As the number of treatments in the therapeutic ultrasound field targeted at fat tissue increase, the performance of the equipment should be evaluated for safety using a fat phantom. In this study, a fat phantom was fabricated using olive oil and a tissue-mimicking material (TMM) phantom. To evaluate the acoustic properties of the TMM phantom according to the changes in the olive oil, the composition ratio of a liquid mixture of olive oil with a surfactant was adjusted from 5-20% in 5% steps. The acoustic properties of the phantom were evaluated using the sound velocity, attenuation coefficient, density, and acoustic impedance. The experimental results showed that the sound velocity decreased with increasing amount of olive oil but the other acoustic properties did not change. In addition, the phantom using an olive-oil mixture with a 15% composition ratio was most similar to the acoustic characteristics of fat tissue with a sound velocity of 1477.35 m/s, an attenuation coefficient of 0.514 dB/MHz-cm, a density of $1.07g/cm^3$, and an acoustic impedance of 1.575 MRayl. These experimental results are expected contribute to the accuracy of the results using a TMM phantom and will be useful for the therapeutic ultrasound field targeted at subcutaneous fat tissue.

Keywords

References

  1. Ing Gon Kim, and Jai Mann Lew, "Suction lipectomy and Ultrasonic liposuction," Diagnosis and Treatment, Vol. 227, pp. 994-997, 2000.
  2. Chan-woo Jeong, Nak-gyeong No, Soo-Hong Kim, Seong-jae Yun, Seok-beom Bak, "Practice of Facial Bone Contouring Surgery using Ultrasound-assisted Lipolysis and Autologous Microfat Graft", Korea Dermatological Association, Vol. 48, No. 10, 2010.
  3. James A. Zagzebski, "Essentials of Ultrasound Physics", Mosby Publication, 1996.
  4. Uwe Haberkorn, Gunter Layer, Volker Rudat, Ivan Zuna, Adolf Lorenz, and Gerhard van Kaick, "Ultrasound image properties influenced by abdominal wall thickness and composition," Journal of Clinical Ultrasound, Vol. 21, Issue 7, pages 423-429, September 1993. https://doi.org/10.1002/jcu.1870210704
  5. P. O. Kotzki, D. Buyck, D. Hans, E. Thomas, F. Bonnel, F. Favier, P. J. Meunier, and M. Rossi, "Influence of fat on ultrasound measurements of the Os calcis," Calcified Tissue International, Vol. 54, Issue 2, pp 91-95, 1994. https://doi.org/10.1007/BF00296057
  6. Martin O. Culjat, David Goldenberg, Priyamvada Tewari, and Rahul S. Singh, "A Review of Tissue Substitutes for Ultrasound Imaging," Ultrasound in Medicine & Biology, Vol. 36, No.6, pp. 861-873, 2010. https://doi.org/10.1016/j.ultrasmedbio.2010.02.012
  7. Kondo T., Kitatuji M., and Kanda H., "New tissue mimicking materials for ultrasound phantoms," Ultrasonics Symposium, 2005 IEEE, Vol. 3, pp. 1664-1667, September 2005.
  8. Ortega R., Leija L., and Vera A., "Measurement of ultrasonic properties of fat biological phantom," Health Care Exchanges, PAHCE 2009. Pan American, pp. 58-61, March 2009.
  9. Louise M. Cannon, Andrew J. Fagan, and Jacinta E. Browne, "Novel Tissue Mimicking Materials for High Frequency Breast Ultrasound Phantom," Ultrasound in Medicine & Biology, Vol. 37, No.1, pp. 122-135, 2011. https://doi.org/10.1016/j.ultrasmedbio.2010.10.005
  10. Sun Yongchen, Dong Yanwu, Tong Jie, and Tang Zhensheng, "Ultrasonic propagation parameters in human tissues," IEEE 1986 Ultrasonics Symposium, pp. 905-908, November 1986.
  11. IEC 60601-2-5 2009, Medical electrical equipment - Part 2-5: Particular requirements for the basic safety and essential performance of ultrasonic physiotherapy equipment.
  12. Scherzinger AL, Belgam RA, Carson PL, Meyer CR, Sutherland JV, Bookstein FL, and Silver TM, "Assessment of ultrasonic computed tomography in symptomatic breast patients by discriminant analysis," Ultrasound Medicine & Biology, Vol. 15, pp. 21-28, 1989.
  13. D'Astous F. T., and Foster F. S., "Frequencydependence of ultrasound attenuation and backscatter in breast-tissue," Ultrasound Medicine & Biology, Vol. 12, pp. 795-808, 1986. https://doi.org/10.1016/0301-5629(86)90077-3
  14. T. D. Mast, "Empirical relationships between acoustic parameters in human soft tissues," Acoustical Society of America, Vol. 1, pp. 37-42, 2000.
  15. J. Oudry, C. Bastard, V. Miette, R. Willinger, and L. Sandrin, "Copolymer-in-oil Phantom Materials for Elastography," Ultrasound Medicine & Biology, Vol. 35, pp. 1185-1197, 2009. https://doi.org/10.1016/j.ultrasmedbio.2009.01.012