Repeatable calibration of Hounsfield units to mineral density and effect of scanning medium

  • Crookshank, Meghan (Biomedical Engineering, University of Toronto) ;
  • Ploeg, Heidi-Lynn (Human Mobility Research Centre, Queen's University & Kingston General Hospital) ;
  • Ellis, Randy (Human Mobility Research Centre, Queen's University & Kingston General Hospital) ;
  • MacIntyre, Norma J. (Human Mobility Research Centre, Queen's University & Kingston General Hospital)
  • Received : 2012.07.13
  • Accepted : 2013.02.13
  • Published : 2013.03.25

Abstract

Computed tomography (CT) is being utilized in orthopaedics and related research to estimate bone strength. These applications benefit from calibration of Hounsfield units to mineral density typical of long bone, up to $1750mg/cm^3$. This study describes a method for establishing repeatable calibration of Hounsfield units to density, and determines the effects of imaging medium on calibration accuracy. Four hydroxyapatite standards were imaged in air on 7 occasions over 19 weeks using a helical multi-slice CT scanner. Each standard was scanned 5 times in different media: porcine soft tissue, water, and air. Calibrated densities were highly repeatable (CV<3.5%). No difference in density was observed between water and soft tissue conditions (p>0.08). This work provides a model for determining repeatable scanner-specific density calibration, demonstrates that the linear relationship between Hounsfield units and density extends to values typical of cortical bone, and supports the practice of imaging calibration standards in an environment similar to that of the target bone.

Keywords

References

  1. Araki, K. and Okano, T. (2011), "The effect of surrounding conditions on pixel value of cone beam computed tomography", Clin. Oral Implants Res., doi: 10.1111/j.1600-0501.2011.02373.x.
  2. Au, A.G., Raso, V.J., Liggins, A.B. and Amirfazli, A. (2007), "Contribution of loading conditions and material properties to stress shielding near the tibial component of total knee replacements", J. Biomech., 40, 1410-1416. https://doi.org/10.1016/j.jbiomech.2006.05.020
  3. Bessho, M., Ohnishi, I., Matsuyama, J., Matsumoto, T., Imai, K. and Nakamura, K. (2007), "Prediction of strength and strain of the proximal femur by a CT-based finite element method", J. Biomech., 40, 1745-1753. https://doi.org/10.1016/j.jbiomech.2006.08.003
  4. Bouxsein, M.L. (2008), "Technology insight: noninvasive assessment of bone strength in osteoporosis", Nat. Clin. Pract. Rheumatol., 4, 310-318.
  5. Braillon, P.M. (2002), "Quantitative computed tomography precision and accuracy for long-term follow-up of bone mineral density measurements", J. Clin. Densitom., 5, 259-266. https://doi.org/10.1385/JCD:5:3:259
  6. Brooks, R.A. and DiChiro, G. (1976), "Beam hardening in reconstructive tomography", Phys. Med. Biol., 21, 390-398. https://doi.org/10.1088/0031-9155/21/3/004
  7. Curry, J.D. (1988), "The effect of porosity and mineral content on the Young's modulus of elasticity of compact bone", J. Biomech., 21, 131-139. https://doi.org/10.1016/0021-9290(88)90006-1
  8. Genant, H.K., Cann, C.E., Etinger, B. and Gordan, G. (1982), "Quantitative computed tomography of vertebral spongiosa: A sensitive method for detecting early bone loss after oophorectomy", Ann. Intern. Med., 97, 699-706. https://doi.org/10.7326/0003-4819-97-5-699
  9. Gong, H., Zhang, M., Fan, Y., Kwok, W.L. and Leung, P.C. (2012), "Relationships between femoral strength evaluated by nonlinear finite element analysis and BMD, material distribution and geometric morphology", Ann. Biomed. Eng., 40, 1575-1585. https://doi.org/10.1007/s10439-012-0514-7
  10. Kalendar, W.A. and Suess, C. (1987), "A new calibration phantom for quantified computed tomography", Med. Phys., 14, 863-866. https://doi.org/10.1118/1.596013
  11. Kalpakcioglu, B.B., Morshed, S., Engelke, K. and Genant, H.K. (2008) "Advanced imaging of bone macrostructure and microstructure in bone fragility and fracture repair", J. Bone Joint Surg. Am., 90, 68-78. https://doi.org/10.2106/JBJS.G.01506
  12. Kazakia, G.J. and Majumdar, S. (2006), "New imaging technologies in the diagnosis of osteoporosis", Rev. Endocr. Metab. Disord., 7, 67-74.
  13. Pietruszczak, S., Gdela, K., Webber, C.E. and Inglis, D. (2007), "On the assessment of brittle-elastic cortical bone fracture in the distal radius", Eng. Frac. Mech., 74, 1917-1927. https://doi.org/10.1016/j.engfracmech.2006.06.005
  14. Reggiani, B., Cristofolini, L., Varini, E. and Viceconti, M. (2006), "Predicting the subject-specific primary stability of cementless implants during pre-operative planning: preliminary validation of subject-specific finite-element models", J. Biomech., 40, 2552-2558.
  15. Schileo, E., Dall'Ara, E., Taddei, F., Malandrino, A., Schotkamp, T., Baleani, M. and Viceconti, M. (2008), "An accurate estimation of bone density improves the accuracy of subject-specific finite element models", J. Biomech., 41, 2483-2491. https://doi.org/10.1016/j.jbiomech.2008.05.017
  16. Steiger, P., Block, J.E., Steiger, S., Heuck, A.F., Friedlander, A., Ettinger, B., Harris, S.T., Gluer, C.C. and Genant, H.K. (1990), "Spinal bone mineral density measured with a quantitative CT: Effect of region of interest, vertebral level, and technique", Radiology, 175, 537-543. https://doi.org/10.1148/radiology.175.2.2326479
  17. Unnikrishnan, G.U. and Morgan, E.F. (2011), "A new material mapping procedure for quantitative computed tomography-based, continuum finite element analyses of the vertebra", J. Biomech. Eng., 133(7), 071001. https://doi.org/10.1115/1.4004190
  18. Van Gompel, G., Van Slambrouck, K., Defrise, M., Batenburg, KJ., de Mey, J., Sijbers, J. and Nuyts, J. (2011), "Iterative correction of beam hardening artifacts in CT", Med. Phys., 38, 36-49. https://doi.org/10.1118/1.3577758

Cited by

  1. Estimating the density of femoral head trabecular bone from hip fracture patients using computed tomography scan data vol.228, pp.6, 2014, https://doi.org/10.1177/0954411914540285
  2. Cortical bone trajectory and traditional trajectory—a radiological evaluation of screw-bone contact vol.157, pp.7, 2015, https://doi.org/10.1007/s00701-015-2432-6
  3. An Anisotropic Analysis of Human Femur Bone with Walking Posture: Experimental and Numerical Analysis pp.2191-1649, 2018, https://doi.org/10.1007/s12668-018-0560-1
  4. Differences in bone mineral density of trajectory between lumbar cortical and traditional pedicle screws vol.14, pp.None, 2014, https://doi.org/10.1186/s13018-019-1169-y
  5. Evaluation of patient tissue selection methods for deriving equivalent density calibration for femoral bone quantitative CT analyses vol.143, pp.None, 2021, https://doi.org/10.1016/j.bone.2020.115759