Abstract
Because of non-coplanar therapy with couch rotation in respiratory gated radiation therapy, the recognition of marker movement due to the change in the distance between the infrared camera and the marker due to the rotation of the couch is called RPM (Real-time The purpose of this paper is to evaluate the accuracy of motion reflections (baseline changes) of 2D gating configuration (two dot marker block) and 3D gating configuration (six dot marker block). The motion was measured by varying the couch angle in the clockwise and counterclockwise directions by $10^{\circ}$ in the 2D gating configuration. In the 3D gating configuration, the couch angle was changed by $10^{\circ}$ in the clockwise direction and compared with the baseline at the reference $0^{\circ}$. The reference amplitude was 1.173 to 1.165, the couch angle at $20^{\circ}$ was 1.132, and the couch angle at $1.0^{\circ}$ was 1.083. At $350^{\circ}$ counterclockwise, the reference amplitude was 1.168 to 1.157, the couch angle at $340^{\circ}$ was 1.124, and the couch angle at $330^{\circ}$ was 1.079. In this study, the phantom is used to quantitatively evaluate the value of the amplitude according to couch change.
호흡동조방사선치료에서 비동일 평면치료는 테이블의 회전이 있기 때문에 테이블의 회전에 따른 적외선 카메라와 마커 사이의 거리 변화로 인하여 마커 움직임의 인식을 Real-time Position Management 시스템의 two dot 마커와 six dot 마커의 움직임 반영의 정확성을 평가하고자 한다. 움직임을 평가는 two dot 마커에서 테이블 각도를 시계방향, 반시계방향으로 각 $10^{\circ}$씩 변화하여 측정하였으며, six dot 마커에서도 테이블 각도를 시계방향으로 각 $10^{\circ}$ 변화하여 기준 $0^{\circ}$에서의 기준과 비교하였을 때 기준 진폭 1.173에서 1.165, 테이블 각도 $20^{\circ}$에서는 1.132, 테이블 각도 $30^{\circ}$에서는 1.083의 진폭값으로 나타났다. 반시계방향으로 $350^{\circ}$에서는 기준 진폭 1.168에서 1.157, 테이블 각도 $340^{\circ}$에서는 1.124, 테이블 각도 $330^{\circ}$에서는 1.079 진폭값으로 나타났다. 본 연구에서는 팬톰을 이용하여 테이블 변화에 따른 진폭의 값을 정량적으로 평가한 점에 학술적 의미를 두고자 한다.