Evaluation of Image Quality for Various Electronic Portal Imaging Devices in Radiation Therapy

방사선치료의 다양한 EPID 영상 질평가

  • Son, Soon-Yong (Department of Radiology, Asan Medical Center) ;
  • Choi, Kwan-Woo (Department of Radiology, Asan Medical Center) ;
  • Kim, Jung-Min (Department of College of Health Science, Radiologic Science, Korea University) ;
  • Jeong, Hoi-Woun (Department of Radiological Technology, Baekseok Culture University) ;
  • Kwon, Kyung-Tae (Department of Radiological Technology, Dongnam Health University) ;
  • Cho, Jeong-Hee (Department of Radiation Oncology, Yonsei Medical Center) ;
  • Lee, Jea-Hee (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Jung, Jae-Yong (Department of Radiation Oncology, Sanggye Paik Hospital) ;
  • Kim, Ki-Won (Department of Radiology, Kyung Hee University Hospital at Gang-dong) ;
  • Lee, Young-Ah (Department of Bio-Technologist and Laboratory Animal, Shingu University College) ;
  • Son, Jin-Hyun (Department of Radiological Technology, Shingu University College) ;
  • Min, Jung-Whan (Department of Radiological Technology, Shingu University College)
  • 손순룡 (서울아산병원 영상의학과) ;
  • 최관우 (서울아산병원 영상의학과) ;
  • 김정민 (고려대학교 방사선학과) ;
  • 정회원 (백석문화대학교 방사선과) ;
  • 권경태 (동남보건대학교 방사선과) ;
  • 조정희 (연세대학교 신촌세브란스 종양학과) ;
  • 이제희 (서울대학교병원 종양학과) ;
  • 정재용 (상계백병원 종양학과) ;
  • 김기원 (강동경희대병원 영상의학과) ;
  • 이영아 (신구대학교 바이오 동물학과) ;
  • 손진현 (신구대학교 방사선과) ;
  • 민정환 (신구대학교 방사선과)
  • Received : 2015.10.31
  • Accepted : 2015.12.21
  • Published : 2015.12.31

Abstract

In megavoltage (MV) radiotherapy, delivering the dose to the target volume is important while protecting the surrounding normal tissue. The purpose of this study was to evaluate the modulation transfer function (MTF), the noise power spectrum (NPS), and the detective quantum efficiency (DQE) using an edge block in megavoltage X-ray imaging (MVI). We used an edge block, which consists of tungsten with dimensions of 19 (thickness) ${\times}$ 10 (length) ${\times}$ 1 (width) $cm^3$ and measured the pre-sampling MTF at 6 MV energy. Various radiation therapy (RT) devices such as TrueBeam$^{TM}$ (Varian), BEAMVIEW$^{PLUS}$ (Siemens), iViewGT (Elekta) and Clinac$^{(R)}$iX (Varian) were used. As for MTF results, TrueBeam$^{TM}$(Varian) flattening filter free(FFF) showed the highest values of $0.46mm^{-1}$ and $1.40mm^{-1}$ for MTF 0.5 and 0.1. In NPS, iViewGT (Elekta) showed the lowest noise distribution. In DQE, iViewGT (Elekta) showed the best efficiency at a peak DQE and $1mm^{-1}DQE$ of 0.0026 and 0.00014, respectively. This study could be used not only for traditional QA imaging but also for quantitative MTF, NPS, and DQE measurement for development of an electronic portal imaging device (EPID).

