A Study on Compression Paddle Materials to Reduce Radiation Exposure Dose During Mammography; PC and PMMA and Carbon

유방촬영 시 피폭선량을 줄이기 위한 압박대 재질의 PC, PMMA, Carbon에 대한 연구

  • Received : 2015.05.11
  • Accepted : 2015.06.06
  • Published : 2015.06.30

Abstract

This study was designed the band material in order to reduce the exposure pressure of the breast and the material was measured of Radiolucent and radiation properties with a radiation materials of PC, PMMA, Carbon. Also the image quality by image analysis to obtain the following results are below: Unfors Xi dosimetry using radiation transmittance when the results of 8.353mGy is measured after removal of the cuff, the PC 6.308mGy, PMMA 6.223mGy, Carbon 7.218mGy were measured respectively. Semi-layer PC 0.375mmAl, PMMA 0.370mmAl, with Carbon 0.360mmAl Carbon, PC, PMMA was higher radiation properties and transparency in order. InLight / OSL NanoDotTM dosimeter was used in the cuff and then removed by placing the dosimeter measured results center on 1.143mGy, at the edge 12.56mGy, on the central PC 8.990mGy, at the edge 10.291mGy, PMMA center on 8.391mGy, the edge 9.654mGy, on Carbon center 9.581mGy, 11.313 mGy were measured at the edge of Carbon, PC, PMMA showed a high permeability in order. Image Pixel average J is then removed from the cuff 976.655, PC 831.032, PMMA 819.069, Carbon 897.118 Carbon, PC, PMMA was measured by high order.

유방의 피폭선량을 줄이기 위해 압박대의 재질을 연구하였으며, 재질로는 PC, PMMA, Carbon을 이용한 방사선 투과성 및 선질 등을 측정하였고, 영상을 통한 화질도 분석하여 다음과 같은 결론을 얻었다. Unfors Xi dosimetry 사용 시 방사선 투과율은 압박대를 제거 후 측정한 결과 8.353 mGy가 나왔으며, PC 6.308 mGy, PMMA 6.223 mGy, Carbon 7.218 mGy가 측정되었다. 반가층은 PC 0.375mmAl, PMMA 0.370mmAl, Carbon 0.360mmAl으로 Carbon, PC, PMMA 순으로 높은 선질과 투과도를 보였다. InLight/OSL NanoDotTM dosimeters 사용 시 압박대를 제거 후 선량계를 위치시켜 측정한 결과 중심부위에서 1.143 mGy, 가장자리에서 12.56 mGy, PC 중심부위에서 8.990 mGy, 가장자리에서 10.291 mGy, PMMA 중심부위에서 8.391 mGy, 가장자리에서 9.654 mGy, Carbon 중심부위에서 9.581 mGy, 가장자리에서 11.313 mGy가 측정되어 Carbon, PC, PMMA 순으로 높은 투과도를 보였다. Image J에서 Pixel 평균값은 압박대를 제거한 후 976.655, PC 831.032, PMMA 819.069, Carbon 897.118로 Carbon, PC, PMMA순으로 높게 측정되었다.

Keywords

References

  1. Sung-Soo Kim: Status and countermeasures of patient exposure during mammography jounal of health science & medical technology, 31(1), 118-121, 2005
  2. Cardenosa: 유방영상학(Breast Imaging), 한미의학, 2005
  3. E klund GW, Cardenos A: The art of mammographic position, RCNA 30(1), 21-53, 1992
  4. Dae-Cheol Kweon, Eun-Mi Lee, Sung-Man Hong, et al.: analysis of the compression force and thickness in the screening mammography korean institute of industrial engineer, 606-610, 2002
  5. 현춘수: 과학용어사전, 뉴턴코리아, 2010
  6. http://starletzzang.blog.me/120165887326
  7. CJ Hourdakis, ABoziari, E Koumbouli: The effect of a compression paddle on energy response, calibration and measurement with mammographic dosimeters using ionization chambers and solid-state detectors. Physics in Medicine and Biology, 54, 1047-1059, 2009 https://doi.org/10.1088/0031-9155/54/4/015
  8. Gray JE: OSL : Dosimetry for the Twenty-First Century. First RCM on Development of Procedures for in-vivo Dosimetry in Radiotherapy. IAEA, Trieste, 2005
  9. Yukihara EG, et al.: High-precision dosimetry for radiotherapy using the optically stimulated luminescence technique and thin dosimeters. Phys Med Biol, 50, 5619-5628, 2005 https://doi.org/10.1088/0031-9155/50/23/014
  10. Aznar MC, et al.: Real-time optical-fibre luminescence dosimetry for radiotherapy, Physical characteristics and applications in photon beams. Phys Med Biol 49, 1655-1669, 2004 https://doi.org/10.1088/0031-9155/49/9/005
  11. Andersen CE, Marckmann CJ, Aznar MC: An algorithm for real-time dosimetry in intensity-modulated radiation therapy using the radioluminescence signal from. Radiat Prot Dosim, 120, 7-13, 2006 https://doi.org/10.1093/rpd/nci600
  12. Jung-Mi Kim, Su-Dong Jun, Keum-Moon Baek, et al.: Characteristic Evaluation of Optically Stimulated Luminescent Dosimeter (OSLD) for Dosimetry. Koean Society for Radiation Therapy, 22(2), 123-129, 2010
  13. 김근형외 5명: 플라스틱 이론과 응용. 문운당, 2010
  14. 과학나눔연구회: 다양한 형태와 천의 얼굴을 가진 탄소재료 탄소섬유. 일진사, 2010
  15. Caroline E. Blane, MD, Mitchell M. Goodsitt, PhD, Jeffrey C. Grimm, et al.: New compression paddle for wire localization in mammography. Academic Radiology, 17(2), 142-145, 2010 https://doi.org/10.1016/j.acra.2009.09.005