Analysis of Output Constancy Checks Using Process Control Techniques in Linear Accelerators

선형가속기의 출력 특성에 대한 공정능력과 공정가능성을 이용한 통계적 분석

  • Oh, Se An (Department of Radiation Oncology, Yeungnam University Medical Center) ;
  • Yea, Ji Woon (Department of Radiation Oncology, Yeungnam University Medical Center) ;
  • Kim, Sang Won (Department of Radiation Oncology, Yeungnam University Medical Center) ;
  • Lee, Rena (Department of Radiation Oncology, School of Medicine, Ewha Womans University) ;
  • Kim, Sung Kyu (Department of Radiation Oncology, Yeungnam University Medical Center)
  • 오세안 (영남대학교 의료원 방사선종양학과) ;
  • 예지원 (영남대학교 의료원 방사선종양학과) ;
  • 김상원 (영남대학교 의료원 방사선종양학과) ;
  • 이레나 (이화여자대학교 의과대학 이대목동병원 방사선종양학교실) ;
  • 김성규 (영남대학교 의료원 방사선종양학과)
  • Received : 2014.04.29
  • Accepted : 2014.09.04
  • Published : 2014.09.30

Abstract

The purpose of this study is to evaluate the results for the quality assurance through a statistical analysis on the output characteristics of linear accelerators belonging to Yeungnam University Medical Center by using the Shewhart-type chart, Exponentially weighted moving average chart (EWMA) chart, and process capability indices $C_p$ and $C_{pk}$. To achieve this, we used the output values measured using respective treatment devices (21EX, 21EX-S, and Novalis Tx) by medical physicists every month from September, 2012 to April, 2014. The output characteristics of treatment devices followed the IAEA TRS-398 guidelines, and the measurements included photon beams of 6 MV, 10 MV, and 15 MV and electron beams of 4 MeV, 6 MeV, 9 MeV, 12 MeV, 16MeV, and 20 MeV. The statistical analysis was done for the output characteristics measured, and was corrected every month. The width of control limit of weighting factors and measurement values were calculated as ${\lambda}=0.10$ and L=2.703, respectively; and the process capability indices $C_p$ and $C_{pk}$ were greater than or equal to 1 for all energies of the linear accelerators (21EX, 21EX-S, and Novalis Tx). Measured values of output doses with drastic and minor changes were found through the Shewhart-type chart and EWMA chart, respectively. The process capability indices $C_p$ and $C_{pk}$ of the treatment devices in our institution were, respectively, 2.384 and 2.136 for 21EX, 1.917 and 1.682 for 21EX-S, and 2.895 and 2.473 for Novalis Tx, proving that Novalis Tx has the most stable and accurate output characteristics.

이 연구의 목적은 본원이 보유하고 있는 선형가속기들의 출력 특성을 Shewhart-type Chart, EWMA Chart, 공정능력지수 $C_p$$C_{pk}$을 이용한 통계적 분석으로 품질보증에 대한 결과를 평가하고자 한다. 측정값은 의학물리사에 의하여 2012년 9월부터 2014년 4월까지 매월 측정된 각각 치료기들(21EX, 21EX-S, Novalis Tx)의 출력측정값을 사용하였다. 치료기들의 출력 특성은 IAEA TRS-398의 가이드라인을 따랐으며, 측정 에너지는 광자선 6 MV, 10 MV, 15 MV와 전자선 4 MeV, 6 MeV, 9 MeV, 12 MeV, 16 MeV, 20 MeV였다. 매월 측정하여 교정한 출력특성에 대한 통계학적 분석이며, 가중인자와 측정값의 관리한계의 폭은 ${\lambda}=0.10$, L=2.703로 계산되었으며, 공정능력 $C_p$$C_{pk}$는 모든 선형가속기(21EX, 21EX-S, Novalis Tx)의 모든 에너지에서 1이상이었다. Shewhart-type Chart를 통하여 출력선량의 측정값의 큰 변화점을 찾을 수 있었고, EWMA Chart를 통하여 출력선량의 측정값의 미세한 변화점을 알아 볼 수 있었다. 본원의 치료기의 공정능력지수 $C_p$$C_{pk}$를 통하여 21EX가 2.384와 2.136, 21EX-S가 1.917과 1.682, Novalis Tx가 2.895와 2.473으로 Novalis Tx가 가장 안정적이고 정확한 출력특성을 나타내고 있었다.

Keywords

References

  1. Oh SA, Kang MK, Kim SK, Yea JW: Comparison of IMRT and VMAT Techniques in Spine Stereotactic Radiosurgery with International Spine Radiosurgery Consortium Consensus Guidelines. Prog. Med. Phys. 24(3):145-153 (2013) https://doi.org/10.14316/pmp.2013.24.3.145
  2. Andreo P, Burns D T, Hohlfeld K, et al: Absorbed dose determination in external beam radiotherapy: An international Code of Practice for dosimetry based on standards of absorbed dose to water IAEA Technical Report Series No 398 IAEA, Vienna (2000)
  3. Boyer A, Biggs P, Galvin J, et al: Basic applications of multileaf collimators. Report of the AAPM Radiation Therapy Committee Task Group No. 50. Med Phys (2001)
  4. ICRU Report 83: Prescribing, recording and reporting photon- beam intensity-modulated radiation therapy (IMRT). International Commission on Radiation Units and Measurements, 10:1-106 (2010)
  5. Sanghangthum T, Suriyapee S, Srisatit S, Pawlicki T: Retrospective analysis of linear accelerator output constancy checks using process control techniques. J Appl Clin Med Phys 14(1):147-60 92013) https://doi.org/10.1120/jacmp.v14i1.4032
  6. Ekaette E, Lee RC, Cooke DL, Iftody S, Craighead P: Probabilistic fault tree analysis of a radiation treatment system. Risk Analysis 27(6):1395-410 (2007) https://doi.org/10.1111/j.1539-6924.2007.00976.x
  7. Huq MS, Fraass BA, Dunscombe PB, et al: A method for evaluating quality assurance needs in radiation therapy. Int J Radiat Oncol Biol Phys 71(1 Suppl):S170-S173 (2008) https://doi.org/10.1016/j.ijrobp.2007.06.081
  8. Pawlicki T, Whitaker M, Boyer A: Statistical process control for radiotherapy quality assurance. Med Phys 32(9):2777-86 (2005) https://doi.org/10.1118/1.2001209
  9. Tennant R, Mohammed MA, Coleman JJ, Martin U: Monitoring patients using control charts: a systematic review. Int J Qual Health Care 19(4):187-94 (2007) https://doi.org/10.1093/intqhc/mzm015
  10. Nordstrom F, Wetterstedt S, Johnsson S, Ceberg C, Back SAJ: Control chart analysis of data from a multicenter monitor unit verification study. Radiother Oncol 102(3):364.70 (2012)
  11. Carey RG and Lloyd RC: Measuring quality improvement in healthcare: a guide to statistical process control application. Wisconsin, WI: Quality Press (2001)
  12. Klein EE, Hanley J, Bayouth J, et al: Task Group 142 report: quality assurance of medical accelerators. Med Phys 36(9):4197-212 (2009) https://doi.org/10.1118/1.3190392
  13. Grattan MW and Hounsell AR: Analysis of output trends from Varian 2100C/D and 600C/D accelerators. Phys Med Biol 56(1):N11-N19 (2011)
  14. Reynolds MR and Stoumbos ZG: Should exponentially weighted moving average and cumulative sum charts be used with Shewhart limits Technometrics 47(4):409-24 (2005) https://doi.org/10.1198/004017005000000382