A Comparison of the Implant Stability Among the Bone Density Groups: Prospective Study

골밀도와 임플란트 안정성 변화에 대한 전향적 연구

  • Ko, Sok-Min (Youngtong-Bestden Dental Clinic) ;
  • Park, Seong-Jae (ITplant Dental Clinic) ;
  • Lee, In-Kyung (Department of Periodontics, Jukjeon Dental Hospital, College of Dentistry, Dankook University)
  • 고석민 (영통베스트덴치과) ;
  • 박성재 (IT플란트치과) ;
  • 이인경 (단국대학교 치과대학 죽전치과병원 치주과)
  • Received : 2012.12.07
  • Accepted : 2013.03.25
  • Published : 2013.03.30

Abstract

The purpose of this study was to get a proper time when implant was loaded and superstructure was fabricated by tracing a change of the implant stability in bone type for 1 year. We carried out RFA(resonance frequency analysis) of 67 implants in 36 patients at the time of surgery, 2, 6, 10, 14weeks, and 1year postoperatively for each implant, and analyzed data for different bone density. The ISQ value at the time of 10 & 14weeks postoperatively were significant(P<0.01) in subjective grouping, while that were not significant in objective grouping(P>0.05). The change pattern of ISQ value could be expressed $y=y_0+a{\times}\{1-\exp(-b{\times}x)\}$ for 1 year in this study, ISQ value at surgery might be used to estimate the time of superstructure fabrication consequentially.

임플란트 식립 후 상부구조 제작 시기를 결정하기 위해 임플란트 고정체의 골유착과 안정성의 평가는 중요하며 임플란트 안정성이 변화하지 않는 고원(plateau) 현상이 나타나는 시점을 찾는다면 임플란트 저작하중에 안정적인 시기를 판단하는 객관적 지표가 될 수 있다. 본 연구의 목적은 1년간의 전향적인 연구를 통해 골밀도 간 임플란트 안정성 변화를 추적하여 저작하중을 부담하기에 적당한 안정된 시점을 구함으로써 상부구조 제작 시기를 예측하고자 하였다. 총 36명의 환자에게 식립된 67개의 임플란트를 대상으로 식립 직후, 2주, 6주, 10주, 14주, 1년 등 총 6회에 걸쳐 RFA를 시행하여 임플란트 안정성의 변화추이를 골밀도에 따라 비교 분석하였다. 식립 직후와 각 치유기간 동안의 ISQ값 비교에서 주관적으로 평가된 골질군 간에는 식립 10주와 14주에서 임플란트 안정성에 차이가 있었으나(P<0.01), 객관적으로 평가된 골질군 간에는 임플란트 안정성 변화에 차이가 없었다(P>0.05). 본 연구의 1년간의 추적조사에서 ISQ값의 변화는 $y=y_0+a{\times}\{1-\exp(-b{\times}x)\}$로 나타낼 수 있었으며, 이를 이용하면 임플란트식립 당시의 ISQ값으로 상부구조 제작 시점을 유추할 수 있다.