MV방사선 치료는 둘러싸여 있는 정상조직의 피폭선량을 최소화 하면서, target volume 내에 정확하게 선량을 전달하는데 있어 중요한 요인이다. 본 연구에서는 방사선 치료의 높은 정확성을 유지하기 위하여 megavoltage X-ray imaging (MVI)에서 edge block 을 사용한 digital radiography (DR) system 검출기의 modulation transfer function (MTF: 변조전달함수), the noise power spectrum (NPS: 잡음전력스펙트럼) and the detective quantum efficiency (DQE: 양자검출효율)를 측정하고자 한다. 우리는 텅스텐으로 구성된 19 (thickness) ${\times}$ 10 (length) ${\times}$ 1 (width) $cm^3$의 edge block을 사용하였으며, 다음과 같은 setting들로 pre-sampling modulation transfer function (MTF)를 계산하였다: 6-megavolt (MV) energy를 사용하고, 다양한 Radiotherapy장비인 TrueBeamTM (Varian), BEAMVIEWPLUS (Siemens), iViewGT (Elekta), ClinacR iX (Varian) 를 사용하였다. MTF결과에서 Varian TrueBeamTM flattening filter free가 MTF의 50% ($mm^{-1}$)에서 0.46, 10% ($mm^{-1}$) 에서 1.40로 가장 highest value를 보였다. Noise 분포는 Elekta iViewGT가 가장 낮은 분포를 보였다. DQE에서는 E lekt a iViewGT가 peak DQE에서 0.0026 그리고 $1mm^{-1}$ DQE 에서 0.00014로 가장 높았다. 본 연구는 Edge method를 이용하여 MTF와 DQE산출을 재현하였으며, 현재 임상에서 사용되는 DR 시스템 측정의 높은 정확성을 유지할 수 있었으며 이러한 연구는 전통적인 QA 영상화뿐만 아니라 검출기 개발 연구에 있어서 정량적인 MTF, NPS, DQE 측정에 더욱 더 효율적으로 사용될 수 있다는 것을 알 수 있다.

