Measurement of neutron spectra in MC50 cyclotron using Bonner sphere spectrometer with LiI scintillation detector

LiI 섬광검출기 기반의 보너구 스펙트로메터를 이용한 MC50 사이클로트론의 중성자스펙트럼 측정

  • 하위호 (한국원자력의학원 국가방사선비상진료센터) ;
  • 박세영 (한국원자력의학원 국가방사선비상진료센터) ;
  • 유재룡 (한국원자력의학원 국가방사선비상진료센터) ;
  • 윤석원 (한국원자력의학원 국가방사선비상진료센터) ;
  • 이승숙 (한국원자력의학원 국가방사선비상진료센터) ;
  • 김정호 (한국표준과학연구원 방사선표준센터) ;
  • 김종경 (한양대학교 원자력공학과)
  • Received : 2013.07.24
  • Accepted : 2013.08.22
  • Published : 2013.09.30

Abstract

Operational nuclear facilities such as nuclear power plants and particle accelerators show various neutron spectra according to the type of facilities and specific position. Necessities of neutron dose management and neutron monitoring for radiation protection of radiation workers in such a kind of facilities have continuously increased in recent years. Bonner sphere spectrometers are widely used for measurement of neutron spectra. Data on response function of neutron detector, default neutron spectra and count rates of Bonner sphere spectrometer are required to obtain unfolded neutron spectra in specific workplaces. In this study, we carried out measurement of neutron spectra produced in MC50 cyclotron using Bonner sphere spectrometer with LiI scintillation detector. Additionally, we estimated quantitative data on neutron flux, mean neutron energy and ambient dose equivalent rate according to the incident proton energies and positions in MC50 cyclotron.

원자력발전소 및 입자 가속기 등 상용화된 원자력관계시설에서는 해당 시설 내부의 구조 및 위치 등에 따라 다양한 에너지를 가진 중성자스펙트럼을 나타내며 이에 대한 중성자선량 평가 및 작업자의 방사선방호 조치에 대한 수요는 점차 증가하고 있다. 중성자스펙트럼을 측정하기 위해서 일반적으로 보너구 스펙트로메터가 많이 이용된다. 보너구 스펙트로메터를 이용한 중성자 스펙트럼 측정의 경우 중성자검출기의 반응함수, 초기추정 스펙트럼, 계수율 측정값을 사용하여 중성자스펙트럼 언폴딩을 수행하게 된다. 본 연구에서는 MC50 사이클로트론에서 발생하는 중성자스펙트럼을 LiI 섬광검출기 기반의 보너구 스펙트로메터를 이용하여 측정하였으며 표적에 입사되는 양성자 에너지와 위치에 따른 중성자선속, 중성자 평균에너지 및 선량률 자료를 정량적으로 평가하였다.

Keywords

References

  1. International Commission on Radiation Protection. Conversion coefficients for use in radiological protection against external radiation. ICRP Publication 74. 1996.
  2. International Commission on Radiation Protection. Conversion coefficients for radiological protection quantities for external radiation exposures. ICRP Publication 116. 2010.
  3. International Atomic Energy Agency. Compendium of neutron spectra and detector response for radiation protection purpose. IAEA. 2001; IAEA Technical Report No. 403.
  4. Thomas DJ. Neutron spectrometry. Radiation Measurement. 2010;45:1178-1185. https://doi.org/10.1016/j.radmeas.2010.04.016
  5. Los Alamos National Laboratory. MCNPXTM User's Manual Version 2.6.0. LA-CP-07-1473. 2008.
  6. 박세영, 하위호, 유재룡, 윤석원, 이승숙. LiI 섬광체를 이용한 보너구 스펙트로메터의 반응함수 계산. 2012년도 대한방사선방어학회 춘계학술발표회논문집. 2012.
  7. http://physics.nist.gov/cgi-bin/Star/compos.pl.
  8. Kim BH, Kim JS, Kim JL, Kim YS, Yang TG, Lee MY. Determination of the neutron spectra in the neutron therapy room of KIRAMS. Radiat. Prot. Dosimet. 2007;126:348-389.
  9. S.E. International, INC. Universal Radiation Spectrum Analyzer Operation Manual. 2003.
  10. Physikalisch-Technische Bundesanstalt. UMG FC Manual. 2004.
  11. Mares V, Schraube H. Evaluation of the response matrix of a Bonner sphere spectrometer with LiI detector from thermal energy to 100 MeV. Nucl. Instr. and Meth. in Phys. Res. 1994;A337:461-473
  12. Vega-Carrillo HR, Donaire I, Gallego E, Manzanares-Acuna E, Lorente A, Iniguez MP, Martin-Martin A, Gutierrez-Villanueva JL. Calculation of Response matrix of a BSS with 6LiI scintillator. Rev. Mex. de Fís. 2008;54(1):57-62.
  13. http://www-nds.iaea.org/exfor/endf.htm.