DOI QR코드

DOI QR Code

Research of Efficiency for Gas Electron Multiplier Detector to Monitor Low Energy Gamma-Ray and Beta-Ray

낮은 에너지 감마선과 베타선 모니터링을 위한 Gas Electron Multiplier 검출기의 효율성에 대한 연구

  • Lee, Soonhyouk (Medical Research Institute, Ewha Womans University) ;
  • Jung, Jae Hoon (Medical Research Institute, Ewha Womans University) ;
  • Lee, Rena (Department of Radiation Oncology, School of Medicine, Ewha Womans University)
  • 이순혁 (이화여자대학교 의과학연구소) ;
  • 정재훈 (이화여자대학교 의과학연구소) ;
  • 이레나 (이화여자대학교 의과대학 이대목동병원 방사선종양학교실)
  • Received : 2014.04.29
  • Accepted : 2014.05.13
  • Published : 2014.06.30

Abstract

Radiation monitoring is one of the most important process in all places where radioactive material is used including hospital. In this preliminary study, we made GAS electron multiplier (GEM) detector and acquired relative efficiencies in order to see if GEM detector can be useful in radiation monitoring system. The relative efficiency was acquired by using the ratio of GEM detector efficiency to CdTe detector efficiency. The relative efficiency of 72% and 4% was acquired for beta-ray and gamma-ray respectively.

병원을 포함해서서 방사선 물질을 사용하는 모든 사업장에서 방사선 물질의 모니터링은 방사선 안전에 가장 중요한 요소의 하나이다. 본 연구는 방사선 모니터링 시스템에 있어서 GEM 검출기의 활용 가능성을 알아보기 위한 선행 연구로서, GEM 검출기를 제작하고 CdTe 검출기와의 상대적 효율을 구한 결과, 베타와 감마선에 대한 평균 상대 효율이 각각 72%와 4%로서 매우 우수한 성능을 나타냄을 보였다.

Keywords

References

  1. 박인규: 방사선 검출기와 국제연구. 물리학과 첨단 기술. 25-30 (2011)
  2. Simon Frank: Commissioning of the GEM Detectors in the COMPAS Experiment. Diploma Thesis. Physik Department, Technische Universitat Munchen (2011)
  3. Buzulutskov A: Radiation detectors based on gas electron multipliers (Reviews). Instrum Exp Tech 50(3):287-310 (2007) https://doi.org/10.1134/S0020441207030013
  4. 곽정원, 신정욱, 김진성 등: 전립선암에 대한 양성자치료에서 금마커에 의한 방사선 선량분포의 영향. 의학물리 21(3):291-297 (2010)
  5. 전경수, 오영기, 백종근 등: 이온전리함 및 TLD 법을 이용한 multi-detector computed tomography의 흡수선량 및 체표면 선량 평가. 의학물리 24(1):35-40 (2013)
  6. Tsyganov Edward, Antich Peter, Parkey Robert, et al: Gas electron multiplying detectors for medical applications. Nucl Instru and Meth 597(2):257-265 (2008) https://doi.org/10.1016/j.nima.2008.09.002
  7. Fraga F, Margato L, Fetal S, et al: Luminescence and imaging with gas electron multipliers. Nucl Instru and Meth 513(1):379-387 (2003) https://doi.org/10.1016/j.nima.2003.08.066
  8. Fabio Sauli: Development and applications of the gas electron multiplier. Nucl Instru and Meth 513(1):115-119 (2001)
  9. Peskov V, Fonte R, Danielsson M, et al: The study and optimization of new micropattern gaseous detectors for high rate applications. IEEE Trans Nucl Sci 48(4) (2001)
  10. Yong HS, IG Kim, Hie HC, et al: Energy resolution of a single GEM detector. Journal of the Korean Physical Society 55(6):2366-2371 (2009) https://doi.org/10.3938/jkps.55.2366
  11. Lepy Marie-Christine: Detection efficiency, IAEA-ALMERA Technical Visit. (2010)

Cited by