Abstract
Fatigue tests were performed in various simulated body environments reflecting various factors (such as body fluids, artificial saliva) relevant within a living body. First, the fatigue limit under a simulated body environment (artificial saliva) was evaluated and the governing factors of implant fatigue strength were looked into by observing the fracture mode. The fatigue life of an implant decreased in the artificial saliva environment compared with that in the ringer environment. Furthermore, in the artificial saliva environment, the implant fracture mode was fatigue failure of fixture as opposed to the abutment screw mode in the ringer environment. In the fatigue test, corrosion products were observed on the implant in the simulated body environment. A larger amount of corrosion products were generated on the artificial saliva specimen than on the ringer specimen. It is thought that the stronger corrosion activity on the artificial saliva specimen as compared with that on the ringer specimen led to an overall decrease of fatigue life of the former specimen. In the case of the implant with a nitrided abutment screw eliminated hardened layer (TixN), a several times increase in fatigue life is achieved in comparison with tungsten carbide-coated implants.
치과용 임플란트를 대상으로 하여 건전환경 및 생체 내의 인자들을 유사하게 반영한 생체유사환경 하에서 피로시험을 수행하였다. 생체유사환경으로는 링거액 환경과 인공타액환경을 각각 사용하였다. 우선 건전환경과 생체유사환경 하에서의 피로수명 및 피로한도를 평가하였다. 또한 피로수명을 향상시킬 수 있는 한 가지 방법을 제안하였다. 인공타액 하에서의 피로수명이 링거환경보다 대체적으로 감소하였다. 시편에 대한 부식작용은 인공타액이 링거액보다 훨씬 크다. 이러한 부식의 차이가 두 경우의 피로수명의 차이로 나타난 것으로 판단된다. 질화처리된 지대주 나사의 질화경화층을 제거하고, 질소확산층을 최표면으로 하는 지대주 나사를 사용한 임플란트의 피로수명은 텅스텐카바이드를 코딩한 기존 제품에 비해 최대 8배의 피로수명 향상효과가 확인되었다.