동적 세기조절방사선 치료 시 선량률 변화에 따른 선량학적엽간격 변화 분석

Analysis of dosimetric leaf gap variation on dose rate variation for dynamic IMRT

  • 양명식 (전북대학교병원 방사선종양학과, 전북대학교 의학전문대학원 방사선종양학 교실) ;
  • 박주경 (전북대학교병원 방사선종양학과, 전북대학교 의학전문대학원 방사선종양학 교실) ;
  • 이승훈 (전북대학교병원 방사선종양학과, 전북대학교 의학전문대학원 방사선종양학 교실) ;
  • 김양수 (전북대학교병원 방사선종양학과, 전북대학교 의학전문대학원 방사선종양학 교실) ;
  • 이선영 (전북대학교병원 방사선종양학과, 전북대학교 의학전문대학원 방사선종양학 교실) ;
  • 차석용 (전북대학교병원 방사선종양학과, 전북대학교 의학전문대학원 방사선종양학 교실)
  • Yang, Myung Sic (Department of Radiation Oncology, Chonbuk National University Hospital, Department of Radiation Oncology, Chonbuk National University Medical School) ;
  • Park, Ju Kyeong (Department of Radiation Oncology, Chonbuk National University Hospital, Department of Radiation Oncology, Chonbuk National University Medical School) ;
  • Lee, Seung Hun (Department of Radiation Oncology, Chonbuk National University Hospital, Department of Radiation Oncology, Chonbuk National University Medical School) ;
  • Kim, Yang Su (Department of Radiation Oncology, Chonbuk National University Hospital, Department of Radiation Oncology, Chonbuk National University Medical School) ;
  • Lee, Sun Young (Department of Radiation Oncology, Chonbuk National University Hospital, Department of Radiation Oncology, Chonbuk National University Medical School) ;
  • Cha, Seok Yong (Department of Radiation Oncology, Chonbuk National University Hospital, Department of Radiation Oncology, Chonbuk National University Medical School)
  • 투고 : 2016.05.27
  • 심사 : 2016.06.23
  • 발행 : 2016.06.30

초록

동적 세기조절방사선치료 시 선량률 변화에 따른 다엽콜리메이터의 엽의 위치를 반영하는 선량학적엽간격과 다엽콜리메이터 투과계수 변화를 분석하여 다엽콜리메이터의 정확성을 평가하고자 하였다. Millennium 120 MLC 시스템이 장착된 선형가속기의 6 MV와 10 MV X선으로 물 팬텀의 깊이 10 cm에서 CC13과 FC-65G 전리함을 이용하여 선량률을 200, 300, 400, 500, 600 MU/min으로 변화시켜 선량학적엽간격과 다엽콜리메이터 투과계수를 측정하였다. 400 MU/min의 선량률 기준으로 200, 300, 400, 500, 600 MU/min으로 선량률을 변경하여 선량학적엽간격 값을 측정한 결과, 6 MV의 경우 각각 -2.59, -1.89, 0.00, -0.58, -2.89%의 차이가 나타났고, 10 MV에서는 각각 ?2.52, -1.69, 0.00, +1.28, -1.98%의 차이가 나타났다. 다엽콜리메이터 투과계수는 두 종류의 에너지와 모든 선량률에서 약 ${\pm}1%$ 이내의 범위로 측정되었다. 본 연구는 동적 세기조절방사선치료 시 선량률 변화에 대하여 다엽콜리메이터의 선량학적엽간격과 투과계수 변화를 평가하였다. 선량률 변화에 따라서 다엽콜리메이터의 투과계수의 차이는 미미했지만, 선량학적엽간격의 차이는 큰 것으로 확인하였다. 따라서 동적 세기조절방사선치료 시 임의로 선량률을 변경하면 종양에 전달되는 선량에 많은 영향을 미치므로 치료 중에 선량률을 변화시키지 않는 것이 더욱 정확한 방사선치료 방법이라 사료된다.

To evaluate the position accuracy of the MLC. This study analyzed the variations of the dosimetric leaf gap(DLG) and MLC transmission factor to reflect the location of the MLC leaves according to the dose rate variation for dynamic IMRT. We used the 6 MV and 10 MV X-ray beams from linear accelerator with a Millennium 120 MLC system. We measured the variation of DLG and MLC transmission factor at depth of 10 cm for the water phantom by varying the dose rate to 200, 300, 400, 500 and 600 MU/min using the CC13 and FC-65G chambers. For 6 MV X-ray beam, a result of measuring based on a dose rate 400 MU/min by varying the dose rate to 200, 300, 400, 500 and 600 MU/min of the difference rate was respectively -2.59, -1.89, 0.00, -0.58, -2.89%. For 10 MV X-ray beam, the difference rate was respectively ?2.52, -1.69, 0.00, +1.28, -1.98%. The difference rate of MLC transmission factor was in the range of about ${\pm}1%$ of the measured values at the two types of energy and all of the dose rates. This study evaluated the variation of DLG and MLC transmission factor for the dose rate variation for dynamic IMRT. The difference of the MLC transmission factor according to the dose rate variation is negligible, but, the difference of the DLG was found to be large. Therefore, when randomly changing the dose rate dynamic IMRT, it may significantly affect the dose delivered to the tumor. Unless you change the dose rate during dynamic IMRT, it is thought that is to be the more accurate radiation therapy.

