DOI QR코드

DOI QR Code

경피적 관상동맥 확장 성형술 도중 우관상동맥 뒤가쪽가지에서 발생한 유도철사 부러짐

Guide wire fracture during percutaneous coronary intervention

  • Kim, Hak Ro (Department of Internal Medicine, Sunlin Hospital) ;
  • Yim, Tae Hoon (Department of Internal Medicine, Sunlin Hospital) ;
  • Kim, Byung Chul (Department of Internal Medicine, Sunlin Hospital) ;
  • Lee, Ho Jun (Department of Internal Medicine, Sunlin Hospital) ;
  • Oh, Hong Geun (Department of Internal Medicine, Sunlin Hospital) ;
  • Ju, Hyun Sik (Department of Internal Medicine, Sunlin Hospital) ;
  • Kim, Tae Jin (Department of Internal Medicine, Sunlin Hospital) ;
  • Kim, Young Bok (Department of Internal Medicine, Sunlin Hospital)
  • 투고 : 2015.01.06
  • 심사 : 2015.04.02
  • 발행 : 2016.06.30

초록

Guide wire fracture during percutaneous coronary intervention (PCI) is rare. It can cause fatal complications such as thrombus formation, embolization, and perforation. Guide wire fracture could occur during intervention for severely calcified stenotic lesions, and rarely from distal small branches of stenotic lesions. There are several methods for its management depending on the material character, position, length of the remnant, and the patient's condition. If percutaneous retrieval was not achieved, the surgical procedure should be considered for prevention of potential risks, although the remnant guide wire does not usually cause complications. We experienced a patient with a guide wire fracture during PCI, and managed to prevent its complications through surgical removal of the remnant wire. We report this case here.

키워드

참고문헌

  1. Hartzler GO, Rutherford BD, McConahay DR. Retained percutaneous transluminal coronary angioplasty equipment components and their management. Am J Cardiol 1987;60:1260-4. https://doi.org/10.1016/0002-9149(87)90604-7
  2. Khonsari S, Livermore J, Mahrer P, Magnusson P. Fracture and dislodgment of floppy guidewire during percutaneous transluminal coronary angioplasty. Am J Cardiol 1986;58:855-6. https://doi.org/10.1016/0002-9149(86)90370-X
  3. Hong YM, Lee SR. A case of guide wire fracture with remnant filaments in the left anterior descending coronary artery and aorta. Korean Circ J 2010;40:475-7. https://doi.org/10.4070/kcj.2010.40.9.475
  4. Al-Moghairi AM, Al-Amri HS. Management of retained intervention guide-wire: a literature review. Curr Cardiol Rev 2013;9:260-6. https://doi.org/10.2174/1573403X11309030010
  5. Woodfield SL, Lopez A, Heuser RR. Fracture of coronary guidewire during rotational atherectomy with coronary perforation and tamponade. Cathet Cardiovasc Diagn 1998;44:220-3. https://doi.org/10.1002/(SICI)1097-0304(199806)44:2<220::AID-CCD20>3.0.CO;2-P
  6. Al-Amri HS, AL-Moghairi AM, Calafiore AM. Left main approach for retrieval of retained guidewire fragment. J Card Surg 2012;27:307-8. https://doi.org/10.1111/j.1540-8191.2011.01352.x
  7. Cho YH, Park S, Kim JS, Lee SY, Choi D, Jang Y, et al. Rescuing an entrapped guidewire using a Tornus catheter. Circ J 2007;71:1326-7.
  8. Kim TJ, Kim JK, Park BM, Song PS, Kim DK, Kim KH, et al. Fatal subacute stent thrombosis induced by guidewire fracture with retained filaments in the coronary artery. Korean Circ J 2013;43:761-5. https://doi.org/10.4070/kcj.2013.43.11.761
  9. Karabulut A, Daglar E, Cakmak M. Entrapment of hydrophilic coated coronary guidewire tips: which form of management is best? Cardiol J 2010;17:104-8.