Offline fusion of co-registered intravascular ultrasound and frequency domain optical coherence tomography images for the analysis of human atherosclerotic plaques,” Euro Intervention 8(1), 98-108 (2012).
  25. J. Li, T. Ma, D. Mohar, E. Steward, M. Yu, Z. Piao, Y. He, K. Shung, Q. Zhou, P. Patel, and Z. Chen. “Ultrafast optical-ultrasonic system and miniaturized catheter for imaging and characterizing atherosclerotic plaques in vivo,” Scientific Reports 5, 18406. (2015). https://doi.org/10.1038/srep18406