Radiation Exposure in Coronary Angiography: A Comparison of Cineangiography and Fluorography

  • Hwang, Jongmin (Cardiovascular Center, Pusan National University Yangsan Hospital, Division of Cardiology, Department of Internal Medicine, Pusan National University School of Medicine) ;
  • Lee, Soo Yong (Cardiovascular Center, Pusan National University Yangsan Hospital, Division of Cardiology, Department of Internal Medicine, Pusan National University School of Medicine) ;
  • Chon, Min Ku (Cardiovascular Center, Pusan National University Yangsan Hospital, Division of Cardiology, Department of Internal Medicine, Pusan National University School of Medicine) ;
  • Lee, Sang Hyun (Cardiovascular Center, Pusan National University Yangsan Hospital, Division of Cardiology, Department of Internal Medicine, Pusan National University School of Medicine) ;
  • Hwang, Ki Won (Cardiovascular Center, Pusan National University Yangsan Hospital, Division of Cardiology, Department of Internal Medicine, Pusan National University School of Medicine) ;
  • Kim, Jeong Su (Cardiovascular Center, Pusan National University Yangsan Hospital, Division of Cardiology, Department of Internal Medicine, Pusan National University School of Medicine) ;
  • Park, Yong Hyun (Cardiovascular Center, Pusan National University Yangsan Hospital, Division of Cardiology, Department of Internal Medicine, Pusan National University School of Medicine) ;
  • Kim, June Hong (Cardiovascular Center, Pusan National University Yangsan Hospital, Division of Cardiology, Department of Internal Medicine, Pusan National University School of Medicine) ;
  • Chun, Kook Jin (Cardiovascular Center, Pusan National University Yangsan Hospital, Division of Cardiology, Department of Internal Medicine, Pusan National University School of Medicine)
  • Received : 2015.02.14
  • Accepted : 2015.06.16
  • Published : 2015.11.30

Abstract

Background and Objectives: Coronary angiography (CAG) is the gold standard for diagnosing coronary artery disease. However, exposure to ionizing radiation delivered during CAG has various negative biological effects on humans. In this study, there was an evaluation of whether fluorography resulted in decreased radiation exposure, as compared with cineangiography. Subjects and Methods: Fifty-five patients were prospectively enrolled and divided into two CAG groups, in accordance with the operator's professional discretion: a conventional cineangiography group versus a fluorography group. Fluorography refers to the photography of fluoroscopic images that are retrospectively stored, e.g., using the "Store fluoro" function of the Siemens cardiac angiography system. The primary outcomes included the air kinetic energy released per unit mass {air kerma (AK) mGy} and the dose (kerma)-area product (DAP; ${\mu}Gy{\cdot}m^2$), both measured using built-in software in the Siemens system. The secondary outcomes included the total procedure time and amount of contrast agent used with each CAG method. Results: The total AK and DAP were significantly lower in the fluorography group ($159.3{\pm}64.9mGy$ and $1337.9{\pm}629.6{\mu}Gy{\cdot}m^2$, respectively) than in the cineangiography group ($326.9{\pm}107.5mGy$ and $2341.1{\pm}849.9{\mu}Gy{\cdot}m^2$, respectively; p=0.000 for both). The total procedure time (cineangiography vs. fluorography, $12.8{\pm}4.7$ vs. $12.5{\pm}2.9min$; p=0.779) and contrast agent amount ($136.1{\pm}28.3$ vs. $126.3{\pm}25.7$, p=0.214) were comparable between the two groups. Conclusion: Fluorography is a useful method to decrease the radiation exposure in selected patients requiring CAG.

Keywords

Acknowledgement

Supported by : Pusan National University

References

  1. Nickoloff EL, Lu ZF, Dutta A, So J, Balter S, Moses J. Influence of flatpanel fluoroscopic equipment variables on cardiac radiation doses. Cardiovasc Intervent Radiol 2007;30:169-76. https://doi.org/10.1007/s00270-006-0096-6
  2. Smith-Bindman R, Miglioretti DL, Johnson E, et al. Use of diagnostic imaging studies and associated radiation exposure for patients enrolled in large integrated health care systems, 1996-2010. JAMA 2010;307:2400-9.
  3. Cousins C, Miller DL, Bernardi G, et al. ICRP PUBLICATION 120: radiological protection in cardiology. Ann ICRP 2013;42:1-125.
  4. Poll LW, Cohnen M, Brachten S, Ewen K, Mödder U. Dose reduction in multi-slice CT of the heart by use of ECG-controlled tube current modulation (“ECG pulsing”): phantom measurements. Rofo 2002;174:1500-5. https://doi.org/10.1055/s-2002-35945
  5. Delewi R, Hoebers LP, Råmunddal T, et al. Clinical and procedural characteristics associated with higher radiation exposure during percutaneous coronary interventions and coronary angiography. Circ Cardiovasc Interv 2013;6:501-6. https://doi.org/10.1161/CIRCINTERVENTIONS.113.000220
  6. Roguin A, Goldstein J, Bar O, Goldstein JA. Brain and neck tumors among physicians performing interventional procedures. Am J Cardiol 2013;111:1368-72. https://doi.org/10.1016/j.amjcard.2012.12.060
  7. Shah B, Mai X, Tummala L, et al. Effectiveness of fluorography versus cineangiography at reducing radiation exposure during diagnostic coronary angiography. Am J Cardiol 2014;113:1093-8. https://doi.org/10.1016/j.amjcard.2013.12.013
  8. Olcay A, Guler E, Karaca IO, et al. Comparison of fluoro and cine angiographic modes in coronary stenting procedure: a preliminary feasibility study. Int J Cardiol 2014;177:595-6. https://doi.org/10.1016/j.ijcard.2014.08.144