DOI QR코드

DOI QR Code

Accuracy and Satisfaction with IVIC300 (Intravenous infusion controller)

IVIC300(Intravenous infusion controller)의 정확성과 만족도

  • Park, Jung Hee (Department of Nursing, Konyang University Hospital Team) ;
  • Yang, Nam Young (Department of Nursing Science, Konyang University) ;
  • Na, Moon Jun (Department of Pulmonology, Konyang University Hospital) ;
  • Go, Young Jin (Department of Nursing, Konyang University Hospital Head Nurse in 81ward) ;
  • Kim, Ki Suk (Department of Nursing, Outpatient of Konyang University Hospital Head Nurse in Outpatient) ;
  • Kim, Young Aue (Department of Nursing, Konyang University Hospital Nurse of Injection Room)
  • 박정희 (건양대학교병원 간호부 간호팀) ;
  • 양남영 (건양대학교 간호학과) ;
  • 나문준 (건양대학교병원 호흡기내과) ;
  • 고영진 (건양대학교병원 81병동 파트) ;
  • 김기숙 (건양대학교병원 외래 파트) ;
  • 김영애 (건양대학교병원 주사실)
  • Received : 2013.10.24
  • Accepted : 2014.04.21
  • Published : 2014.04.30

Abstract

Purpose: The purpose of this study was to determine the accuracy and satisfaction with IVIC300 (Intravenous infusion controller). Method: The subjects consisted of 90 nurses and 100 patients. Data collected during July 2013 were analyzed using descriptive statistics, t-test, ANOVA, and $X^2$-test (Chi-Square). Result: The error value of IVIC300 was less than that of Dosi-flow. Differences of error values according to variable of fluid infusion (needle's gauge, patient's moving, patient's calls, and nurse's arabitary calls) were not significant. The mean scores for satisfaction with IVIC300 of nurses ($3.67{\pm}.76$), patients ($3.75{\pm}.63$) were above average. Satisfaction with IVIC300 of nurses differed significantly according to age, work unit, and clinical career. Satisfaction with IVIC300 of patients was not significantly different according to general characteristics. Conclusion: These findings indicate that it is proven the accuracy of IVIC300, and is considered individual characteristics in use IVIC300.

Keywords

References

  1. American Society of Hospital Pharmacists (1993). ASHP guidelines on prevention medication errors in hospital. American Journal of Hospital Pharmacy, 50(2), 305-314.
  2. Bai, S. J. (2010). Fluid therapy : classification and characteristics of intravenous fluids. Journal of Korean medical Association, 53(12), 1103-1112. http://dx.doi.org/10.5124/jkma.2010.53.12.1103
  3. Crass, R. E., & Vance, J. R. (1985). In vivo accuracy of gravity-flow i.v. infusion systems. American Journal of Hospital Pharmacy, 42(2), 328-331.
  4. Dennis, L. K., Eugene, B., Anthony, S. F., Stephen, L. H., Dan, L. L., & J. Larry, J. (2005). Harrison's principles of Intrernal Medicine. 16th ed. New York: McGraw-Hill
  5. Faul, F., Erdfelder, E., Lang, A. G., & Buchner, A. (2007). G*Power 3 : A flexible statistical power analysis power analysis program for the social, behavior, and biomedical sciences. Behavior Research Methods, 39(2), 175-191. https://doi.org/10.3758/BF03193146
  6. Flack, F. C., & Whyte, T. D. (1974). Behaviour of standard gravity-fed administration sets used for intravenous infusion. The British Medical Journal, 17(3), 439-443.
  7. Flack, F. C., & Whyte, T. D. (1975). Variations of drop size in disposable administration sets used for intravenous infusion. Journal of clinical Pathology, 28(6), 510-512. https://doi.org/10.1136/jcp.28.6.510
  8. Han, H. J. (1996). Medication error. Journal of Korean Society of Hospital Pharmacists, 13(3), 226-271.
  9. Hong, S. H., & Lee, E. H. (1996). Development of infusion Pump System Using Drop Sensors. WonKwang Institute of Biomedical Engineering Research, 3(1), 19-32.
  10. Jemma, C. R., & David, A. G. (2007). Errors of intravenous fluid infusion rates in medical inpatients. Clinical Medicine, 7(5), 482-485. https://doi.org/10.7861/clinmedicine.7-5-482
  11. Jeong, Y. B., Lee, M. S., Choi, B. M., Chin, J. H., & Noh, G. J. (2007). A clinical study to evaluate the safety and efficacy of a patient-controlled analgesia pump in post-surgical patients. Korean Journal Anesthesiol, 52(2), 161-165. https://doi.org/10.4097/kjae.2007.52.2.161
  12. Joint Commission International (2011). Joint commission international accreditation standards for hospitals, 4th edition.
  13. Kaushal, R., Bates, D. W., Landrigan, C., McKenna, K. J., Clapp, M. D., Federico, F., & Goldmann, D. A. (2001). Medication errors and adverse drug events in pediatric inpatients. Journal of the American Medical Association, 285, 2114-2120. https://doi.org/10.1001/jama.285.16.2114
  14. Kim, B. J., Lee, S. B., Han, C. W., Shin, B. S., & Sung, H. J. (2004). A new flow controller for medical injection. Measurement, 36, 67-72. https://doi.org/10.1016/j.measurement.2004.04.010
  15. Kim, J. I., Lee, J. H., & Chang, O. J. (2013). Perceived importance and performance of intravenous fluid therapy by nurses in small-medium general hospitals. Korean Journal of Fundamentals of Nursing, 20(4), 372-380. http://dx.doi.org/10.7739/jkafn.2013.20.4.372
  16. Kim, J. H., Wang, S. J., Lee, S. W., Kang, M. S., O, S. H., & You, K. C. (2008). A Report for the research about the accuracy of a flow regulator. The Korean Society of Emergency Medicine, 19(1), 109-113.
  17. Kim, M. R. (2011). Concept Analysis of Patient Safety. Journal of Korean Academy Nursing Administration, 41(1), 1-8. https://doi.org/10.4040/jkan.2011.41.1.1
  18. Lee, H. K. & Sohn, K. H. (2009). Safe Medication Management in Hospital- Emphasis on Dispensing, and Administration. Korean Society for Pharmacoepidermiology and Risk Management, 2(1), 14-25.
  19. Leff, R. D., & Stull, J. C. (1988). Accuracy, continuity, and pattern of flow from five macrorate infusion pumps. American Journal of Hospital Pharmacy, 45(2), 361-365.
  20. Mousavi, M., Khalili, H., & Dashti-khavidaki, S. (2012). Erros in fluid therapy in medical wards. International Journal of Clinical Pharmacy, 34, 374-381. https://doi.org/10.1007/s11096-012-9620-8
  21. Park, J. A., & Kim, B. J. (2009) Critical Thinking Disposition and Clinical Competence in General Hospital Nurses. Journal of Korean Academic of Nursing, 39(6), 840-850 https://doi.org/10.4040/jkan.2009.39.6.840
  22. Park, K., Lee, J., Kim, S. Y., Kim, J., Kim, I., Choi, S. P., Jeong, S. & Hong, S. (2012). Infusion Control and Calculation with Metronome and Drop Counter Based Infusion Therapy Helper. International Journal of Nursing Practice, 19(3), 257-264.
  23. Stull, J. C., Erenberg, A., & Leff, R. D. (1988). Flow rate variability from electronic infusion devices. Critical Care Medicine, 16(9), 888-891. http://dx.doi.org/10.1097/00003246-198809000-00014