DOI QR코드

DOI QR Code

Determination of Trace Amounts of Formaldehyde in Water Using High Performance Liquid Chromatography and Acetylacetone as a Derivative Reagent

아세틸아세톤 유도체화 시약과 HPLC를 이용한 미량 포름알데하이드 수질분석

  • Lee, Ki-Chang (Gyeongsangbuk-do Institute of Health and Environment) ;
  • Park, Jae-Hyung (Gyeongsangbuk-do Institute of Health and Environment) ;
  • Lee, Wontae (Department of Environmental Engineering, Kumoh National Institute of Technology)
  • 이기창 (경상북도보건환경연구원) ;
  • 박재형 (경상북도보건환경연구원) ;
  • 이원태 (금오공과대학교 환경공학전공)
  • Received : 2015.01.08
  • Accepted : 2015.02.11
  • Published : 2015.02.28

Abstract

A simple analytical method to quantify formaldehyde in water at lower levels (${\mu}g/L$) was developed using a high performance liquid chromatography (HPLC) and acetylacetone as a derivative reagent. Unlike conventional methods, no extraction and/or concentration were required. The derivative reagent was added into samples and reacted for 30 minutes at $80^{\circ}C$ prior to the analysis of formaldehyde using HPLC. The method detection limit and the limit of quantification for this method were 1.6 and $5.0{\mu}g/L$, respectively. This method also achieved high precision (0.6-3.0%) and accuracy (91.6-106.3%). The recovery rates for various environmental samples ranged from 92.0 to 115.2%.

수중에 존재하는 포름알데하이드(formaldehyde)를 ${\mu}g/L$ 단위 농도까지 정량할 수 있는 간단한 분석법을 개발하였다. 시료 전처리시 유도체화 시약으로 아세틸아세톤(acetylacetone)을 사용하였고, 기존 분석방법과 달리 추출 및 농축과정을 거치지 않는 것이 장점이다. 시료에 유도체화 시약을 첨가하고 $80^{\circ}C$에서 30분 반응시킨 후 HPLC를 이용하여 포름알데하이드를 분석하였다. 본 방법에 대한 정도보증 결과, 방법검출한계 및 정량한계는 각각 1.6, $5.0{\mu}g/L$로 기존 분석법에 비하여 낮았다. 정밀도 및 정확도는 각각 0.6~3.0%, 91.6~106.3%로 우수하였다. 또한 다양한 환경시료에 대한 회수율 검증에서도 92.0~115.2%로 양호하게 나타났다.

