Evaluation of Image Quality and dose with the Change of kVp and BMI in the Liver CT

CT 검사 시 관전압과 BMI 변화에 따른 화질 및 피폭평가

  • 김동현 (부산가톨릭대학교 보건과학대학 방사선학과) ;
  • 고성진 (부산가톨릭대학교 보건과학대학 방사선학과) ;
  • 강세식 (부산가톨릭대학교 보건과학대학 방사선학과) ;
  • 김정훈 (부산가톨릭대학교 보건과학대학 방사선학과) ;
  • 최석윤 (부산가톨릭대학교 보건과학대학 방사선학과) ;
  • 김창수 (부산가톨릭대학교 보건과학대학 방사선학과)
  • Received : 2013.03.22
  • Accepted : 2013.04.11
  • Published : 2013.06.28

Abstract

CT for follow-up visits because of liver disease, body mass index (BMI) and kVp according to the change of the image quality and radiation dose to evaluate for changes. March 2010 to June 2011 at Pusan P University Hospital, abdominal CT scans a patient BMI (Body Mass Index. Less BMI) index was less than 25 in the treatment of subjects had a 48-person Noise and SNR at 100kVp abdominal image is lager than the 120kVp image. CTDI volume value at by the analysis of the radiation dose is 4.47mGy(100kVp) and 9.01mGy(120kVp). So CTDIvol in 100kVp is smaller than CTDIvol in 120kVp(decrease by 44.1%). And, effective dose is 7.1mSv(100kVp) and 12.51mSv(120kVp). So effective dose in 100kVp is smaller than effective dose in 120kVp(decrease by 43%). Evaluation of image quality is that Unacceptable 0 person, Suboptimal 0 person, Adequate 0 person, Good 1 person, Excellent 47 person. In case of repeatly patient, we examinate abdomianl CT scan by using low kVp and body mass index less than 25. We can has good quality image and benefit of low radiation dose.

간질환을 주소로 추적검사를 위해 내원한 CT(Computer Tomography, 이하 CT) 검사자를 대상으로 체질량 지수와 관전압 변화에 따른 영상의 화질 및 방사선 피폭선량변화에 대하여 알아보고자 하였다. 2010년 3월부터 2011년 6월까지 부산 P대학병원에 복부 CT 를 검사한 환자 중 체질량지수(Body Mass Index, 이하 BMI)가 25이하인 환자를 대상으로 하였고 대상자는 48명이었다. 영상의 질의 객관적 평가로 신호대 잡음비와 유효선량을 비교하였다. 복부 영상의 화질평가는 영상의학과 의사2명이 한국의료영상품질관리원에서 선정한 임상영상 평가의 기준을 근거로 해 1점에서 20점까지 점수를 매겨 평가하였다. 피폭선량분석에서 CTDIvol값은 관전압이 100kVp일 때 120kVp보다 약44.1%가 감소하였다. 그리고 유효선량은 관전압 100kVp일 때 120kVp보다 약43%가 감소하였다. 영상의 화질 평가는 반복적으로 CT검사를 위해 내원한 총48명의 검사자 영상 중 Good 1명, Excellent 47명으로 평가되었다. 추적검사를 시행하는 환자 중 BMI지수가 25이하인 환자들을 대상으로 저관전압을 적용한 복부 CT검사 시 영상의 질적 저하없이 진단 가치가 있는 영상의 획득과 피폭선량 감소효과를 얻을 수 있다고 사료된다.

