Quantitative Indices of Small Heart According to Reconstruction Method of Myocardial Perfusion SPECT Using the 201Tl

201Tl을 이용한 심근관류 SPECT에서 재구성 방법에 따른 작은 용적 심장의 정량 지표 변화

  • Kim, Sung Hwan (Department of Nuclear Medicine, Asan Medical Center) ;
  • Ryu, Jae Kwang (Department of Nuclear Medicine, Asan Medical Center) ;
  • Yoon, Soon Sang (Department of Nuclear Medicine, Asan Medical Center) ;
  • Kim, Eun Hye (Department of Nuclear Medicine, Asan Medical Center)
  • 김성환 (서울아산병원 핵의학과) ;
  • 류재광 (서울아산병원 핵의학과) ;
  • 윤순상 (서울아산병원 핵의학과) ;
  • 김은혜 (서울아산병원 핵의학과)
  • Received : 2013.02.06
  • Accepted : 2013.04.08
  • Published : 2013.04.27

Abstract

Purpose: Myocardial perfusion SPECT using $^{201}Tl$ is an important method for viability of left ventricle and quantitative evaluation of cardiac function and now various reconstruction methods are used to improve the image quality. But in case of small sized heart, you should always be careful because of the Partial Volume Effect which may cause errors of quantitative indices at the reconstruction step. So, In this study, we compared those quantitative indices of left ventricle according to the reconstruction method of myocardial perfusion SPECT with the Echocardiography and verified the degree of the differences between them. Materials and Methods: Based on ESV 30 mL of Echocardiography, we divided 278 patients (male;98, female;188, Mean age;$65.5{\pm}11.1$) who visited the Asan medical center from February to September, 2012 into two categories; below the criteria to small sized heart, otherwise, normal or large sized heart. Filtered and output each case, we applied the method of FBP and OSEM to each of them, and calculated EDV, ESV and LVEF, and we conducted statistical processing through Repeated Measures ANOVA with indices that measured in Echocardiography. Results: In case of men and women, there were no significant difference in EDV between FBP and OSEM (p=0.053, p=0.098), but in case of Echocardiography, there were meaningful differences (p<0.001). The change of ESV especially women in small sized heard, significant differences has occurred among FBP, OSEM and Echocardiography. Also, in LVEF, there were no difference in men and women who have normal sized heart among FBP, OSEM and Echocardiography (p=0.375, p=0.969), but the women with small sized heart have showed significant differences (p<0.001). Conclusion: The change in quantitative indices of left ventricle between Nuclear cardiology image reconstruction, no difference has occurred in the patients with normal sized heart but based on ESV, under 30 mL of small sized heart, especially in female, there were significant differences in FBP, OSEM and Echocardiography. We found out that overestimated LVEF caused by PVE can be reduced in average by applying OSEM to all kinds of gamma camera, which are used in analyzing the differences.

$^{201}Tl$을 이용한 심근관류 SPECT 검사는 좌심실의 생존능 및 심장 기능의 정량적 평가를 함에 있어 중요한 방법으로서 현재 영상의 질을 향상시키기 위해 다양한 재구성 방법들이 이용 되고 있다. 하지만 작은 용적 심장에서는 부분용적효과로 인해 재구성 단계에서 정량 지표 값의 오류를 야기 할 수 있으므로 항상 주의 해야 한다. 이에 본 연구는 심근관류 SPECT 검사의 재구성 방법에 따른 좌심실의 정량적 지표 값을 심장 초음파와 서로 비교 함으로써 그 차이의 정도를 확인 한다. 2012년 2월부터 9월까지 본원에 내원하여 심근관류 SPECT 및 심장 초음파 검사를 실시한 278명의 환자(남자 90명, 여자 188명, 평균$65.5{\pm}11.1$세)를 심장 초음파의 ESV 30 mL를 기준으로 삼아 그 이하를 작은 용적 심장, 그 이상을 보통 또는 큰 용적심장으로 구분하였다. 각각 여과 후 FBP 및 OSEM의 방법을 적용하여 EDV, ESV 그리고 LVEF를 산출하였으며, 이를 심장 초음파에서 측정된 지표들과 함께 반복측정 분산분석 방법(Repeated Measures ANOVA)으로 분석하였다. 남녀 간의 EDV는 FBP, OSEM 간 유의한 차이가 없었으나 (p=0.053, p=0.098), 심장 초음파와의 비교에서는 유의한 차이를 보였다(p<0.001). ESV의 변화는 특히 작은 용적 심장을 가진 여성에서 FBP, OSEM, 심장 초음파 모두 유의한 차이(p<0.001)를 보였다. 또한 LVEF에서도 보통 용적 심장을 가진 남녀 모두 FBP, OSEM, 심장 초음파 간 유의한 차이는 보이지 않았으나(p=0.375, p=0.969), 작은 용적 심장을 가진 여성에서 모두 유의한 차이(p<0.001)를 보였다. 핵의학 영상 재구성 방법 간 좌심실의 정량적 지표 값의 변화는 보통 용적 심장을 가진 환자에서는 유의한 차이를 발견 할수 없었으나, ESV를 기준으로 30 mL 이하의 작은 용적 심장, 특히 여성에서는 FBP, IR_OSEM, 심장 초음파 간 유사한 차이를 확인 할 수 있었다. 하지만 이러한 차이는 분석에 사용된 3종류의 모든 감마 카메라에서 OSEM 적용 시FBP 보다 평균적으로 심장 초음파와의 오차가 적은 LVEF값이 산출됨을 확인 할 수 있었다.