DOI QR코드

DOI QR Code

The Scattering Beam Measurement of the RBC and the Fabrication of the Micro Cell Biochip

적혈구의 산란빔 측정과 마이크로 세포 분석 바이오칩 제작

  • Byun, In Soo (Department of Biomedical Engineering, Yeungnam University) ;
  • Kwon, Ki Jin (Department of Biomedical Engineering, Yeungnam College of Science & Technology) ;
  • Lee, Joon Ha (Department of Biomedical Engineering, Yeungnam University)
  • Received : 2014.05.09
  • Accepted : 2014.06.02
  • Published : 2014.06.30

Abstract

Next future, The bio technology will be a rapidly developing. This paper is the scattering beam measurement of the red blood cell (RBC) and the fabrication of the micro cell biochip using the bio micro electro mechanical system (Bio-MEMS) process technology. The Major process method of Bio-MEMS technology was used the buffered oxide etchant (BOE), electro chemical discharge (ECD) and ultraviolet sensitive adhesives (UVSA). All experiments were the 10 times according to the process conditions. The experiment and research are required the ultraviolet expose, the micro fluid current, the cell control and the measurement of the output voltage Vpp (peak to peak) waveform by scattering angles. The transmitting and receiving of the laser beam was used the single mode optical fiber. The principles of the optical properties are as follows. The red blood cells were injected into the micro channel. The single mode optical fiber was inserting in the guide channel. The He-Ne laser beam was focusing in the single mode optical fiber. The transmission He-Ne laser beam is irradiating to the red blood cells. The manufactured guide channel consists of the four inputs and the four outputs. The red blood cell was allowed with the cylinder pump. The output voltage Vpp waveform of the scattering beam was measured with a photo detector. The receiving angle of the output optical fiber is $0^{\circ}$, $5^{\circ}$, $10^{\circ}$, $15^{\circ}$. The magnitude of the output voltage Vpp waveform was measured in the decrease according to increase of the reception angles. The difference of the output voltage Vpp waveform is due differences of the light transmittance of the red blood cells.

본 본문은 Bio-MEMS 공정으로 제작한 마이크로 세포 분석 바이오칩을 사용하여 적혈구의 광학적 특성을 전압으로 측정한 실험이다. Bio-MEMS 공정을 이용하여 세포의 원활한 이동과 측정 분석에 사용되는 글라스에 채널 패턴 에칭을 위하여 포토리소그래피(photolithography)와 산화완충식각(BOE: buffered oxide etchant) 공정 조건, 세포 분석과 정보 전달에 사용되는 광섬유의 에칭을 위하여 산화완충식각 공정 조건, 세포나 유체를 칩과 외부의 전달 등에 사용되는 글라스의 홀을 위하여 전기화학방전(ECD: electro chemical discharge) 공정 조건, 글라스 접합을 위한 자외선반응접합(UVSA: ultraviolet sensitive adhesives) 공정 조건을 정립하였다. 또한 유체나 세포의 흐름 제어를 위한 라미나 흐름 조건, 적혈구세포에 대한 산란빔 파형을 측정하였다. 적혈구 실험을 통하여 출력 광섬유의 각도에 따른 산란빔이 출력측의 광섬유각도가 $0^{\circ}$일 때 약 17 V, 각도가 $5^{\circ}$일 때 약 10 V, 각도가 $10^{\circ}$일 때 약 6 V, 각도가 $15^{\circ}$일 때 약 4 V의 전압(Vpp)으로 측정되었다. 따라서 마이크로 세포 분석 바이오칩 제작의 소형화, 단순화, 공정신간 단축, 정량화하였고 적혈구의 광학적 특성을 측정을 측정함으로써 의공학(biomedical), 바이오칩공학(biochip), 반도체공학(semiconductor), 생물정보학(bioinformatics) 등의 응용과학 분야 발전에 기여할 것으로 기대한다.

Keywords

References

  1. The Korean Federation of Science and Technology Societies. 2020 year bio economic opening. Science and technology 375:42-45, 1599-7340 (2000)
  2. Jing XM, Chen D, Fang DM, Huang C, Liu JQ: Multilayer microstructure fabrication by combining bulk silicon micromachining and UV-LIGA technology. Microelectronics J 38:120-124 (2007) https://doi.org/10.1016/j.mejo.2006.09.003
  3. Bu MI, T Melvin, GJ Ensell: A new masking technology for deep glass etching and its microfluidic application. Sensors and Actuators A 115:476-482 (2004) https://doi.org/10.1016/j.sna.2003.12.013
  4. Hu CY, Lo SL, Li CM, Kuan WH: Treating chemical mechanical polishing (CMP) wastewater by electro coagulation flotation process with surfactant. J Hazardous Mater 120:15-20 (2005) https://doi.org/10.1016/j.jhazmat.2004.12.038
  5. 정안목, 전의식, 김철호: A study on the metal-glass bonding using ultra- sonic. J the Semiconductor & Display Technology 10(2):June (2011)
  6. Zbigniew D, Paul R, Harry C: Flow cytometry. Academic press (2nd):part A-B (1995)
  7. David PS, Christopher TC, Stephen CJ, Michael R: Microchip flow cytometry using electro kinetic focusing. Analytical Chemistry 71(19):4173-4177 (1999) https://doi.org/10.1021/ac990372u
  8. Kim S, Williams RT, Lee HK, Midori M: Turner size characterization of magnetic cell sorting micro beads using flow field flow fractionation and photon correlation spectroscopy. J Magnetism and Magnetic Materials 194:248-253 (1999) https://doi.org/10.1016/S0304-8853(98)00551-4
  9. Daniel S, Young AM, Gray ML, Senturia SD: A micro fabricated flow chamber for optical measurements in fluids. IEEE: 219-224 (1993)
  10. Yoo DS, Sim KS: Track distribution of recoil protons in PN-3 dosimeters etched in NaOH solution. Korean Society of Medical Physics 2(2):129-139 (1991)
  11. Lee BY, Yoon HG, Hyun KS, Kwon YH, Yun IG: Investigation of manufacturing variations of planar InP/InGaAs avalanche photodiodes for optical receivers. Microelectronics J 35:635-640 (2004) https://doi.org/10.1016/j.mejo.2004.04.009
  12. Al-Mufti R, Hambley H, Albaiges G, Lees C, Nicolaides KH: Increased fetal erythroblast in women who subsequently develop preeclampsia. Hum Reprod 15:1624-1628 (2000) https://doi.org/10.1093/humrep/15.7.1624
  13. 신동오, 홍성언, 이병용, 이명자: The construction of solid state detector system using commercially available diode and its application. Korean Society of Medical Physics 1(1):91-95 (1990)
  14. Jian J, Lili C, Chengrong S, Fu XZ: A study of the effect of PTFE electret on fibroblast cell cycle with flow cytometry (FCM). Proceedings of 10th International Symposium on Electrets:171-173 (1999)