A Study on Expiration Date on Ethylene Oxide Gas Sterilization Products - Related to Storage Environment -

Ethylene Oxide 가스 멸균물품의 유효기간에 관한 연구 - 보관환경을 중심으로 -

  • Son, Jeong Sook (Logistics Management Team, Seoul Metropolitan Government Boramae Medical Center, Nursing Graduate School, Chungang University) ;
  • Yu, Il Kyong (Seoul Metropolitan Government Boramae Medical Center, Logistics Management Team)
  • 손정숙 (서울특별시보라매병원, 물류자산팀, 중앙대학교 대학원) ;
  • 유일경 (서울특별시보라매병원, 물류자산팀 중앙공급파트)
  • Received : 2013.12.09
  • Accepted : 2014.03.25
  • Published : 2014.05.31

Abstract

Purpose: This was a study on the expiration date of Ethylene Oxide (EO) gas sterilization and effects of the environmental factors of temperature, humidity and type of cabinet in sterile goods storage area on the expiration date. Methods: Sterile goods storage areas from 13 departments in one hospital were selected and 455 EO gas sterilization samples were prepared and kept in those areas over the 14 months of the study. Each sample was tested with a microbiological culture in the laboratory every week. If the result was positive, the sample was regarded as contaminated. The researcher visited once a month to check the temperature, humidity and type of cabinet. Results: With the exception of 1 sample which was positive at 56th week. 454 samples were confirmed as negative. The environment of the samples storage area was measured monthly. The annual average temperature was $24.2{\pm}1.6^{\circ}C$, and the mean relative humidity $34.7{\pm}15.2%$. The types of cabinet were 7 open and 6 closed. Conclusion: The results of the microbiological culture at 13 months showed that none of the samples were contaminated. Therefore the hospital's existing Expiration Date can be extended from 6 months to 13~14 months.

Keywords

References

  1. Association for the Advancement of Medical Instrumentation. (1993). Good hospital practice: Steam sterilization and sterility assurance. Association for the Advancement of Medical Instrumentation. ST46-93.
  2. Association of Operating Room Nurses Standards. (1994). Implementation of an Event-Related Sterility Plan. Association of perioOperative Registered Nurses Journal, 60, 598-602.
  3. Association of Operating Room Nurses Standards. (1996). Proposed recommended practices for selection and use of packaging system. Association of perioOperative Registered Nurses Journal, 63, 949-953.
  4. Atkinson, L. J., & Fortunato, N. H. (1992). Berry & Kohn's Operating room technique. 8th Ed. St. Louis: Mosby-Year Book, Inc.
  5. Cho, S. S., & Kim, M. A. (1999). The measurement of the contamination period about the storage environment of ethylene oxide gas sterilized packs using in the operating room. Journal of Korean Academic Society of Adult Nursing, 11, 343-357.
  6. Choi, J. S., & Kim, K. S. (2004). A study for safe Storage time and Cost analysis for sterilized products. Korean Journal of Nosocomial Infection Control, 9, 131-138.
  7. Garner, J. S., & Favero, M. S. (1986). CDC Guideline for handwashing and hospital environmental control, 1985. Infection Control, 7, 231-243. https://doi.org/10.1017/S0195941700084022
  8. Jung, H. J., Kim, W. J., Kim, M. J., Park, S. C., & Cho, K. H 1995). Nosocomial infection surveillance in intensive care unit. Korean Journal of Infectious Diseases, 27, 105-117.
  9. Kim, G. S., Kim, G. J., Park, D. M., Park, M. O., & Choi, M. G. (2001). A study on validity periods of steam sterilization products- centering around packing materials and storage environments. Journal of Korean Clinical Nursing Research, 6, 73-90.
  10. Kim, Y. H., & Ham, G. J. (2005). Disinfection and sterilization. Seoul: Korea Medical Book Publisher.
  11. Korea Association of Central Supply Department Nurses. (2004). The principal of sterilization and practices. Seoul: Hyunmoon Publishing.
  12. Korean Society for Nosocomial Infection Control. (2011). Infection control and prevention in healthcare facilities, the 4th edition. Seoul: Hanmi Medical Publishing.
  13. Kyun, Y. H. (1996). A study on validity periods of sterilization products with concerning by storage environments. Korea Association of Operating Room Nurses, 4, 39-51.
  14. Lim, S. H. (2003). A study for contamination time for steam sterilized products, Unpublished master's thesis. Dongeui University, Busan.
  15. Mayworm, D. (1984). Sterile Shelf Life and Expiration Dating. Journal of Hospital Supply. Processing and Distribution, 2(6), 32-35.
  16. Schwartz, R. S., Bradley, D. V., Butt, W. E., & Mayhew, R. B. (1992). Safe storage times for sterile instrument packs. Military Medicine. 157, 406-408.
  17. Song, N. S. (2009). A contamination study for resetting of the validity terms of steam sterilized items in a general hospital. Unpublished master's thesis, Chungnam National University, Daejeon.
  18. Yoon, K. S. (1993). Research on central supply room disinfection of hospitals nationwide basis. The Book of Korea Association of Central Supply Department Nurses, 3-17.
  19. Yoon, K. S., Kim, J. H., Yang, S. H., Chae, J. Y., Lee, Y. M., & Cho, K. S. (1998). A study for safe storage time for in-house sterilized products in a korean hospital. Journal of Korean Society of Quality Assurance in Health Care, 5, 258-276.

Cited by

  1. Knowledge and Educational Need of Central Supply Room Nurses for the Sterilization Process vol.20, pp.2, 2015, https://doi.org/10.14192/kjnic.2015.20.2.70
  2. 포장재에 따른 멸균품의 유효기간에 관한 연구 vol.25, pp.3, 2014, https://doi.org/10.22650/jkcnr.2019.25.3.333