The importance of ultrasound examination in the field of medical imaging has been emphasized and the interest in sonographic image evaluation is growing. However image evaluations by the non-standardized criteria and methods, so establishment of legal provisions and objective evaluation criteria are needed. In this study, we used SNR to find out more quantitative way and supplement the limitations of the existing phantom image evaluation. The results of acquired 8 images using ATS-539 multipurpose phantom were compared in SNR of sensitivity and gray-scale dynamic range. In the result of the experiment, excellent equipment of existing phantom images are G1, S1 and G2 in regular sequence. In SNR of sensitivity, G1, S1 and G2 and in SNR of gray-scale dynamic range, S1 G1 and G2 in order. In the conclusion, all the experiment results did not show big difference and regular pattern neither. Therefore, the new evaluation measures should be used with the existing phantom image evaluation method for more objective and quantitative evaluation of the ultrasound imaging device.
영상의학 분야에서 초음파 검사의 중요성이 강조되며 영상평가에 대한 관심이 고조되고 있다. 또한 표준화 되지 않은 기준과 방법으로 영상평가를 시행하고 있어 법적규정과 객관적인 평가기준마련이 시급한 실정이다. 이에 본 연구는 기존의 초음파 영상평가의 한계와 제한적 방법을 보완하고 정량적인 방법을 제시하고자 SNR의 영상평가 실험을 실시하였다. ATS-539 다목적 초음파팬텀을 이용해 획득된 8가지 팬텀영상방법과 예민도, 회색조와 역학범위 두 항목의 SNR 결과를 비교 및 분석하였다. 실험결과 기존의 영상평가결과에서의 우수한 장비는 G1 S1, G2의 순으로 나타났다. 논문에서 제안하는 실험으로 예민도의 SNR 결과는 G1, S1, G2 순으로, 회색조와 역학범위 항목의 SNR 결과는 S1, G1, G2의 순으로 나타났다. 결과에서 기존의 초음파 영상평가 방법과 제시하는 정량적 영상평가 방법이 큰 차이를 보이지도, 일치하지도 않게 나타났다. 객관적이고 정량적인 초음파영상장치의 평가를 위해서는 기존의 팬텀을 이용한 영상평가방법과 SNR을 이용한 영상평가의 측정방법도 함께 평가되어야 할 것으로 사료된다.