DOI QR코드

DOI QR Code

마이크로 도시메트리용 다분할 조직등가비례계수기의 개발과 특성 평가

Development and Characterization of Multi-Segmented Tissue Equivalent Proportional Counter for Microdosimetry

  • 투고 : 2015.03.11
  • 심사 : 2015.03.31
  • 발행 : 2015.03.31

초록

We designed, developed and characterized a multi-segmented tissue equivalent proportional (TEPC) counter for microdosimetry. The energy resolution of the multi-segmented TEPC was about 12% for $^{241}Am$ 5.45 MeV alpha particles. The resolution was better than 33% for a single un-segmented TEPC. A compact and low power consumption TEPC could be made by using digital pulse processor (DPP). We also successfully calibrated the TEPC by using $^{252}Cf$ standard neutron source in Korea Research Institute of Standards and Science (KRISS). According to the results, the TEPC is useful for several application of radiation monitoring such as a neutron monitor, air crew monitor and space dosimeter.

키워드

참고문헌

  1. ICRU Report 36, Microdosimetry, ICRU, pp.1-6, 1983.
  2. Rossi H.H. and Rosenzweig W., "A device for the measurement of dose as a function of specific ionization", Radiology, Vol. 64, pp. 404-410, 1955. https://doi.org/10.1148/64.3.404
  3. D. Moro, P. Colautti, M. Lollo, J. Esposito, V. Conte, L. De Nardo, A. Ferretti, and C. Ceballos, "BNCT dosimetry performed with a mini twin tissue-equivalent proportional counters (TEPC)", Appl. Radiat. Isotopes, Vol. 67, pp.171-174, 2009. https://doi.org/10.1016/j.apradiso.2009.03.042
  4. P. Beck, M. Latocha, S. Rollet, G. Stehno, "TEPC reference measurements at aircraft altitudes during a solar storm", Adv. Space Res., Vol. 36, pp. 1627-1633, 2005. https://doi.org/10.1016/j.asr.2005.05.035
  5. D. Zhou, E. Semones, M. Weyland, S. Johnson, "Radiation measured with TEPC and CR-39 PNTDs in low earth orbit", Adv. Space Res., Vol. 40, pp. 1571-1574, 2007. https://doi.org/10.1016/j.asr.2006.12.006
  6. Lawrence Pinsky, Martin Kroupa, Son Hoang, John Idarraga, Clif Amberboy, Ryan Rios, Jessica Hauss et al, "Timepix-based radiation environment monitor measurements aboard the International Space Station", Nucl. Instrum. Methods in Phys. Res. Sec. A- Accel. Spectrom. Det. Assoc. Equip., Vol. 782, pp. 143-148, 2015. https://doi.org/10.1016/j.nima.2015.02.016
  7. Uk-Won Nam, Jaejin Lee, Jeonghyun Pyo ,Won-Kee Park, Bong-Kon Moon et al, "Measurement of Linear Energy Spectra for 135 MeV/u Carbon beams in HIMAC using prototype TEPC", J. Sensor Sci. & Tech., Vol. 23, No. 3, pp.197-201, 2014. https://doi.org/10.5369/JSST.2014.23.3.197
  8. http://geuz.org/gmsh/ (retrieved on Mar. 11, 2015).
  9. https://csc.fi/web/elmer/elmer (retrieved on Mar. 11, 2015).
  10. http://garfieldpp.web.cern.ch/garfieldpp/ (retrieved on Mar. 11, 2015).
  11. Uk-Won Nam, Jaejin Lee, Bong-Kon Moon, Jeonghyun Pyo ,Won-Kee Park, Tissue Equivalent Proportional Counter, 10-2014-6044789.
  12. http://www.amptek.com/products/dp5-digital-pulse-processor-and-mca/ (retrieved on Mar. 11, 2015).
  13. Jordanov VT, Knoll GF, Huber AC, Pantazis JA, "Digital techniques for real-time pulse shaping in radiation measurements", Nucl. Instrum. Methods Phys. Res. Sect. A-Accel. Spectrom Det. Assoc. Equip., Vol. 353, pp.261-264, 1994. https://doi.org/10.1016/0168-9002(94)91652-7
  14. Delia Perez Nunez, Design, construction and implementation of spherical tissue equivalent proportional counter, Texas A&M University, pp.48-52, 2008.
  15. David R. Farrar, Microdosimetric measurements of shielding effects for iron particles at 500 MeV/nucleon incident upon aluminum and polyethylene, Colorado State University, pp.14-16, 2007.

피인용 문헌

  1. LET Calibration of Fe 500 MeV/u Ions using SSNTD vol.25, pp.1, 2016, https://doi.org/10.5369/JSST.2016.25.1.41