DOI QR코드

DOI QR Code

Design of Unconstrained Baby Monitoring System Based on a Doppler Radar and an Air Mattress

도플러 레이더와 에어 매트리스를 이용한 무구속 방식의 신생아 감시 장치 설계

  • Received : 2014.07.17
  • Accepted : 2014.11.17
  • Published : 2014.12.30

Abstract

Continuous monitoring of heart rates and respiratory rates for newborns or infants is very important since the abnormal breathing and heart problems can threaten the life of newborns or infants. A noncontact baby monitoring system based on a Doppler radar and an air mattress was designed. The Doppler radar was used to acquire respiratory signals and the air mattress was employed to obtain heart rates. The performance of the designed system was evaluated using a commercialized infant simulator ($Simbaby^{TM}$) and a respiration belt transducer was used to measure respiration rates as a reference. Results for respiratory rates revealed that the correlation coefficients between I-and Q-channel and the respiration belt were 0.84 and 0.91 and the mean ${\pm}$ standard deviations of errors between them were $1.66{\pm}1.92$ (bpm) and $0.88{\pm}1.65$ (bpm). Heart rates showed that the correlation coefficient between air mattress and set value of the simulator was 0.73 and the mean ${\pm}$ standard deviation of errors between them was $1.09{\pm}3.45$ (bpm). These results indicate that the designed system holds the potential as an effective monitoring tool for continuous monitoring heart rates and respiratory rates of newborns or infants.

Keywords

References

  1. K. Seo, J.G. Jeon, Y.J. Han, "Obstetric and neonatal causes of Korean neonatal death," Obstetrics & Gynecology Science, vol. 44, no. 10, pp. 1844-1850, 2001.
  2. C.R. Kim, J.W. Oh, M.K. Yum, S.J. Moon, "Risk Factors for Neonatal Sepsis in Premature Infants Admitted to Neonatal Intensive Care Unit," The Korean Society of Pediatric Infectious Diseases, vol. 43, no. 9, 2000.
  3. S.H. Abman, "Monitoring cardiovascular function in infants with chronic lung disease of prematurity," Arch Dis Child Fetal Neonatal, Ed 87, pp. F15-F18, 2002. https://doi.org/10.1136/fn.87.1.F15
  4. T. Matsuda, M. Makikawa, "ECG Monitoring of a car driver using capacitively-coupled electrodes," 30th Annual International IEEE EMBS Conference, Vancouver, Canada, Aug. 2008, pp. 1315-1318.
  5. Y.G. Lim, K.K. Kim, K.S. Park, "ECG Recording on a bed during sleep without direct skin-contact," IEEE Trans. Biomed. Eng., vol. 54, no. 4, pp. 718-725, 2006.
  6. A.K. Abbas, K. Heimann, K. Jergus, T. Orlikowsky, S. Leonhardt, "Neonatal non-contact respiratory monitoring based on real-time infrared thermography," BioMedical Engineering OnLine, vol. 10, no. 93, 2011.
  7. R.M. DiBlasi, "Neonatal noninvasive ventilation techniques: do we really need to intubate?," Respiratory Care, vol. 56, no. 9, pp. 1273-1297, 2011. https://doi.org/10.4187/respcare.01376
  8. B.K. Park, O. Boric-Lubecke, V.M. Lubecke, "Arctangent demodulation with DC offset compensation in quadrature Doppler radar receiver systems," IEEE Trans. Microw. Theory Tech., vol. 55, no. 5, pp. 1073-1079, 2007. https://doi.org/10.1109/TMTT.2007.895653
  9. J.Y. Shin, S.P. Cho, B.J. Jang, H.D. Park, Y.S. Lee, K.J. Lee "A study of noncontact heartbeat and respiration detection using the Doppler radar," The Institute of Electronics Engineers of Korea, vol. 46, no. 1, pp. 1-9, 2009.
  10. N. Hafner, J.C. Drazen, V.M. Lubecke, "Fish heart rate monitoring by body-contact Doppler radar," IEEE Sens. J., vol. 13, no. 1, pp. 408-414, 2013. https://doi.org/10.1109/JSEN.2012.2210400
  11. A. Susan, "Normal vital signs in children: Heart rate, respirations, temperature, and blood pressure." Complex Child E-Magazine, 2008, pp. 1-4.
  12. S.K. Nagarkoti, B. Singh, M. Kumar, "An algorithm for fetal heart rate detection using wavelet transform," 1st Int'l Conf. on Recent Advances in Information Technology, Dhanbad, India, Mar. 2012, pp. 838-840.
  13. C.M. Tenedero, M.A.D. Raya, L.G. Sison, "Design and implementation of a single- channel ECG amplifier with DSP post-processing in Matlab," Third National Electronics & Engineering Conference, Phillipines, Nov. 2002, pp. 1-4.
  14. Y. Xiao, J. Lin, O. Boric-Lubecke and V. M. Lubecke "A Ka-band low power Doppler radar system for remote detection of cardiopulmonary motion," Proc. 27th IEEE Annu. Eng. Med. Biol. Soc. Conf., pp. 7151-7154, 2005.