Keywords

References

  1. Samei E, Flynn MJ. An experimental comparison of detector performance for direct and indirect digital radiography systems. Med Phys, 30, 608, 2003 https://doi.org/10.1118/1.1561285
  2. Kim HK, Cho G, Lee SW, Shin YH, Cho HS. Development and evaluation of a digital radiographic system based on CMOS image sensor. IEEE Trans Nucl Sci, 3, 662, 2001
  3. Jung-Whan Min, Ki-Won Kim, Jung-Min Kim et al: Evaluation of image quality by using a tungsten edge block in a megavoltage (MV) X-ray imaging. Korean Journal of Medical Physics, 23(3), 154, 2012
  4. Hoi-Woun Jeong, Jung-Whan Min, Jung-Min Kim et al: Performance characteristic of a Cs I (TI) Flat panel detector radiography system. Journal of Radiological Science and Technology, 35(2), 109, 2012
  5. White JE, Chen T, McCracken J, Kennedy P, Seydel HG, Hartman G, et al. The influence of radiation therapy quality control on survival, response and sites of relapse in oat cell carcinoma of the lung: preliminary report of a Southwest Oncology Group study. Cancer, 50, 1084, 1982 https://doi.org/10.1002/1097-0142(19820915)50:6<1084::AID-CNCR2820500611>3.0.CO;2-W
  6. Goitein M, Busse J. Immobilization errors: Some theoretical considerations. Radiology, 117, 407, 1975 https://doi.org/10.1148/117.2.407
  7. Marks JE, Haus AG, Sutton GH, Griem ML. Localization error in the radiotherapy of Hodgkin's disease and malignant lymphoma with extended mantle fields. Cancer, 34, 83, 1974 https://doi.org/10.1002/1097-0142(197407)34:1<83::AID-CNCR2820340112>3.0.CO;2-P
  8. Marks JE, Haus AG, Sutton GH, Griem ML. The value of frequent treatment verification films in reducing localization error in the irradiation of complex fields. Cancer, 37, 2755, 1976 https://doi.org/10.1002/1097-0142(197606)37:6<2755::AID-CNCR2820370628>3.0.CO;2-J
  9. Gopal A, Samant SS. Validity of the line-pair bar-pattern method in the measurement of the modulation transfer function (MTF) in megavoltage imaging. Med Phys, 35, 270, 2008 https://doi.org/10.1118/1.2816108
  10. Min JW, Son JH, Jeong HW, Kim JM, Seoung YH, Son SY, et al. Comparison of Noise Power Spectrum Methodologies in Measurements by Using Megavoltage X-ray Energies. J Korean Phys Soc, 60, 129, 2012 https://doi.org/10.3938/jkps.60.129
  11. Giger ML, Doi K. Investigation of basic imaging properties in digital radiography. 1. Modulation transfer function. Med Phys, 287, 11, 1984
  12. Fujita H, Tsai DY, Itoh T, Doi K, Morishita J, Ueda K, et al. A simple method for determining the modulation transfer function in digital radiography. IEEE Trans Med Imaging, 11, 34, 1992 https://doi.org/10.1109/42.126908
  13. Samei E, Flynn MJ, Reimann DA. A method for measuring the presampled MTF of digital radiographic systems using an edge test device. Med Phys, 25, 102, 1998 https://doi.org/10.1118/1.598165
  14. Greer PB, van Doorn T. Evaluation of an algorithm for the assessment of the MTF using an edge method. Med Phys, 27, 2048, 2000 https://doi.org/10.1118/1.1288682
  15. Cunningham A, Reid BK. Signal and noise in modulation transfer function determinations using the slit, wire and edge techniques. Med Phys, 19, 1037, 1992 https://doi.org/10.1118/1.596821
  16. Hillen W, Schiebel U, Zaengel T. Imaging performance of a digital storage phosphor system. Med Phys, 14, 744, 1987 https://doi.org/10.1118/1.596127
  17. Rogge F, Bosmans H, Marchal G. A practical method for detected quantum efficiency (DQE) assessment of digital mammography systems in the radiological environment. Proc SPIE, 4682, 645, 2002
  18. Cremers F, Frenzel T, Kausch C, Albers D, Schonborn T, Schmidt R. Performance of electronic portal imaging devices (EPIDs) used in radiotherapy: image quality and dose measurements. Med Phys, 31, 985, 2004 https://doi.org/10.1118/1.1688212
  19. Siewerdsen H, Antunuk LE, El-Mohri Y, Yorkston J, Huang W, Boudry JM, et al. Empirical and theoretical investigation of the noise performance of indirect detection, active matrix flat-panel imagers (AMFPIs) for diagnostic radiology. Med Phys, 24, 71, 1997 https://doi.org/10.1118/1.597919
  20. Min JW, Kim KW, Seoung YH, Kim JM, Choi IS, Jeong HW, et al. Performance of an Edge Block Used in a Configuration Detector: Image Quality Measurements. J Korean Phys Soc, 64, 732, 2014 https://doi.org/10.3938/jkps.64.732
  21. Min JW, Kim KW, Kim JM, Choi IS, Jeong HW, Son SY, et al. Evaluation of the Modulation Transfer Function of Megavoltage X-rays. J Korean Phys Soc, 65, 1969, 2014 https://doi.org/10.3938/jkps.65.1969
  22. Son SY, Choe BY, Lee JW, Kim JM, Jeong HW, Min JW, et al. Evaluation of an Edge Method for Computed Radiography and an Electronic Portal Imaging Device in Radiotherapy: Image Quality Measurements. J Korean Phys Soc, 65, 1976, 2014 https://doi.org/10.3938/jkps.65.1976
  23. Andrew G, Amit J, Daniel RB, Kenneth RH, Stephen R. Accurate MTF measurement in digital radiography using noise response. Med Phys. 37, 724, 2010 https://doi.org/10.1118/1.3284376
  24. Gopal A, Samant SS. Use of a line-pair resolution phantom for comprehensive quality assurance of electronic portal imaging devices based on fundamental imaging metrics. Med Phys, 36, 2006, 2009 https://doi.org/10.1118/1.3099559