Keywords

References

  1. Yang, S. T. and Lee, H. S., "A study for determination of formaldehyde in water," J. Korean Soc. Water Environ., 16(2), 275-282(2000).
  2. Li, O., Oshima, M. and Motomizu, S., "Flow-injection spectrofluorometric determination of trace amounts of formaldehyde in water after derivatization with acetoacetanilide," Talanta., 72(5), 1675-1680(2007). https://doi.org/10.1016/j.talanta.2007.01.054
  3. Chae, H. J., Kim, H. K., Kim, S. K., Pyo, H. S. and Hong, J. K., "Analysis and risk assessment of formaldehyde in water from water purification plant in korea," Anal, Sci. Technol., 22(5), 386-394(2009).
  4. Jeong, D. H., Shin, J. S., Shin, K. S., Kim, J. H., Kim, Y. S. and Rhew, D. H., "A study on the determination of formaldehyde effluent limitation in the industrial wastewater," J. Environ. Impact Assess., 22(3), 203-217(2013). https://doi.org/10.14249/eia.2013.22.3.203
  5. Trenholm, R. A., Rosario-Ortiz, F. L. and Snyder, S. A., "Analysis of formaldehyde formation in wastewater using on-fiber derivatization-solid-phase microextraction-gas chromatography-mass spectrometry," J. Chromatogr. A, 1210(1), 25-29(2008). https://doi.org/10.1016/j.chroma.2008.09.044
  6. Shin, S. K. and Song, J. H., "Modeling and simulations of the removal of formaldehyde using silver nano-particles attached to granular activated carbon," J. Hazard. Mater., 194, 385-392(2011). https://doi.org/10.1016/j.jhazmat.2011.08.001
  7. Rivero, R. T. and Topiwala, V., "Quantitative determination of formaldehyde in cosmetics using a combined solid-phase microextraction-isotope dilution mass spectrometry method," J. Chromatogr. A, 1029(1), 217-222(2004). https://doi.org/10.1016/j.chroma.2003.12.054
  8. Shin, H. S. and Ahn, H. S., "The study on the measurement of formaldehyde in saliva and urine by GC-MS," Anal. Sci. Technol., 19(2), 149-154(2006).
  9. Iqbal, M. Z. and Novalin, S., "Analysis of formose sugar and formaldehyde by high-performance liquid chromatography," J. Chromatogr. A, 1216(26), 5116-5121(2009). https://doi.org/10.1016/j.chroma.2009.04.065
  10. Wang, T., Gao, X., Tong, J. and Chen, L., "Determination of formaldehyde in beer based on cloud point extraction using 2,4-dinitrophenylhydrazine as derivative reagent," Food Chem., 131(4), 1577-1582(2012). https://doi.org/10.1016/j.foodchem.2011.10.021
  11. Huang, H., Simon, X. H., Ip, H. S. and Yu, J. Z., "Determination of trace amounts of formaldehyde in acetone," Anal. Chim. Acta, 604(2), 134-138(2007). https://doi.org/10.1016/j.aca.2007.10.011
  12. Michels, J. J., "Improved measurement of formaldehyde in water-soluble polymers by high-performance liquid chromatography coupled with post-column reaction detection," J. Chromatogr. A, 914(1), 123-129(2001). https://doi.org/10.1016/S0021-9673(00)01267-X
  13. Moussavi, G., Bagheri, A. and Khavanin, A., "The investigation of degradation and mineralization of high concentrations of formaldehyde in an electro-Fenton process combined with the biodegradation," J. Hazard. Mater., 237-238, 147-152(2012). https://doi.org/10.1016/j.jhazmat.2012.08.022
  14. Moussavi, G., Yazdanbakhsh, A. and Heidarizada, M., "The removal of formaldehyde from concentrated synthetic wastewater using $O_3/MgO/H_2O_2$ process integrated with the biological treatment," J. Hazard. Mater., 171, 907-913(2009). https://doi.org/10.1016/j.jhazmat.2009.06.090
  15. Li, O., Sritharathikhum, P., Oshima, M. and Motomizu, S., "Development of novel detection reagent for simple and sensitive determination of trace amounts of formaldehyde and its application to flow injection spectrophotometric analysis," Anal. Chim. Acta, 612(2), 165-172(2008). https://doi.org/10.1016/j.aca.2008.02.028
  16. Kiba, N., Sun, L., Yokose, S., Kazue, M. T. and Suzuki, T. T., "Determination of nano-molar levels of formaldehyde in drinking water using flow-injection system with immobilized formaldehyde dehydrogenase after off-line solid-phase extraction," Anal. Chim. Acta, 378(1), 169-175(1999). https://doi.org/10.1016/S0003-2670(98)00636-9
  17. Bruyn, W. J., Clark, C. D., Pagel, L. and Takehara, C., "Photochemical production of formaldehyde, acetaldehyde and acetone from chromophoric dissolved organic matter in coastal waters," J. Photochem. Photobiol. A: Chem., 226(1), 16-22(2011). https://doi.org/10.1016/j.jphotochem.2011.10.002
  18. Lo, K. M. and Yung, Y. L., "Integration of headspace solid phase micro-extraction with gas chromatography for quantitative analysis of formaldehyde," Bull. Korean Chem. Soc., 34(1), 139-142(2012). https://doi.org/10.5012/bkcs.2013.34.1.139

Cited by