유방촬영용 X선 진단장치에서 IEC 60601-2-45 표준규격을 적용한 직선성 평가

Evaluation of Linearity Air Kerma Applied to the IEC 60601-2-45 Standard in Mammographic X-ray

  • 홍동희 (한서대학교 보건의료학과) ;
  • 정홍량 (극동대학교 방사선학과) ;
  • 임청환 (극동대학교 방사선학과) ;
  • 한범희 (서남대학교 방사선학과) ;
  • 한상현 (서남대학교 방사선학과) ;
  • 이상호 (서남대학교 방사선학과) ;
  • 모은희 (전북대학교 방사선과학기술학과) ;
  • 김기정 (극동대학교 방사선학과) ;
  • 이미화 (극동대학교 방사선학과) ;
  • 유인규 (극동대학교 방사선학과)
  • Hong, Dong-Hee (Dept. of Health Care, Hanseo University) ;
  • Jung, Hong-Ryang (Dept. of Radiological Science, Fareast University) ;
  • Lim, Cheong-Hwan (Dept. of Radiological Science, Fareast University) ;
  • Han, Beom-Hee (Dept. of Radiological Science, Seonam University) ;
  • Han, Sang-Hyun (Dept. of Radiological Science, Seonam University) ;
  • Lee, Sang-Ho (Dept. of Radiological Science, Seonam University) ;
  • Mo, Eun-Hee (Dept. of Radiation Science & Technology, Chonbuk National University) ;
  • Kim, Ki-Jung (Dept. of Radiological Science, Fareast University) ;
  • Lee, Mi-Hwa (Dept. of Radiological Science, Fareast University) ;
  • You, In-Gyu (Dept. of Radiological Science, Fareast University)
  • 투고 : 2014.02.04
  • 심사 : 2014.03.20
  • 발행 : 2014.03.28

초록

현재 우리나라 유방촬영장치의 정도관리 항목에는 국제규격에서 원하는 직선성이 빠져있다. 직선성은 방사선량 및 방사선질의 조정을 위한 요구사항이다. 본 연구에서는 유방촬영장치 5대를 대상으로 IEC 60601-2-45 규격에 맞는 직선성을 시험한 결과 모두 적합한 결과를 보였으며 앞으로 좀 더 발전된 정도관리를 위해서는 지속적인 측정관리가 필요하다.

The quality control items of mammography devices in South Korea do not include the linearity, which is required by international standards. The linearity is a requirement for the adjustment of radiation dose and radiation quality. This study tested the linearity, which was suitable for the IEC 60601-2-45 standard, of the 5 mammography devices. All showed adequate results. Consistent measurement management is required for more developed quality control in the future.

키워드

참고문헌

  1. Jung-min kim, "Diagnostic x-ray equipment to keep the performance management method" Korea Food & Drug Administration, Radiation Health Newletter, vol. 19, no.1, pp.15, 2012
  2. In-Kyu You, Chung-hwan lim, Sang-Ho Lee, Man-ku lee, "", Journal of the Korean Radiological Society, vol.6, no.6, pp.447-454, 2012 https://doi.org/10.7742/jksr.2012.6.6.447
  3. Ministry of Health and Welfare, No.186. Radiation generator rules regarding the safety of. 2001.
  4. Ministry of Health and Welfare, No.234. Radiation generator rules regarding the safety of. 2003
  5. Ministry of Health and Welfare, No.349. Radiation generator rules regarding the safety of. 2006.
  6. IEC International Organization, http://www.iec.ch
  7. Medical electrical equipment - Part 2-45: Particular requirements for basic safety and essential performance of mammographic X-ray equipment and mammomagraphic stereotactic devices
  8. Yun, Jeong-Seop; Kim, Chun-Sik; Go, Sin-Gwan, "Medical X-ray equipment on the management of the maintenance and performance review", *Journal of radiological science and technoligy, vol.15, no.1, pp.25-26, 1992.
  9. Bum-Hee Han, Jong-Il Kim, Sang-Ho Lee, Sang-Hyeon Han, In-Kyu You, "Consistent Comparison for The Linearity Air Kerma of IEC Standards and Commercial Load in Diagnosing DR Generators", https://doi.org/10.7742/jksr.2012.6.5.389
  10. Ministry of Commerce, Industry and Energy, Korean Agency for Technology and Standards, http://standard.ats.go.kr
  11. Do-Wan Kim, "", Korea university Graduate School of Biomedical Sciences, 's, 2010
  12. Joo-Hoon Park, In-Cheol Im, Kyeong-Rae Dong, Sae-Sick Kang, "", The Journal of the Korean Association for Radiation Protection,vol.34 no.1, pp.31-36, 2009.