DOI QR코드

DOI QR Code

Multiple Pivot loading 방법을 이용한 액체 환경에서의 수평방향 힘 교정

Lateral Force Calibration in Liquid Environment using Multiple Pivot Loading

  • Kim, Lyu-Woon (School of Mechanical Engineering, University of Ulsan) ;
  • Chung, Koo-Hyun (School of Mechanical Engineering, University of Ulsan)
  • 투고 : 2013.01.04
  • 심사 : 2013.02.16
  • 발행 : 2013.04.30

초록

Quantifying the nanoscale force between the atomic force microscopy (AFM) probe of a force-sensing cantilever and the sample is one of the challenges faced by AFM researchers. The normal force calibration is straightforward; however, the lateral force is complicated due to the twisting motion of the cantilever. Force measurement in a liquid environment is often needed for biological applications; however, calibrating the force of the AFM probes for those applications is more difficult owing to the limitations of conventional calibration methods. In this work, an accurate nondestructive lateral force calibration method using multiple pivot loading was proposed for liquid environment. The torque sensitivity at the location of the integrated probe was extrapolated based on accurately measured torque sensitivities across the cantilever width along a few cantilever lengths. The uncertainty of the torque sensitivity at the location of the integrated tip was about 13%, which is significantly smaller than those for other calibration methods in a liquid environment.

키워드

참고문헌

  1. Binnig, G., Quate, C. F., and Gerber, C., "Atomic Force Microscope," Phys Rev Lett, Vol. 56, pp. 930- 933, 1986. https://doi.org/10.1103/PhysRevLett.56.930
  2. Chung, K. H., Bhadriraju, K., Spurlin, T. A., Cook, R. F., and Plant, A. L., "Nanomechanical Properties of Thin Films of Type I Collagen Fibrils," Langmuir, Vol. 26, pp. 3629-3636, 2010. https://doi.org/10.1021/la903073v
  3. Mahaffy, R. E., Shih, C. K., MacKintosh, F. C., and KAs, J., "Scanning Probe-Based Frequency-Dependent Microrheology of Polymer Gels and Biological Cells," Phys Rev Lett, Vol. 85, pp. 880-883, 2000. https://doi.org/10.1103/PhysRevLett.85.880
  4. Stan, G., Krylyuk, S., Davydov, A. V., Levin, I., and Cook, R. F., "Ultimate Bending Strength of Si Nanowires," Nano Lett., Vol. 12, pp. 2599-2604, 2012. https://doi.org/10.1021/nl300957a
  5. Hutter, J. L. and Bechhoefer, J., "Calibration of Atomic-Force Microscope Tips," Rev. Sci. Instrum., Vol. 64, pp. 1868-1873, 1993. https://doi.org/10.1063/1.1143970
  6. Torii, A., Sasaki, M., Hane, K., and Okuma, S., "A Method for Determining the Spring Constant of Cantilevers for Atomic Force Microscopy," Meas Sci Technol, Vol. 7, pp. 179-184, 1996. https://doi.org/10.1088/0957-0233/7/2/010
  7. Gates, R. S. and Reitsma, M. G., "Precise Atomic Force Microscope Cantilever Spring Constant Calibration using a Reference Cantilever Array," Rev. Sci. Instrum., Vol. 78, pp. 086101, 2007. https://doi.org/10.1063/1.2764372
  8. Sader, J. E., Chon, J. W. M., and Mulvaney, P., "Calibration of Rectangular Atomic Force Microscope Cantilevers," Rev. Sci. Instrum., Vol. 70, pp. 3967- 3969, 1999. https://doi.org/10.1063/1.1150021
  9. Cain, R. G., Biggs, S., and Page, N. W., "Force Calibration in Lateral Force Microscopy," J. Colloid Interface Sci., Vol. 227, pp. 55-65, 2000. https://doi.org/10.1006/jcis.2000.6840
  10. Sader, J. E. and Green, C. P., "In-Plane Deformation of Cantilever Plates with Applications to Lateral Force Microscopy," Rev. Sci. Instrum., Vol. 75, pp. 878-883, 2004. https://doi.org/10.1063/1.1667252
  11. Ogletree, D. F., Carpick, R. W., and Salmeron, M., "Calibration of Frictional Forces in Atomic Force Microscopy," Rev. Sci. Instrum., Vol. 67, pp. 3298- 3306, 1996. https://doi.org/10.1063/1.1147411
  12. Varenberg, M., Etsion, I., and Halperin, G., "An Improved Wedge Calibration Method for Lateral Force in Atomic Force Microscopy," Rev. Sci. Instrum., Vol. 74, pp. 3362-3367, 2003. https://doi.org/10.1063/1.1584082
  13. Li, Q., Kim, K. -., and Rydberg, A., "Lateral Force Calibration of an Atomic Force Microscope with a Diamagnetic Levitation Spring System," Rev. Sci. Instrum., Vol. 77, pp. 065105, 2006. https://doi.org/10.1063/1.2209953
  14. Chung, K. H. and Reitsma, M. G., "Note: Lateral Force Microscope Calibration using Multiple Location Pivot Loading of Rectangular Cantilevers," Rev. Sci. Instrum., Vol. 81, pp. 026104, 2010. https://doi.org/10.1063/1.3276709
  15. Wang, F. and Zhao, X., "Effect of Contact Stiffness on Wedge Calibration of Lateral Force in Atomic Force Microscopy," Rev. Sci. Instrum., Vol. 78, pp. 043701, 2007. https://doi.org/10.1063/1.2720723
  16. Wagner, K., Cheng, P., and Vezenov, D., "Noncontact Method for Calibration of Lateral Forces in Scanning Force Microscopy," Langmuir, Vol. 27, pp. 4635-4644, 2011. https://doi.org/10.1021/la1046172
  17. Ryu, S. and Franck, C., "In Situ Hydrodynamic Lateral Force Calibration of AFM Colloidal Probes," Langmuir, Vol. 27, pp. 13390-13399, 2011. https://doi.org/10.1021/la201033e
  18. Higgins, M. J., Proksch, R., Sader, J. E., Polcik, M., Endoo, S. M., Cleveland, J. P., and Jarvis, S. P., "Noninvasive Determination of Optical Lever Sensitivity in Atomic Force Microscopy," Rev. Sci. Instrum., Vol. 77, pp. 013701, 2006. https://doi.org/10.1063/1.2162455
  19. Green, C. P., Lioe, H., Cleveland, J. P., Proksch, R., Mulvaney, P., and Sader, J. E., "Normal and Torsional Spring Constants of Atomic Force Microscope Cantilevers," Rev. Sci. Instrum., Vol. 75, pp. 1988- 1996, 2004. https://doi.org/10.1063/1.1753100
  20. Chung, K. H., Shaw, G. A., and Pratt, J. R., "Accurate Noncontact Calibration of Colloidal Probe Sensitivities in Atomic Force Microscopy," Rev. Sci. Instrum., Vol. 80, pp. 065107, 2009. https://doi.org/10.1063/1.3152335
  21. John R. Taylor. An Introduction Error Analysis: The Study of Uncertainties in Physical Measurements. Sausalito, Caifornia: University Science Books, 1996.