Keywords

References

  1. 김문찬, "복부 CT검사시 technical factor 변화에 따른 환자피폭선량의 비교", 대한전산화단층기술학회지, 제4권, 제1호, p.91, 2003.
  2. D. Brenner, C. Elliston, and W. Berdon, "Estimated risks of radiation-induced fatal cancer from pediatric CT," AJR Am J Roentgenol, Vol.176, pp.289-296, 2001. https://doi.org/10.2214/ajr.176.2.1760289
  3. D. J. Brenner and C. D. Elliston, Estimated radiaton risks potentially associated with full-body CT screening, Radiology, Vol.232, pp.735-738, 2004. https://doi.org/10.1148/radiol.2323031095
  4. D. J. Brenner, Radiaton risks potentially associated with low-dose CT screening of adult smokers for lung cancer, Radiology, Vol.231, pp.440-445, 2004. https://doi.org/10.1148/radiol.2312030880
  5. S. Yoshinaga, K. Mabuchi, A. J. Sigurdson, M. M. Doody, and E. Ron, Cancer risks among radiologists and radiologic technologists: review of epidermiologicstudies, Radiology, Vol.233, pp.313-321, 2004. https://doi.org/10.1148/radiol.2332031119
  6. 고인호, 김경근, 김동윤, Textbook of Computed Tomography, p.62, 청구문화사, 2003.
  7. S. Gryspeerdt, G. Hoe, L. V. Marchal, and A. L. Baert, "Evaluation of hepatic perfusion disorders with double-phase spiral CT," RadioGraphics, Vol.17, pp.337-348, 1997. https://doi.org/10.1148/radiographics.17.2.9084076
  8. J. P. Heiken, J. A. Brink, and M. W. Vannier, "Spiral(helical)CT," Radiology, Vol.189, pp.647-656, 1993. https://doi.org/10.1148/radiology.189.3.8234684
  9. R. K. Zeman and S. H. Fox, "Silverman PM,etal.Helical(spiral)CT of abdomen," AJR Vol.160, pp.719-725, 1993. https://doi.org/10.2214/ajr.160.4.8456652
  10. M. A. Hoon, S. J. Lee, and S. A. Kim, "Usefulness of Follow-up Computed Tomography after Surgery for Early Gastric Cancer," J Korean Radioi Soc, Vol.47, pp.513-518, 2002. https://doi.org/10.3348/jkrs.2002.47.5.513
  11. W. Huda, E. M. Scalzetti, and G. Levin, "Technique factors and image quality as functions of patient weight at abdominal CT," Radiology, Vol.217, pp.430-435, 2002.
  12. A. B. Sigal-Cinqualbre, R. Hennequin, H. T. Abada, X. Chen, J. F. Paul, "Low-kilovoltage multi-detector row chest CT in adults:feasibility and effect on image quality and iodine dose," Radiology, Vol.231, pp.169-174, 2004. https://doi.org/10.1148/radiol.2311030191
  13. Y. Nakayama, K. Awai, Y. Funama, M. Hatemura, M. Imuta, and T. Nakaura, "Abdominal CT with low tube voltage: preliminary observation about radiation dose, contrast enhancement, image quality, and noise," Radiology, Vol.237, pp.945-951, 2005. https://doi.org/10.1148/radiol.2373041655
  14. Y. C. Nam, "복부 CT 검사시 SDCT와 MDCT의 선량비교 및 16채널 MDCT에서 scan parameter의 변화에 의한 선량비교", 대한전산화단층기술학회지, 제5권, 제1호, pp.55-63, 2006.
  15. 보건복지가족부, 식품의약품안전평가원, 대한영상의학회, 대한방사선협회, "CT 엑스선검사에서의 환자선량 권고량 가이드 라인", 방사선안전관리시리즈, No.19, 2009.
  16. UNSCEAR 2000 Report Vol. I Sources and Effects of Ionization Radiation, Annex D Medical radiation exposure, UNSCEAR (2000)
  17. B. Jung, A. H. Mahnken, A. Stargardt, J. Simon, T. G. Flohr, and S. Schaller, "Individual weigt-adepted examination protocol in retrospectively ECG-gated MSCT of the heart," Eur Radiol, Vol.13, pp.2560-2566, 2003. https://doi.org/10.1007/s00330-003-2111-5
  18. T. Irle and H. Inoue, "Individual modulation of the tube current-seconds to achieve similar levels of image noise in contrast-enhanced abdominal CT," AJR Am J Roentgenol, Vol.184, pp.1514-1518, 2005. https://doi.org/10.2214/ajr.184.5.01841514
  19. K. T. Bae, B. A. Seeck, C. F. Hildebolt, C. Tao, F. Zhu, and M. Kanematsu, "Contrast enhancement in cardiovascular MDCT: effect of body weight, height, body surface area, body mass index, and obesity," AJR Am J Roentgenol, Vol.190, pp.777-784, 2008. https://doi.org/10.2214/AJR.07.2765
  20. 김윤경, "체질량 지수 및 관전압 변화에 따른 CT 관상동맥 조영술에서 영상의 질 및 방사선 피폭량 비교," 대한 영상 의학회지, 제62권, pp.29-35, 2010.
  21. C. Suess and X. Chen, "Dose optimization in pediatric CT: current technology and future innovation," Pediatr Radiol, Vol.32, pp.729-734, 2002. https://doi.org/10.1007/s00247-002-0800-x
  22. B. Wintersperger, T. Jakobs, P. Herzog, S. Schaller, and K. Nikolaou, "Aorto-iliac multidetector-row CT angiograpy with low kV setting:improved vessel enhancement and simulataneous reduction of radiation dose," Eur Radiol, Vol.15, No.2, pp.334-341, 2005. https://doi.org/10.1007/s00330-004-2575-y