DOI QR코드

DOI QR Code

Accuracy of Pedicle Screw Insertion Using Fluoroscopy-Based Navigation-Assisted Surgery : Computed Tomography Postoperative Assessment in 96 Consecutive Patients

  • Lee, Keong Duk (Department of Neurosurgery, Ulsan University Hospital) ;
  • Lyo, In Uk (Department of Neurosurgery, Ulsan University Hospital) ;
  • Kang, Byeong Seong (Department of Radiology, Ulsan University Hospital) ;
  • Sim, Hong Bo (Department of Neurosurgery, Ulsan University Hospital) ;
  • Kwon, Soon Chan (Department of Neurosurgery, Ulsan University Hospital) ;
  • Park, Eun Suk (Department of Neurosurgery, Ulsan University Hospital)
  • Received : 2014.01.23
  • Accepted : 2014.07.15
  • Published : 2014.07.28

Abstract

Objective : Two-dimensional fluoroscopy-based computerized navigation for the placement of pedicle screws offers the advantage of using stored patient-specific imaging data in providing real-time guidance during screw placement. The study aimed to describe the accuracy and reliability of a fluoroscopy-based navigation system for pedicle screw insertion. Methods : A total of 477 pedicle screws were inserted in the lower back of 96 consecutive patients between October 2007 and June 2012 using fluoroscopy-based computer-assisted surgery. The accuracy of screw placement was evaluated using a sophisticated computed tomography protocol. Results : Of the 477 pedicle screws, 461 (96.7%) were judged to be inserted correctly. Frank screw misplacement [16 screws (3.3%)] was observed in 15 patients. Of these, 8 were classified as minimally misplaced (${\leq}2mm$); 3, as moderately misplaced (2.1-4 mm); and 5, as severely misplaced (>4 mm). No complications, including nerve root injury, cerebrospinal fluid leakage, or internal organ injury, were observed in any of the patients. Conclusion : The accuracy of pedicle screw placement using a fluoroscopy-based computer navigation system was observed to be superior to that obtained with conventional techniques.

