Determination of Optimal Scan Time for the Measurement of Downstream Metabolites in Hyperpolarized 13C MRSI

  • Lee, Hansol (Department of Electrical & Electronic Engineering, Yonsei University) ;
  • Lee, Joonsung (Center for Neuroscience Imaging Research, Institute for Basic Science, Sungkyunkwan University) ;
  • Joe, Eunhae (Department of Electrical & Electronic Engineering, Yonsei University) ;
  • Yang, Seungwook (Department of Electrical & Electronic Engineering, Yonsei University) ;
  • Choi, Young-suk (Department of Radiology, Yonsei University College of Medicine) ;
  • Wang, Eunkyung (Department of Radiology, Yonsei University College of Medicine) ;
  • Song, Ho-Taek (Department of Radiology, Yonsei University College of Medicine) ;
  • Kim, Dong-Hyun (Department of Electrical & Electronic Engineering, Yonsei University)
  • Received : 2015.06.23
  • Accepted : 2015.09.10
  • Published : 2015.12.31

Abstract

Purpose: For a single time-point hyperpolarized $^{13}C$ magnetic resonance spectroscopy imaging (MRSI) of animal models, scan-time window after injecting substrates is critical in terms of signal-to-noise ratio (SNR) of downstream metabolites. Pre-scans of time-resolved magnetic resonance spectroscopy (MRS) can be performed to determine the scan-time window. In this study, based on two-site exchange model, protocol-specific simulation approaches were developed for $^{13}C$ MRSI and the optimal scan-time window was determined to maximize the SNR of downstream metabolites. Materials and Methods: The arterial input function and conversion rate constant from injected substrates (pyruvate) to downstream metabolite (lactate) were precalibrated, based on pre-scans of time-resolved MRS. MRSI was simulated using two-site exchange model with considerations of scan parameters of MRSI. Optimal scan-time window for mapping lactate was chosen from simulated lactate intensity maps. The performance was validated by multiple in vivo experiments of BALB/C nude mice with MDA-MB-231 breast tumor cells. As a comparison, MRSI were performed with other scan-time windows simply chosen from the lactate signal intensities of pre-scan time-resolved MRS. Results: The optimal scan timing for our animal models was determined by simulation, and was found to be 15 s after injection of the pyruvate. Compared to the simple approach, we observed that the lactate peak signal to noise ratio (PSNR) was increased by 230%. Conclusion: Optimal scan timing to measure downstream metabolites using hyperpolarized $^{13}C$ MRSI can be determined by the proposed protocol-specific simulation approaches.

Keywords

References

  1. Golman K, Zandt RI, Lerche M, Pehrson R, Ardenkjaer-Larsen JH. Metabolic imaging by hyperpolarized $^{13}C$ magnetic resonance imaging for in vivo tumor diagnosis. Cancer Res 2006;66:10855-10860 https://doi.org/10.1158/0008-5472.CAN-06-2564
  2. Day SE, Kettunen MI, Gallagher FA, et al. Detecting tumor response to treatment using hyperpolarized 13C magnetic resonance imaging and spectroscopy. Nat Med 2007;13:1382-1387 https://doi.org/10.1038/nm1650
  3. Ardenkjaer-Larsen JH, Fridlund B, Gram A, et al. Increase in signal-to-noise ratio of > 10,000 times in liquid-state NMR. Proc Natl Acad Sci U S A 2003;100:10158-10163 https://doi.org/10.1073/pnas.1733835100
  4. Golman K, Zandt RI, Thaning M. Real-time metabolic imaging. Proc Natl Acad Sci U S A 2006;103:11270-11275 https://doi.org/10.1073/pnas.0601319103
  5. Darpolor MM, Yen YF, Chua MS, et al. In vivo MRSI of hyperpolarized [1-(13)C]pyruvate metabolism in rat hepatocellular carcinoma. NMR Biomed 2011;24:506-513 https://doi.org/10.1002/nbm.1616
  6. Harris T, Eliyahu G, Frydman L, Degani H. Kinetics of hyperpolarized $^{13}C1$-pyruvate transport and metabolism in living human breast cancer cells. Proc Natl Acad Sci U S A 2009;106:18131-18136 https://doi.org/10.1073/pnas.0909049106
  7. Lau AZ, Chen AP, Ghugre NR, et al. Rapid multislice imaging of hyperpolarized $^{13}C$ pyruvate and bicarbonate in the heart. Magn Reson Med 2010;64:1323-1331 https://doi.org/10.1002/mrm.22525
  8. Schroeder MA, Cochlin LE, Heather LC, Clarke K, Radda GK, Tyler DJ. In vivo assessment of pyruvate dehydrogenase flux in the heart using hyperpolarized carbon-13 magnetic resonance. Proc Natl Acad Sci U S A 2008;105:12051-12056 https://doi.org/10.1073/pnas.0805953105
  9. Kohler SJ, Yen Y, Wolber J, et al. In vivo 13 carbon metabolic imaging at 3T with hyperpolarized $^{13}C$-1-pyruvate. Magn Reson Med 2007;58:65-69 https://doi.org/10.1002/mrm.21253
  10. Zierhut ML, Yen YF, Chen AP, et al. Kinetic modeling of hyperpolarized $^{13}C1$-pyruvate metabolism in normal rats and TRAMP mice. J Magn Reson 2010;202:85-92 https://doi.org/10.1016/j.jmr.2009.10.003

Cited by

  1. Alternating Acquisition Technique for Quantification of in vitro Hyperpolarized [1-13C] Pyruvate Metabolism vol.20, pp.1, 2015, https://doi.org/10.13104/imri.2016.20.1.53