Implementation and Evaluation of Chair-type ECG Monitoring System using Unconstraint Electrode

무구속 전극을 이용한 의자형 심전도 모니터링 시스템의 구현 및 평가

  • Received : 2015.03.30
  • Accepted : 2015.05.01
  • Published : 2015.04.30

Abstract

In this study, we implemented an unconstraint ECG monitoring system on a chair. Nowadays, modern people spend most of their time sitting on the chair. Therefore, it is necessary to have an unconstraint ECG monitoring system that can be used for a long time. The implemented system can perform measurement even with clothes on and it has great advantage on motion artifacts. A pair of unconstraint electrodes are placed on the back of the chair. Amplifier and filters are designed to remove motion artifacts, The baseline noise and power line noise are filtered and very low level of bio-signal is amplified to give a final measurement. Control unit and wireless transmission unit are implemented. Analog signal is converted into digital signal and transmits biological signal to the PC and the smart phone. Therefore continuous ECG monitoring in daily life is made possible. A comparison experiment between Ag / AgCl electrode and unconstraint electrode is conducted to evaluate the performance of the implemented system. As a result, we confirm our unconstraint system can be used for daily life ECG monitoring.

본 연구에서는 현대인들의 일상생활 중 대부분의 시간을 보내는 좌식생활 중에 무구속적으로 장시간 심장활동상태를 모니터링하기 위한 의자형 무구속 심전도 모니터링 시스템을 설계 및 구현하였다. 구현된 시스템은 옷을 입은 상태에서도 측정이 가능하며, 동잡음에도 비교적 강한 장점을 가진다. 이를 위하여 의자의 등받이 부분에 무구속 전극을 배치하였으며, 미세한 생체신호를 증폭하고, 측정 시 발생 할 수 있는 동잡음 및 기저선의 변화, 상용전원에 의한 전원잡음을 제거하기 위하여 증폭기 및 필터를 설계하였다. 또한 제어부와 무선전송부를 구현하여 아날로그 신호를 디지털 신호로 변환하고 생체신호를 PC와 스마트폰으로 전송함으로써 일상생활 중 지속적인 심장활동 상태의 모니터링이 가능하다. 구현된 시스템의 성능을 평가하기 위하여 Ag/AgCl 전극을 사용한 시스템과 비교 실험을 수행하였으며, 실험 결과 구현된 시스템의 무구속적으로 심장활동 상태 모니터링이 가능함을 확인하였다.

Keywords

References

  1. 지역별 건강수명의 형평성 분석과 정책과제, 한국보건사회연구원, 2011.
  2. 김창수, "헬스케어 서비스 및 실용화 동향 IoT 활용해다양한시스템으로시장확대", electronic science, 2014.
  3. Nemati, Ebrahim, M. Jamal Deen, and Tapas Mondal. "A wireless wearable ECG sensor for long-term applications." Communications Magazine, IEEE, vol. 50, no. 1, pp.36-43, 2012
  4. Yan, L., Yoo, J., Kim, B., & Yoo, H. J., "A 0.5-V 12-W wirelessly powered patch-type healthcare sensor for wearable body sensor network", Solid-State Circuits, IEEE Journal of, vol. 45, no. 11, pp.2356-2365, 2010.
  5. A. Arcelus, M. Sardar and A. Mihailidis, "Design of a capacitive ECG sensor for unobtrusive heart rate measurements", IEEE International International Instrumentation and Measurement Technology Conference, pp. 407-410, 2013.
  6. S. Junnila, A. Akhbardeh, L. C. Barna, I. Defee, and A. Varri, "A wireless ballistocardiographic chair," in Engineering in Medicine and Biology Society, 2006. EMBS'06. 28th Annual International Conference of the IEEE, 2006, pp. 5932-5935.
  7. J. Alametsa, E. Rauhala, E. Huupponen, A. Saastamoinen, A. Varri, A. Joutsen, et al., "Automatic detection of spiking events in EMFi sheet during sleep" Medical engineering & physics, vol. 28, pp. 267-275, 2006. https://doi.org/10.1016/j.medengphy.2005.07.008
  8. 이재호, 이영재, 임민규, 박희정, 이정환, "이용량성 결합(Capacitive coupling)을 이용한 비접촉 생체전위 측정 전극 개발", 대한전기학회, 2013.
  9. B. B. Winter and J. G. Webster, "Driven-right-leg circuit design", IEEE Trans. Biomed. Eng., vol. BME-30, pp. 62-66, 1983. https://doi.org/10.1109/TBME.1983.325168