Keywords

References

  1. Adell R, Eriksson B, Lekholm U, Brånemark PI, Jemt T. Long-term follow-up study of osseointegrated implants in the treatment of totally edentulous jaws. Int J Oral Maxillofac Implants 1990;5:347-59.
  2. Nevins M, Langer B. The successful use of osseointegrated implants for the treatment of the recalcitrant periodontal patient. J Periodontol 1995;66;150-7. https://doi.org/10.1902/jop.1995.66.2.150
  3. Esposito M, Hirsch JM, Lekholm U, Thomsen P. Biological factors contributing to failures of osseointegrated oral implants.(II). Etiopathogenesis. Eur J Oral Sci 1998;106:721-64. https://doi.org/10.1046/j.0909-8836..t01-6-.x
  4. Adell R, Lekholm U, Rockler B, Branemark P-I. A 15 year study of osseointegrated implants in the treatment of the edentulous jaw. Int J Oral Surg 1981;6:387-416.
  5. Misch CE. Stress Factor: Influence on treatment planning. In: Misch CE. Dental implant Prosthetics. St Louis: Mosby 2005:71-90.
  6. Collaert B, De Bruyn H. Immediate functional loading of TiOblast dental implants in full-arch edentulous maxillae: a 3-year prospective study. Clin Oral Implants Res 200819:1254-60.
  7. Lekholm U, Zarb GA. Patient selection and preparation. In: Branemark P-I, Zarb GA, Albrektsson T. Tissue Integrated Prostheses: Osseointegration in Clinical Dentistry. Chicago: Quintessence 1985:199-209.
  8. Misch CE: Bone character:second vital implant criterion. Dent Today 1988;7:39-40.
  9. Misch CE, Kirocos LT. Diagnostic imaging and techniques. In Misch CE, editor, Contemporary implant dentistry, ed 2, St Louis: Mosby 1999:73-87.
  10. Meredith N, Book K, Friberg B, Jemt T, Sennerby L. Resonance frequency measurements of implant stability in vivo. A cross-sectional and longitudinal study of resonance frequency measurements on implants in the edentulous and partially dentate maxilla. Clin Oral Implants Res. 1997;8:226-33. https://doi.org/10.1034/j.1600-0501.1997.080309.x
  11. Bischof M, Nedir R, Szmukler-Moncler S, Bernard JP, Samson J. Implant stability measurement of delayed and immediately loaded implants during healing. Clin Oral Implants Res. 2004;15:529-39. https://doi.org/10.1111/j.1600-0501.2004.01042.x
  12. Nkenke E, Hahn M, Weinzierl K, et al. Implant stability and histomorphometry: a correlation study in human cadavers using stepped cylinder implants. Clin Oral Implants Res. 2003;14:601-9. https://doi.org/10.1034/j.1600-0501.2003.00937.x
  13. Jaffin RA, Berman CL. The excessive loss of Brånemark fixtures in the type IV bone: a 5-year analysis. J Periodontol 1991;62:2-4. https://doi.org/10.1902/jop.1991.62.1.2
  14. Meredith N, Shagaldi F, Alleyne D, Sennerby L, Cawley P. The application of resonance frequency measurements to study the stability of titanium implants during healing in the rabbit tibia. Clin Oral Implants Res. 1997;8:234-43. https://doi.org/10.1034/j.1600-0501.1997.080310.x
  15. Lang NP, WilsonTG, Corbet EF. Biological complications with dental implants: their prevention, diagnosis and treatment. Clin Oral Implants Res. 2000;11 Suppl 1:146-55. https://doi.org/10.1034/j.1600-0501.2000.011S1146.x
  16. Sennerby L, Dasmah A, Larsson B, Iverhed M. Bone tissue responses to surface-modified zirconia implants: A histomorphometric and removal torque study in the rabbit. Clin Implant Dent Relat Res. 2005;7 Suppl 1:S13-20 https://doi.org/10.1111/j.1708-8208.2005.tb00070.x
  17. Klokkevold PR, Johnson P, Dadgostari S, et al. Early endosseous integration enhanced by dual acid etching of titanium: a torque removal study in the rabbit. Clin Oral Implants Res 2001;12:350-7. https://doi.org/10.1034/j.1600-0501.2001.012004350.x
  18. Brandt J, Bierogel C, Holweg K, Hein W, Grellmann W. Extended push-out test to characterize the failure of bone-implant interface. Biomed Tech 2005;50: 201-6. https://doi.org/10.1515/BMT.2005.028
  19. Dhert WJ, Verheyen CC, Braak LH, et al. A finite element analysis of the push-out test: influence of test conditions. J Biomed Mater Res 1992;26:119-30. https://doi.org/10.1002/jbm.820260111
  20. Lachmann S, Laval JY, Jäger B, et al. Resonance frequency analysis and damping capacity assessment. Part 2: peri-implant bone loss follow-up. An in vitro study with the Periotest and Osstell instruments. Clin Oral Implants Res. 2006;17:80-4. https://doi.org/10.1111/j.1600-0501.2005.01174.x
  21. Olive J, Aparicio C. The periotest method of osseointegrated oral implant stability. Int J Oral Maxillofac Implants 1990;5:390-400.
  22. Earthman JC, Li Y, VanSchoiack LR, Sheets CG, Wu JC. Reconstructive materials and bone tissue engineering in implant dentistry. Dent Clin N Am 2006;50:229-44. https://doi.org/10.1016/j.cden.2005.11.010
  23. Balleri P, Cozzolino A, Ghelli L, Momicchioli G, Varriale A. Stability measurements of osseointegrated implants using Osstell in partially edentulous jaws after 1 year of loading: a pilot study. Clin Implant Dent Relat Res 2002;4:128-32. https://doi.org/10.1111/j.1708-8208.2002.tb00162.x
  24. Jae-Min Kim, Sun-Jong Kim, Inho Han, et al. A comparison of the implant stability among various implant systems: clinical study. J Adv Prosthodont 2009;1:31-6. https://doi.org/10.4047/jap.2009.1.1.31
  25. Turkyilmaz I, McGlumphy EA. Influence of bone density on implant stability parameters and implant success: a retrospective clinical study. BMC Oral Health 2008;8:32-9. https://doi.org/10.1186/1472-6831-8-32