Abstract
Currently, there are many studies being conducted around the world to reduce exposure dose to radiation for patients to receive medical treatments in a safe environment. We developed and fabricated of this shield that the patients are protected from the radiation and are need of safety control during breast imaging. In this study, for breast imaging, GE Senography 2000D were used and set at SID 65cm, 28kVp, and 63mAs. The measuring instrument was Fluke's Victoreen 6000-529. And we performed Face Block on with 30 patients. The chamber on the actual thyroid glands to take CC and MLO and measure the dosage before and after wearing the Face Block. For the results, after wearing the Face Block, exposure was decreased by 53.8%-100% and 65.8% in average in CC View and by 50%-100% and 60.7% in average in MLO View. The development of the Face Block that practically decreased the exposure dose of thyroid glands, crystalline eyes during breast imaging and reduced the patients' anxiety during breast imaging. The Face Block is expected to improve patients' satisfaction and contribute to reducing patients' exposure dose, but more efforts should be made to reduce exposure dose to medical radiation.
최근 환자가 안심하고 진료를 받을 수 있도록 의료 환경의 변화에 따라 의료 피폭 저감화 대책이 국내, 국제적으로 추진되고 있다. 연구자는 유방촬영 시 보다 안전하게 검사를 받을 수 있도록 환자의 방사선 피폭에 대한 방어 및 안전 관리가 절실 하다고 생각되어 본 차폐체를 개발하고, 제작하게 되었다. 임상에서 사용 중인 유방촬영 장치로 SID 65 cm, 조건 28 kVp, 63 mAs, 측정기는 Victoreen 6000-529(Fluke Co.)를 이용하여 제작한 Face Block을 사용하여, 환자 30명을 대상으로 실제 갑상샘 위치에 Chamber를 놓고 CC, MLO를 촬영해 Face Block장착 전, 후 선량을 측정하였다. 그 결과 Face Block 장착 후 CC View에서 평균 65.9%의 선량이 감소되었고, MLO View에서는 평균 60.7%의 선량이 감소되었으며, 영상의 질에는 큰 변화가 없었다. 따라서 유방촬영 시 갑상샘의 피폭을 줄이고, 실용적이고, 유용한 실제 Face Block을 개발한 것에 큰 성과가 있었다고 생각되며, 유방촬영 시 환자들의 피폭에 관한 불안감을 해소시켜, 환자 만족도 증대와 함께 환자 피폭선량 경감에 크게 이바지 할 것으로 기대하며, 피폭선량을 줄일 수 있는 방안을 끊임없이 노력해야 할 것이다.