키워드

참고문헌

  1. Louis Lee, Quynh-Thu Le, Lei Xing: Retrospective IMRT dose reconstruction based on cone-beam CT and MLC log-file. International Journal of Radiation Oncology Biology Physics 2008;70:634-644 https://doi.org/10.1016/j.ijrobp.2007.09.054
  2. Youngyih Han: Review on the Pre-treatment Quality assurance for intensity modulated radiation therapy. Progress in MEDICAL PHYSICS 2013;24(4):213-219 https://doi.org/10.14316/pmp.2013.24.4.213
  3. So Yeon Park, Yang Kyun Park, Jong Min Choi, Chang Heon: MU Fluence reconstruction based-on delivered leaf position for IMRT quality assurance. JOURNAL OF RADIATION PROTECTION 2011;38(1):28-34
  4. Graves1 Maria N. , Antoinette V. Thompson, Mary K. Martel, Daniel L. McShan1 and Benedick A. Fraass: Calibration and quality assurance for rounded leaf-end MLC systems. Medical Physics 2001;28(11):2227-2233 https://doi.org/10.1118/1.1413517
  5. Jeong Woo Lee, Semie Hong, Yon Lae Kim, Kyoung Sik Choi, Jin Beom Chung, Doo Hyun Lee, Tae Suk Suh: Effects of static dosimetric leaf gap on MLC-based small beam dose distribution for intensity-modulated radiosurgery. Journal of Applied Clinical Medical Physics 2007;8(4):54-65 https://doi.org/10.1120/jacmp.v8i4.2397
  6. Kayla N. Kielar, Ed Mok, Annie Hsu, Lei Wang and Gary Luxton : Verification of dosimetric accuracy on the TrueBeam STx: Rounded leaf effect of the high definition MLC. Medical Physics 2012;39(10):6360-6371 https://doi.org/10.1118/1.4752444
  7. Eric E. Klein, Joseph Hanley, John Bayouth, Fang-Fang Yin et al: Task Group 142 report: Quality assurance of medical accelerators. Medical Physics 2009;36(9):4197-4212 https://doi.org/10.1118/1.3190392
  8. Soon Sung Lee, Sang Hyoun Choi, Chul Kee Min, Woo Chul Kim, Young Hoon Ji, Seungwoo Park, Haijo Jung, Mi-Sook Kim, Hyung Jun Yoo, Kum Bae Kim:Assessment for the Utility of Treatment Plan QA System according to Dosimetric Leaf Gap in Multileaf Collimator, Medical Physics 2015;26(3):168-177
  9. Zygmanski Piotr, Florin Rosca, Dnyanesh Kadam, Friedlieb Lorenz, Adrian Nalichowski, Laurence Court, and Lee Chin:Determination of depth and field size dependence of multileafcollimator transmission in intensity?modulated radiation therapy beams. Journal of Applied Clinical Medical Physics 2007;8(4):2693
  10. Rangel Alejandra and Peter Dunscombe: Tolerances on MLC leaf position accuracy for IMRT delivery with a dynamic MLC. Medical Physics 2009;36(7): 3304-3309 https://doi.org/10.1118/1.3134244
  11. Neil A Campbell: The Effect of Maximum MLC Speed and Dose Rate Constraints on VMAT Plan Quality, Accuracy and Efficiency, Degree of Master of Science in Medical Physics, 2013
  12. Thomas LoSasso, Chen shou chui, and C. Clifton Ling: Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy, Medical Physics 1998;25:1921-1927
  13. Varian Medical Systems: dosimetric leaf gap Measurement Guide.
  14. K Joseph Maria Das, Karthick Raj Mani, Sa Yoganathan:Effect of Leaf Motion Calculator Parameters in Head and Neck IMRT, Medical Physics 2010;37(6):3222
  15. 임경달, 제영완, 윤일규, 이제희, 유숙현:동적다엽콜리메이터를 이용한 세기변조방사선 치료 시 선량분포상의 선량률 변화에 따른 효과 2012;12:1-10
  16. 강민규, 김성준, 신현수 김성규: 세기조절방사선치료에 서 선량율 변화에 따른 선량분포 특성, 의학물리, 2010;21:218-222
  17. Jonas D. Fontenot: Evaluation of a novel secondary check tool for intensity modulated radiotherapy treatment planning. Journal of Applied Clinical Medical Physics 2014;15(5):207-215 https://doi.org/10.1120/jacmp.v15i5.4990
  18. E Wasbo, H valen: Dosimetric discrepancies caused by differing MLC parameters for dynamic IMRT, PHYSICS IN MEDICINE AND BIOLOGY, 2008;53:405-415 https://doi.org/10.1088/0031-9155/53/2/008
  19. Xiangyang Mei1, Ian Nygren1 and J. Eduardo Villarreal-Barajas: On the use of the MLC dosimetric leaf gap as a quality control tool for accurate dynamic 세기 조절 방사선 치료 delivery, Medical Physics 2011;38(4):2246-2255 https://doi.org/10.1118/1.3567148
  20. Danielle Fraser, William Parker, Jan Seuntjens: Characterization of cylindrical ionization chambers for patient specific IMRT QA, ournal of Applied Clinical Medical Physics 2009;10(4):241-251 https://doi.org/10.1120/jacmp.v10i4.2923
  21. 김선영, 이두현, 조정근, 정도형, 김호식, 최계숙: 세기 조절 방사선치료의 환자 정도관리에서 다양한 이온전리함 볼륨이 정확도에 미치는 영향, 대한방사선치료학회지 2006;18(1):1-5