Abstract
A whole body counter (WBC) is used in nuclear power plants (NPP) to identify and measure internal radioactivity of workers who is likely to ingest or inhale radionuclides. WBC has several counting geometry, i.e. the thyroid, lung, whole body and gastrointestinal tract, considered with the location where radionuclides are deposited in the body. But only whole body geometry is used to detect internal radioactivity during whole body counting at NPPs. It is overestimated internal exposure dose because this measured values are indicated as the most conservative radioactivity values among the them of others geometry. In this study, experiments to measure radioactivity depending on the counting geometry of WBC were carried out using a WBC, a phantom, and standard radiation sources in order to improve overestimated internal exposure dose. Quantitative criteria, could be selected counting geometry according to ratio of count rates of the upper and lower detectors of the WBC, are provided through statistical analysis method.
원전에서는 방사성핵종의 체내 섭취에 따른 작업종사자의 내부 방사능을 측정하기 위하여 전신계측기(WBC)를 이용하고 있다. 이 계측기는 인체 내 방사성핵종의 침적위치를 고려하여 다양한 측정 모드를 선택하여 측정할 수 있으나, 대부분 전신 모드를 적용하고 있다. 그런데 전신 모드를 적용한 내부방사능 측정값은 WBC 측정 모드 중에서 가장 보수적인 값을 나타내는 특성이 있고, 따라서 내부피폭 방사선량이 과대평가되는 문제점이 있다. 이러한 문제점을 해결하기 위해 WBC, 팬텀, 표준 방사선원을 이용하여 WBC의 측정 모드별 방사능측정 실험을 수행하였다. 이 결과에 대해 통계적 분석방법을 적용하여 WBC의 상부 및 하부 검출기 계측비율에 따라 WBC의 측정 모드를 선정할 수 있는 정량적인 기준을 제시하였다.