Keywords

References

  1. Castro WH, Halm H, Jerosch J, Malms J, Steinbeck J, Blasius S : Accuracy of pedicle screw placement in lumbar vertebrae. Spine (Phila Pa 1976) 21 : 1320-1324, 1996 https://doi.org/10.1097/00007632-199606010-00008
  2. Farber GL, Place HM, Mazur RA, Jones DE, Damiano TR : Accuracy of pedicle screw placement in lumbar fusions by plain radiographs and computed tomography. Spine (Phila Pa 1976) 20 : 1494-1499, 1995 https://doi.org/10.1097/00007632-199507000-00010
  3. Gertzbein SD, Robbins SE : Accuracy of pedicular screw placement in vivo. Spine (Phila Pa 1976) 15 : 11-14, 1990 https://doi.org/10.1097/00007632-199001000-00004
  4. Holly LT : Image-guided spinal surgery. Int J Med Robot 2 : 7-15, 2006 https://doi.org/10.1002/rcs.69
  5. Ju SM, Kim YS, Kim SB, Ko Y, Oh SH : Radiologic evaluation of proper pedicle screw placement after pedicle screw fixation in degenerative lumbar disc disease. J Korean Neurosurg Soc 38 : 265-268, 2005
  6. Kalfas IH, Kormos DW, Murphy MA, McKenzie RL, Barnett GH, Bell GR, et al. : Application of frameless stereotaxy to pedicle screw fixation of the spine. J Neurosurg 83 : 641-647, 1995 https://doi.org/10.3171/jns.1995.83.4.0641
  7. Kosmopoulos V, Schizas C : Pedicle screw placement accuracy : a meta-analysis. Spine (Phila Pa 1976) 32 : E111-E120, 2007 https://doi.org/10.1097/01.brs.0000254048.79024.8b
  8. Laine T, Lund T, Ylikoski M, Lohikoski J, Schlenzka D : Accuracy of pedicle screw insertion with and without computer assistance : a randomised controlled clinical study in 100 consecutive patients. Eur Spine J 9 : 235-240, 2000 https://doi.org/10.1007/s005860000146
  9. Laine T, Makitalo K, Schlenzka D, Tallroth K, Poussa M, Alho A : Accuracy of pedicle screw insertion : a prospective CT study in 30 low back patients. Eur Spine J 6 : 402-405, 1997 https://doi.org/10.1007/BF01834068
  10. Learch TJ, Massie JB, Pathria MN, Ahlgren BA, Garfin SR : Assessment of pedicle screw placement utilizing conventional radiography and computed tomography : a proposed systematic approach to improve accuracy of interpretation. Spine (Phila Pa 1976) 29 : 767-773, 2004 https://doi.org/10.1097/01.BRS.0000112071.69448.A1
  11. Lonstein JE, Denis F, Perra JH, Pinto MR, Smith MD, Winter RB : Complications associated with pedicle screws. J Bone Joint Surg Am 81 : 1519-1528, 1999 https://doi.org/10.2106/00004623-199911000-00003
  12. Merloz P, Tonetti J, Pittet L, Coulomb M, Lavallee S, Sautot P : Pedicle screw placement using image guided techniques. Clin Orthop Relat Res (354) : 39-48, 1998
  13. Oertel MF, Hobart J, Stein M, Schreiber V, Scharbrodt W : Clinical and methodological precision of spinal navigation assisted by 3D intraoperative O-arm radiographic imaging. J Neurosurg Spine 14 : 532-536, 2011 https://doi.org/10.3171/2010.10.SPINE091032
  14. Shin MH, Ryu KS, Park CK : Accuracy and safety in pedicle screw placement in the thoracic and lumbar spines : comparison study between conventional C-arm fluoroscopy and navigation coupled with O-arm$^{(R)}$ guided methods. J Korean Neurosurg Soc 52 : 204-209, 2012 https://doi.org/10.3340/jkns.2012.52.3.204
  15. Tian NF, Huang QS, Zhou P, Zhou Y, Wu RK, Lou Y, et al. : Pedicle screw insertion accuracy with different assisted methods : a systematic review and meta-analysis of comparative studies. Eur Spine J 20 : 846-859, 2011 https://doi.org/10.1007/s00586-010-1577-5
  16. Tian NF, Xu HZ : Image-guided pedicle screw insertion accuracy : a meta-analysis. Int Orthop 33 : 895-903, 2009 https://doi.org/10.1007/s00264-009-0792-3
  17. Vaccaro AR, Rizzolo SJ, Balderston RA, Allardyce TJ, Garfin SR, Dolinskas C, et al. : Placement of pedicle screws in the thoracic spine. Part II : An anatomical and radiographic assessment. J Bone Joint Surg Am 77 : 1200-1206, 1995 https://doi.org/10.2106/00004623-199508000-00009
  18. Van de Kelft E, Costa F, Van der Planken D, Schils F : A prospective multicenter registry on the accuracy of pedicle screw placement in the thoracic, lumbar, and sacral levels with the use of the O-arm imaging system and StealthStation Navigation. Spine (Phila Pa 1976) 37 : E1580-E1587, 2012 https://doi.org/10.1097/BRS.0b013e318271b1fa
  19. Weinstein JN, Spratt KF, Spengler D, Brick C, Reid S : Spinal pedicle fixation : reliability and validity of roentgenogram-based assessment and surgical factors on successful screw placement. Spine (Phila Pa 1976) 13 : 1012-1018, 1988 https://doi.org/10.1097/00007632-198809000-00008

Cited by

  1. Accuracy of Current Techniques for Placement of Pedicle Screws in the Spine: A Comprehensive Systematic Review and Meta-Analysis of 51,161 Screws vol.126, pp.None, 2019, https://doi.org/10.1016/j.wneu.2019.02.217