DOI QR코드

DOI QR Code

Design of pulsatile pump and performance test for pulsatile flow generation

맥동 유동 발생을 위한 맥동 펌프의 설계 및 특성 분석 연구

  • 주윤하 (경북대학교 기계공학부) ;
  • 김경원 (경북대학교 대학원 기계공학과) ;
  • 이연호 (경북대학교 대학원 기계공학과) ;
  • 곽문규 (경북대학교 기계공학부) ;
  • 이춘영 (경북대학교 기계공학부) ;
  • 이종민 (경북대학교 의과대학 영상의학과) ;
  • 박철우 (경북대학교 기계공학부)
  • Received : 2012.11.16
  • Accepted : 2013.02.13
  • Published : 2013.03.31

Abstract

For in-depth research of blood flow, it is important to create pulsating flow like the blood flow from heart beat. In this study, we developed a heart mimicking pulsatile pump and evaluated its performances. Main body of pump was produced using a piston pump, and its rpm and duty ratio was modulated by DC motor and encoder. To determine the part dimensions, principle stress theory and simple fluidic pressure analysis were used. The performance of pulsating pump was evaluated by comparing the pressure values and their deviations according to experimental variables. For the results, the output value of pressure followed the distribution of pulsating flow and its deviation was negligible. Through this study, we expect the established pulsating pump can be widely used in study of blood flow produce easy ways to related researchers.

혈류 유동 연구를 위해서는 심장의 맥동 유동을 만들어 내는 것이 중요하다. 본 연구에서는 동맥혈류의 유동 해석을 위해 새롭게 설계된 심장 박동 모사 맥동펌프의 제작 및 그 성능에 관하여 시험을 수행하였다. 피스톤 펌프를 이용한 기본몸체 설계를 수행하였고 직류모터와 엔코더를 사용하여 회전수와 가속비를 설정할 수 있도록 하였으며, 시험결과는 전 영역에서 맥동 유동의 분포를 잘 따르고, 반복되는 맥동 주기의 편차도 작게 나타남을 확인하였다.

Keywords

References

  1. J. J. H. Spek, J. A. Staessen, R. H. Fagard, A. P. Hoeks, H. A. Boudier, and L. M. Bortel, "Effect of age on brachial artery wall properties differs from the aorta and is gender dependent: a population study," Hypertension, vol. 35, no. 2, pp. 637-642, 2000. https://doi.org/10.1161/01.HYP.35.2.637
  2. W. W. Nichols and M. F. O'Rourke, McDonald's Blood Flow in Arteries (5th Ed.), Oxford University Press, 2005.
  3. A. A. Benam, M. T. Shadpour, N. Fatouraee, A. Pashaiee, and M. M. Khani, "A new system to analyze pulsatile flow characteristics in elastic tubes for hemodynamic applications," American Journal of Applied Sciences, vol. 5, no. 12, pp. 1730-1736, 2008 https://doi.org/10.3844/ajassp.2008.1730.1736
  4. N. D. Manring and Y. Zhang, "The improved volumetric-efficiency of an axial-piston pump utilizing a trapped-volume design," Journal of Dynamic Systems, Measurement, and Control, vol. 123, pp. 479-487, 2001. https://doi.org/10.1115/1.1389311
  5. K. Imaichi, Y. Murakami, H. Tsurusaki, and K. R. Jo, The Basis of Pump Design, Daeyoung Sa, 2002.
  6. I. S. Jun, K. J. Lee, K. O. Jo, and C. H. Jo, Theory and Reality of Pump, Kunsur Seoul Metropolitan, 2004.
  7. W. F. Hosford, Solid Mechanic, Cambridge University Press, 2010.
  8. K. X. Qian, P. Zeng, W. M. Ru, H. Y. Yuan, Z. G. Feng, and I. Li, "How to produce a pulsatile flow with low hemolysis?," Journal of Medical Engineering and Technology, vol. 24, no. 5, pp. 227-229, 2000. https://doi.org/10.1080/03091900050204278

Cited by

  1. Effects of stenotic severity on the flow structure in a circular channel under a pulsatile flow vol.38, pp.2, 2014, https://doi.org/10.5916/jkosme.2014.38.2.140
  2. Analysis of the time-velocity curve in phase-contrast magnetic resonance imaging: a phantom study vol.24, pp.suppl2, 2019, https://doi.org/10.1080/24699322.2019.1649066