DOI QR코드

DOI QR Code

Exposure dose Reduction using Pb Banding of own manufacturing

자체제작 Pb 밴딩을 이용한 피폭선량 감소

  • Received : 2013.04.28
  • Accepted : 2013.06.20
  • Published : 2013.06.30

Abstract

Exposure dose to the examinee was measured using glass dosimeter in the test using panorama device at the time of dental treatment. As a result of measuring expose dose to lens according to the different sizes of Pb banding of own manufacturing to reduce exposure dose to lens especially sensitive to radiation, it was verified that exposure dose to lens varied depending on the size of the Pb banding. With the size of Pb banding of $3{\times}20{\times}0.2cm$, exposure dose tended to increase higher than normal value, and with the size of or more than $5{\times}20{\times}0.2cm$, it decreased. And also, the obtained image with the size of $7{\times}20{\times}0.2cm$ was not suitable for diagnosis. Therefore, it is expected that exposure dose would be reduced by using Pb banding of the size of not less than $5{\times}20{\times}0.2cm$ and not more than $6{\times}20{\times}0.2cm$ in the test, to minimize exposure dose and conduct panorama test efficiently.

치과 진료시 파노라마 장치를 이용한 검사에서 유리선량계를 사용하여 피검자의 피폭선량을 측정하였다. 특히 방사선에 민감한 수정체의 피폭선량을 줄이기 위하여 자체 제작한 Pb밴딩의 크기에 따라 수정체 피폭선량을 측정한 결과 Pb밴딩의 크기에 따라서 수정체의 피폭선량이 다르다는 것을 확인할 수 있었다. Pb밴딩의 크기가 $3{\times}20{\times}0.2cm$에서는 정상치보다 피폭선량이 증가하는 경향을 보였으며 $5{\times}20{\times}0.2cm$ 이상의 크기에서는 피폭선량이 감소하는 결과를 보였다. 또한 획득되어진 영상 $7{\times}20{\times}0.2cm$ 크기에서 진단에 부적합한 영상으로 판정되었다. 그러므로 피폭선량을 최소화하고 효과적인 파노라마 검사를 수행하기 위해서는 Pb밴딩 $5{\times}20{\times}0.2cm$이상 $6{\times}20{\times}0.2cm$이하 크기를 사용하여 검사에 활용하면 피폭선량이 감소될 것으로 기대한다.

Keywords